Introduction

This guide is written to help guide scientists and technicians in setting up for and measuring section measurements with IMS version 9.2. For more detailed information regarding the SRM – IMS software, please refer to the SRM-IMS User guide.

Initializing the Program

Figure 1. IMS-SRM desktop shortcut.

Open IMS Main from the Desktop shortcut or from the Windows Start Menu. Icon is shown in Figure 1.
The program will begin an initialization process to check communications with the M-drive motor, the degausser, and the SQUIDS. The software will load the most recent configuration files and will home the tray on the track.

Figure 2. IMS-SRM main screen.



Once the program launches, the user will see the main IMS-SRM window (Figure 2). The leftmost panel contains the controls for changing the setup and for starting the SRM measurements. The center panel allows the user to view the data as it is collected and monitor the measurement process. The right panel display indicates which measurement sequence is selected and which step is being executed, as well as other pertinent information.

...

  1. Select DAQ > Measurement Editor from the drop-down menu. The Measurement Editor window will open (Figure 3).


    Figure 3. Measurement Editor Window.

  2. Set the desired sample interval, number of points to average, leader/trailer length, and measurement speed.
  3. If the sample interval has changed, you will likely receive a warning (DAQ Valid? indicator). When the DAQ Valid? indicator is red, the user must measure the tray background.
  4. Click OK to exit the Measurement Editor window.

...

To take a background measurement, select Instruments > SRM: Section Tray Background or Instruments > SRM: Discrete Tray Background (depending on the tray type used). The window in Figure 4 will appear.

Image Modified

Figure 4.Section Tray and Discrete Tray Background Windows 

  1. Set the desire demagnetization level.
    The default action for the tray background measurement is to demagnetize the tray and then to measure the empty tray. The user may adjust the demagnetization level (maximum of 80 mT) or chose to measure without demagnetizing the tray.
  2. Select START to begin the background measurement.
  3. When the measurement has finished the user will be asked if they wish to SAVE the new tray data.

...

The user must define the shape and orientation of the samples to be measured. In most cases this will be Section Half–Archive, but alternative shapes and orientations are available.

Figure 5. Sample Preset Editor Window.

  1. To set up a sample preset shape, select DAQ > Sample Preset Editor (Figure 5).
  2. For an archive section piston core select Section Half–Piston on the left side of the screen and chose Archive Half at the bottom of the screen.
  3. Click in the white square checkbox on the right side of the screen and then click on one of the sample preset buttons. A window will open allowing the user to name the button (Figure 6). Chose Update Button once you have named the preset.
  4. Select OK to exit the Preset Editor.

Figure 6. Edit Preset Button Name

Create a Measurement Sequence

Image Modified

Figure 7. Sequence Editor Window 

The user must create a measurement sequence or select an existing sequence they wish to use. If a new sequence is necessary, the user may create a new sequence prior to starting a measurement. If the desired sequence has already been created, the sequence editor window will not be necessary. Using sequence editor is only necessary if the desired sequence does not already exist.

...

  1. To start a measurement select the START button on the left hand side of the main SRM-IMS screen (Figure 2). The Sample Information window will open (Figure 9).
  2. Check that the Current Sequence listed at the bottom is the sequence you wish to use. If it is not the right sequence, use the Measurement Sequence tab to select the proper sequence.
  3. Ensure the correct sample type and orientation is selected on the left side of the screen.
  4. Place the cursor in the SCAN window.
  5. Scan the Archive End cap label using the barcode scanner.
    The barcode scanner is the easiest method of sample information entry. The user may use the LIMS tab, which allows the user to look in the LIMS database for a core section, or they may manually enter the Sample ID and Text ID. These methods are only recommended if the barcode scanner is not working.
  6. Verify the LIMS Length matches the section length and verify the Sample ID is correct.
    If the LIMS Length does not match the actual section length after the bar code is scanned, select the Manual tab (Figure 10). Adjust the USER length to match the actual core length and select Use ID and Length values.
  7. Click Measure to begin measurement. A window similar to Figure 2 will open and display the measured data in real time.
  8. When the sequence is complete the Core Information window will reopen.
  9. Repeat steps 4–7 for further sections.

Figure 9. Sample Information Window.

Figure 10. Manual Entry Sample Information Window.