Susan Boehm

Summary



New batteries put in both digital micrometers- handheld and petrothin.  I will remove the batteries at end of expedition so that they last longer.

All three vacuum pumps cleaned and oil replaced.

Experimented with procedure to impregnate laminated sediment samples to determine if this is feasible to do on board.


LP-50

Used for slide frosting and lapping.  Very consistent and reliable.  I used one jig for both settings and adjusted the chuck face up or down when needed.

Replaced gaskets in the slurry drum and on the polisher drum, no more leaking. 

Petrothin


PM5 Polisher

Works well but squeaky when WG2 is rotating.

Samples and methods



Miscellaneous


Two Welch vacuum pumps (on Petro Thin and Freeze Dryer) and one Edwards Pump (LP50) were drained, cleaned and refilled with oil.  I ran them for a few minutes, and flushed out the oil two more times because there was significant build up.

...