- Start: Turns on the high voltage which generates an electron beam. Must press start to generate an image.
- Fast: Image updates quickly. Useful to select 'Fast' while changing Field of View.
- Slow: Image updates slowly. If in calm sea state, can produce a higher quality image.
- Reduce: Takes just the center portion of the screen to update.
- Freeze: Stops updating the image and preserves the image as it was at the time of pressing Freeze.
- Magnification: Control to increase or decrease magnification. Ranges from x40 to x30,000.
- Auto B/C: Click this to automatically adjust the brightness and contrast of the image.
- Auto Focus: Click this to automatically focus the image. Image Mode:
- 5 kV / 15 kV / Analy: Sets the beam voltage between either 5kV, 15kV, or Analy
- Quick Save: Saves the current image on the screen. Captures a lower resolution image of 640x520 pixels.
- Save: Re-captures the image and saves it at a higher resolution image of 1280x1040 pixels.
4. Press Start
5. Navigate around using the control knobs on the front of the SEM.
6. When ready to save an image select Quick Save for low resolution images (also good during high seas) or Save for high resolution images (good during calm seas).
7. When done with the microscope select Stop in the upper left corner of the software.
File Naming Convention
Naming convention for SEM files:
- EXP number
- Site
- Hole
- Core
- Section
- Sample
- Taxa/specimen/piece
- No. 1, 2, 3…
E.g.
349_U1431A_1H_1_10-11cm_Emiliania_huxleyi_1,
349_U1431A_1H_1_10-11cm_Emiliania_huxleyi_2,
349_U1431A_1H_1_10-11cm_Emiliania_huxleyi_3, …
Storage convention for SEM files on the local PC:
- SEM
- Site
- Hole
- Sample
E.g.
Folder SEM
Folder 1431
Folder 1341A
Folder U1341A_1_H_1_1-11cm (all files for this SEM sample)
NOTE: All files need to be moved to USERVOL and DATA1 by the end of the expedition.