Total Organic Carbon Analyzer User Guide
This file is for editing content only! Cross-references and hyperlinks do not work. Use track changes and return file to peters@iodp.tamu.edu with copy to techdoc@iodp.tamu.edu.
Manual Information

Author(s):

Y. Vasilyeva (08/29/08)

Reviewer(s):

 

Editor(s):

K. Graber (7/14/08), J. Hesse

Manager Approval (Name, Title, Date):

 

Audience:

RS, Scientists

Origination Date:

 

Current Version:

Draft 12/08/08 LLP

Revised:

 

Domain:

Chemistry

Analysis:

Carbon/Carbonate Analysis


User Guide Contents

Topic

See page…

Introduction

2

Apparatus, Reagents, & Materials

3

Instrument Calibration/Verification

9

Sample Preparation

12

Sample Analysis

17

Quality Assurance/Quality Control

18

LIMS Integration

19

Health, Safety, & Environment

20

Maintenance & Troubleshooting

21

Software Guide

25

Anchor
RTF320035003200380032003a00
RTF320035003200380032003a00
Introduction
Theory of Method
The Total Organic Carbon (TOC) instrument analyzes liquid samples by physically combusting the samples at 680°C for TOC or by chemically combusting the samples in 25% phosphoric acid for inorganic carbon (IC). The instrument measures the amount of carbon degraded to CO2 using nondispersive infrared gas analysis at 4.3 µm wavelength, calculated based on Lambert-Beer's law.
Anchor
RTF4d0065007300750072006500
RTF4d0065007300750072006500
Total Carbon (TC)
To analyze for Total Carbon (TC) a sample is delivered to the TC combustion tube filled with platinum catalyst and heated to 680°C. Carrier gas (purified air or oxygen) , along with the CO2 generated by combustion, flows through the reaction vessel to the dehumidifier and scrubber, where it is cooled, dried, and purified. It then it passes into the nondispersive infrared detector, which generates an analog signal that is digitized and processed by the TOC Control software. The peak area is proportional to the concentration of total carbon in the sample.
Anchor
RTF4d0065006100730075007200
RTF4d0065006100730075007200
Inorganic Carbon (IC)
During inorganic carbon analysis the liquid sample is introduced into the reaction vessel, which is half filled with 25% phosphoric acid. The sample is purged with carrier gas (high purity air) through the acid solution. The phosphoric acid reacts only with inorganic carbon components (CO32- and HCO3), transforming them to CO2. The CO2 is carried fromthe IC vessel through the dehumidifier and scrubber to the NDIR detector.
Total Organic Carbon
TOC concentration can be obtained by subtracting the IC concentration from the TC concentration.
Note: Before operating the TOC system every operator must get either additional training from a senior technician or carefully study all three manufacturer TOC/ASI/Software manuals.
Anchor
RTF360032003300380036003a00
RTF360032003300380036003a00
Apparatus, Reagents, & Materials
Laboratory Apparatus

  • 50 mL Shimadzu autosampler glass vials, 10 pc
  • 6 mL Shimadzu autosampler glass vials, 50 pc
  • Wax paper
  • Plastic ring caps for 50 mL autosampler vials
  • Disposable caps for 6 mL autosampler vials
  • 1000 mL, and 50 mL volumetric flask

Reagents

  • Zero water: Fresh distilled water from the filtering system, carbon (TC) free, stored in a CO2-free container. Use this water to rinse vials, make calibration standards, and run blanks. Do not expose the fresh distilled water to air to prevent dissolution of CO2 into it.
  • Potassium hydrogen phthalate, reagent grade
  • Sodium hydrogen carbonate, reagent grade
  • Sodium carbonate, reagent grade, heated 1 hr at 285°C and cooled in a dessicator
  • Phosphoric acid, 25%

Standard Solutions
Standard solution concentrations are not stable because of instant absorption of CO2 from the atmosphere. To slow down this process, standard solutions should be stored in airtight containers in the refrigerator. 1000 ppm stock solutions can be stored for 2 months; 100 ppm diluted standards can be stored for 1 week.

  • Total Carbon Stock
  • 1000 mg C/L (1000 ppm C) stock solution: dissolve 2.125 g of potassium hydrogen phtalate in zero water in a 1L volumetric flask.
  • At room temperature, potassium hydrogen phtalate doesn't dissolve in water very quickly, and it's grains are almost transparent, so the technician needs to make sure the solution is homogeneous before using it.
  • 100 mg C/L (100 ppm) intermediate stock solution: dilute 10 mL of the the 1000 ppm standard stock solution in zero water in a 100 mL volumetric flask.
  • Inorganic Carbon Stock
  • 1000 mg C/L (1000 ppm C) stock solution: dissolve 3.50 g of sodium hydrogen carbonate and 4.41 g of sodium carbonate in zero water in a 1 L volumetric flask.
  • 100 mg C/L (100 ppm C) intermediate stock solution: dilute 10 mL of the 1000 ppm standard stock solution in zero water in a 100 mL volumetric flask.

