All microscopes, objectives and oculars as well as the transformer boxes were checked and cleaned by the SERCO contractor Emile Meylan. This is the list of findings:
Microscope property number | Microscope Type | Diagnosis |
---|---|---|
90871 | Axioscope A1 POL | Stage counterweights had to be cut down which was accomplished for all three specimen holder sets of that type on board |
90869 | Axioscope A1 POL | XY stage faulty (inner pulley system problem), Serco can’t fix this, needs to go back to Zeiss – tested 3 other stages, 2 out of 3 are broken and will be send to College Station to be forwarded to Zeiss |
91187 | Axio Imager A2 | - known issue with PCBA board, which has missing parts on order, - scope is set up for fluorescence, unsure why in that spot |
91188 | Axio Imager A2 (MBIO) | No issue reported. |
90229 | Axioplan POL | Bertrand lens is de-laminated, attachment not useful, so it was removed. |
90207 | Axioscope POL (Thin Section) | No issues reported. |
90257 | Axiophot | Stage couldn’t be moved far enough to focus on specimen, it was discovered that gear inside wasn’t lined up, Serco had to remove focus knob covers and align 15 O-rings back to where they are needed in order to move stage to correct positions. Recommendation from Serco – don’t tinker with inner workings of scopes if not absolutely necessary, send to Zeiss or have Zeiss rep service call. |
91060 | Stereoscope | Magnification knob doesn’t stop at 1x or 8x, no way to fix it as body is glued shut. Stored in Imaging Lab as fall back for high usage expeditions. |
90812 | Stereoscope | No issue reported. |
91105 | Stereoscope | No issue reported. |
90811 | Stereoscope | No issue reported. |
90812 | Stereoscope | No issue reported. |
90228 | Stereoscope SV11 (reserve in Imaging Lab) | No issue reported. |
ODP00584 | Old grey stereoscope (reserve in Imaging Lab) | No issue reported. |