  • LPA-10kV A Power Module was removed, cleaned, and re-seated with no improvement. The spare module was also cleaned and tested with seemingly no difference. The original module we have been using was replaced back into the unit.
  • The cover of the lower unit was removed to inspect the internals. Everything was clean and dust free.
  • In the upper unit, the 'brains' of the Bathy, boards A00107 and A00108 were taken out, cleaned and inspected for any damage. Resistance was measured across suspect resistors that had failed previously on Exp. 350on a previous expedition. There was no difference between the boards in use and the new spares. The in use boards were installed back into the unit. New unused spares put back into the drawer. 

...