Hardware

  • Shimadzu TOC-5000A instrument
  • Shimadzu ASI-5000A Autosampler
  • HPxw 4400 PC workstation (shared with GC2 system)
  • Whatman/Parker Balston TOC 78-40 Gas Generator (UHP air tank can be substituted)
  • TOC-5000A System Components.

TOC-5000A Analyzer
The TOC-5000A analyzer is configured to directly determine the content of total carbon (TC) and inorganic carbon (IC) in 2–10 mL aqueous samples within a carbon range of 200 ppb to 1000 ppm.
Total organic carbon (TOC) cannot be measured directly in samples but can be calculated:
TOC = TC – IC
Samples run on the TOC-5000A system are analyzed for TC and IC sequentially, followed by automatic software calculation of TOC.
Figures 1-2 and 1-3 show the general configuration of the TOC-5000A hardware.

Instrument

Specification

Method

Combustion/nondispersive infrared gas analysis

Combustion temperature

680°C

Measurement range

4 ppb to 4000 ppm

Anaysis time

2–3 min each for TC and IC

Repeatability

<2000 ppm: 1% of full scale
2000–4000 ppm: 2% of full scale

Sample introduction

Automatic injector

Injection volume

50–2000 µL

Ambient temperature

5°–35°C

Dimensions

480 mm W x 520 mm D x 480 mm H

Weight

~46 kg

  • Schematic of TOC-5000A Interior (Front View).
  • Photograph of TOC-5000A Interior (Front View).
  • Anchor
    RTF340037003600330035003a00
    RTF340037003600330035003a00
    Close-up Front View Flow Lines in Upper Right.

Autosampler
The autosampler allows automated carbon analysis in multiple samples.

  • Anchor
    RTF310031003800320031003a00
    RTF310031003800320031003a00
    Autosampler Back View.
  • Autosampler Closeup.

Gas Generator
The Parker Balston (Whatman)78-40 TOC Gas Generator (Figure 1-7)converts compressed air into high purity carrier/combustion gas for the TOC analyzer with hydrocarbon levels < 0.1 ppm and CO2 level < 1 ppm.
The Gas Generator should be fed from the air compressor, not from an air tank, because of the high rate of air consumption. Incoming air should be relatively free of dust and water (typical air compressor produced air should meet these requirements).

  • Anchor
    RTF320039003800300031003a00
    RTF320039003800300031003a00
    TOC Gas Generator.

Anchor
RTF350036003200300035003a00
RTF350036003200300035003a00
Instrument Calibration/Verification
Anchor
RTF320034003400390030003a00
RTF320034003400390030003a00
Preparing Calibration Standards
Total Carbon
In a 100 mL volumetric flask, make up the following calibration levels to bracket general expected TC levels from the 100 ppm intermediate stock solution:

  • 100 ppb: 0.1 mL of 100 ppm + 99.9 mL zero water
  • 500 ppb: 0.5 mL of 100 ppm + 99.5 mL zero water
  • 1 ppm: 1.0 mL of 100 ppm + 99 mL zero water
  • 2 ppm: 2.0 mL of 100 ppm + 98 mL zero water
  • 5 ppm: 5.0 mL of 100 ppm + 95 mL zero water

Inorganic Carbon
In a 100 mL volumetric flask, make up the following calibration levels to bracket general expected IC levels from the 100 ppm intermediate stock solution:

  • 100 ppb: 0.1 mL of 100 ppm + 99.9 mL zero water
  • 500 ppb: 0.5 mL of 100 ppm + 99.5 mL zero water
  • 1 ppm: 1.0 mL of 100 ppm + 99 mL zero water
  • 2 ppm: 2.0 mL of 100 ppm + 98 mL zero water
  • 5 ppm: 5.0 mL of 100 ppm + 95 mL zero water

Anchor
RTF310030003100310030003a00
RTF310030003100310030003a00
Setting up the Sample Table for Calibration Curve

Step

Action

    From the TOC Control screen select File > Open Sample Table and open a new or previously saved calibration Sample Table (Fig 8).

      Click in the first line of the Sample Table and select Edit > Insert Standard.

        On the TOC Control screen toolbar, select View > Calibration Curve List.

          On the Calibration Curve List dialog box, select New (Fig 9).

            On the Conditions tab (Fig 10), enter the following:

            • Title and File Name for the calibration
            • Analysis Type (TC or IC)
            • Change Calculation Method to Lin. Regression without Zero Shift (FIG).
            • The Range and Injection Volume fields will be autofilled after completing information in the Data tab.

