Instrument/EquipmentPMS TaskFrequencyDate of last maintenance
ALKALINITY/pH - Metrohm 794 Basic Titrino Autotitrator Check 3M KCl level in the electrode, top up if lowMonthly

 


Check storage solution (Metrohm Storage Solution) level in the electrode storage vial, top up if lowMonthly

     


Check DI water level in dispensing tip storage vial, top up if lowMonthly

   


Change out desiccantWhen ~50% pink

  


Check water level in water bath, top up if lowWeekly

        

BALANCE - Cahn Microbalance No PMS necessary--
BALANCE - Mettler-Toledo XS204 Dual-Balance SystemNo PMS necessary--
BARNSTEAD WATER PURIFICATION SYSTEMRun water from the DI tank and nanopure outlet for 5 minutesWeekly

13      


Replace the pre-Barnstead filter under the counter (UTS CM0063)When discolored

 


Replace the three pretreatment cartridges Periodically

 


Replace the Genpure filter cartridgeAnnually or when necessary

 


Replace the UV lamp Every two years or when necessary

  


Replace the 0.2 micron final filter Annually or when necessary

  


Replace the two RO membranes (access through the back panel of the RO unit)Annually or when necessary

  

CARVER PRESSES - Carver 25 Ton Hydraulic PressesCheck hydraulic fluid level in Autopaks - should only be necessary if you have had a large spill and thus removed a significant amount of the fluid - top up with Carver hydraulic fluid (under the sink)When necessary

 

CENTRIFUGE - Eppendorf Centrifuge 5810No PMS necessary--
CHLORINITY - Metrohm 785 DMP Titrino AutotitratorCheck DI water level in the electrode storage vial, top up if lowWeekly

    


Check DI water level in the dispensing tip storage vial, top up if lowWeekly

   


Change out desiccantWhen ~50% pink

   

CHNS - Thermo FlashEA Elemental Analyzer  Change out magnesium perchlorate in the water trap (if using trap) (the new instrument did not come with a trap installed)Once a month when in use

 

COULOMETER - UIC CM5015 CoulometerNo PMS necessary


FREEZE-DRYER - Labconco FreeZone 12 Freeze Dry SystemCheck vacuum pump oil level - should be between low and high mark - top up if lowOnce a month when in use

    


Empty the vacuum pump oil mist filter and discard oilWhen necessary

  


Change vacuum pump oilWhen dirty

 

FUME HOODS - Labconco Chemical and Acid Digestion HoodsInspect the drains for trash/debris and remove if present. Pour/run water down the drain in fume hood #3 and #4. Pour water down the fume hood #2 drain at the end of an expedition. Fume hood #1 drain is capped. Do not pour water in fume hood #1.   Once an expedition

  

GC GC2 - Agilent 6890N Gas ChromatographReplace gas trapsEvery 2 years or when necessary

   


Replace front inlet septum When necessary

  


Replace autosampler tension cordWhen needles are getting bent

 


Replace autosampler syringe plungerWhen necessary

  


Replace autosampler syringeWhen necessary

 

GC NGA1 - Agilent 7890A Gas ChromatographNo PMS necessary--
GC NGA2 - Agilent 7890A Gas ChromatographNo PMS necessary--
HYDROGEN GENERATORS - Parker-BalstonCheck water level in the generator(s) in use, top up if lowWeekly

  


Generator 1 (SRA): Replace "6-month maintenance kit" componentsAnnually or when necessary

 


Generator 1 (SRA): Replace "24-month maintenance kit" componentsEvery few years

 


Generator 2 (NGAs): Replace "6-month maintenance kit" componentsAnnually or when necessary

Generator 2 (NGAs): Replace "24-month maintenance kit" componentsEvery few years

  


Generator 3 (NGAs): Replace "6-month maintenance kit" componentsAnnually or when necessary

Generator 3 (NGAs): Replace "24-month maintenance kit" componentsEvery few years

 

IC - Metrohm 850 Professional Ion ChromatographCirculate eluent through the system for one hourWeekly

  


Replace soda lime in eluent bottle CO2 scrubbers When ~50% pink or once a year

 


Replace inline filters (1 on cation flow path, 3 on anion flow path)Once an expedition (in the beginning)

   


Replace anion guard columnOnce an expedition (in the beginning)

      


Replace cation guard columnOnce an expedition (in the beginning)

  


Replace anion columnWhen necessary

 


Replace cation columnWhen necessary

 


Replace the CO2 suppressor CO2 scrubber cartridge (soda lime)  When ~50% pink

 


Replace the beads in the CO2 suppressor H2O scrubber cartridge (put old beads in the aft oven at > 140 C on a kimwipe-lined tray and bake back to dark orange color)When beads lose their color

   


Replace the autosampler ultrafiltration membraneEvery 200 samples or when necessary

    


Replace all peristaltic pump tubing for autosamplerPeriodically

   


Replace all peristaltic pump tubing for suppressorPeriodically

 


Clean bearings on peristaltic pump for suppressorPeriodically

 

ICP-OES - Agilent 5110 Inductively Coupled Plasma Optical Emission SpectrometerClean the instrument water inlet filterPeriodically

 


Clean the air inlet filterPeriodically

    


Clean the water chillerPeriodically

 


Clean the coneWhen necessary

   


Clean the snoutPeriodically

  


Replace the HR torchWhen necessary

  


Replace the IW torchWhen necessary



Replace the HR nebulizer (CM5228, Agilent G8010-60255)Periodically or when necessary

 


Replace the IW nebulizer (CM5228, Agilent G8010-60255)Periodically or when necessary



Replace autosampler peristaltic pump tubingPeriodically

 


Replace "blue" peristaltic pump tubingPeriodically

  


Replace "white" peristaltic pump tubingPeriodically

  

OXYGEN SENSOR - AMI Oxygen AnalyzerReplace the sensor (replacement sensor in LO office cabinet)Every two years

 

SPECTROPHOTOMETER - Agilent Cary 100Replace autosampler peristaltic pump tubingWhen necessary

   


Replace instrument peristaltic pump tubingWhen necessary

 


Replace "Lamp, visible source, Varian #5610021700", CM1217Periodically

 

SRA - Weatherford Source Rock AnalyzerReplace moisture trap contentsBefore use

 


Replace CO2 trap (or contents)Before use

  


Replace drieriteBefore use

 

...