              Select the Data tab and enter information requested into the data table.

                Click OK and save the calibration information and Yes to use as method default parameters.

                  Save the Sample Table.

                  • Calibration Sample Table.


                  • Calibration Curve List Dialog Box.


                  • Calibration Curve Conditions Tab.

                  Running a Calibration Curve

                  Step

                  Action

                    Place calibration curve standards in autosampler spaces S1–S8.

                      Verify status of instrument and baseline: View > Background Monitor then close Background Monitor screen.

                        Select Measure > Start or click Start on the instrument control bar. When the Standby screen opens, select Keep Running and click OK.

                          To monitor the status of the measurements, select View > Realtime Window or select Realtime Window on the toolbar.

                            When the calibration runs are complete the instrument status bar will indicate Ready.

                            Evaluating the Calibration Curve

                            Step

                            Action

                              Select View > Calibration Curve or click the Calibration Curve button on the toolbar.

                                Select the Curve tab to see the curve data in graphical format.

                                Wiki Markup
                                *\[Author: Need a screen shot of this.\]*

                                  Check the correlation coefficient (R2), which was calculated using the linear relationship of the concentration vs. area. R2 > 0.995 indicates an acceptable calibration.

                                    Select the Data tab to view standard deviation and coefficient of variation statistics.

                                    Wiki Markup
                                    *\[Author: Need screen shot.\]*

                                      Individual peaks or injections can be viewed using View > Peak Profile > Display.

                                        If the calibration is acceptable, specify the calibration file (CAL) name in the method and close the Calibration window.

                                        Running a Calibration Verification

                                        Step

                                        Action

                                          Run a midpoint curve standard as an "unknown" sample in the Sample Table.

                                            Apply the calibration curve to calculate the sample results.

                                              Open View > Injection Table or click the Injections button on the toolbar to examine the sample results in the Concentration column and the result of the Outlier Test. For acceptable calibration verification, CV should be >0.90.

                                                Close the Injection Table screen.

                                                Sample Preparation
                                                Overview
                                                Samples are prepared by squeezing interstitial waters from sediment cores. Total organic carbon is run on unfiltered interstitial water samples; dissolved organic carbon (DOC) is run on interstitial water samples filtered through a 0.45 µm membrane.
                                                Interstitial waters are diluted ~1:5 before running on the TOC analyzer and the analyzer is calibrated for quantitation of sample results. Running samples involves the following steps:

                                                Step

                                                Action

                                                  Set up hardware (see Setting up the Sample Table for Calibration Curve)

                                                    Set up software (see Preparing TOC Software )

                                                      Set up autosampler (see Preparing the Autosampler)

                                                        Set up sample table (see Setting up a Sample Table and Selecting a Calibration Curve)

                                                          Run samples (see Running Samples)

                                                          Preparing TOC Analyzer Hardware

                                                          Step

                                                          Action

                                                            Make sure tubing and cables are connected properly to the instrument.

                                                              Check compressed air feed line (65–80 psi) for the Gas Generator (if using an ultra-high purity air tank the delivery pressure be set to 80 psi (not to exceed 100 psi).

                                                                Check that the Gas Generator is in ready condition. THC light green and no CHECK SYSTEM/ OVERFLOW lights on.

                                                                Wiki Markup
                                                                *\[Author: Need photo of green light on Gas Generator.\]*

                                                                  Fill the humidifier with distilled water until the water level is between the white lines. Make sure the rubber stopper is fully plugged in.

                                                                    Fill IC vessel with 75–100 mL of 25% phosphoric acid solution.

                                                                      Make sure all drain lines have a negative slope alignment. Empty the drain container, if required.

                                                                        Check position of three-way cock arm. The cock arm should be in the left position (Fig 4).

                                                                          Ensure TOC main power switch is turned on and black switch on front top right side (behind instrument door) is in PC CONTROL position.

                                                                          • Anchor
                                                                            RTF350037003000360037003a00
                                                                            RTF350037003000360037003a00
                                                                            Software TOC Control Screen with Blank Sample Table.

                                                                          Anchor
                                                                          RTF360033003000380035003a00
                                                                          RTF360033003000380035003a00
                                                                          Preparing TOC Software

                                                                          Step

                                                                          Action

                                                                            Click on the TOC Control icon on the computer desktop (see #1 on Fig 11). Type DAQ for username and password to open the TOC Control screen.

                                                                              From the Toolbar on the TOC Control screen, select Measure > Connect (#3 on Fig 11), and at the bottom of the Control screen click the TOC button (#4 on Fig 11).
                                                                              The message bar (#5 on Fig 11) should say Connected.

                                                                                Check the carrier gas flow meter and pressure gage on the front panel of the TOC analyzer.

                                                                                Wiki Markup
                                                                                *\[Author: Need a photo of this meter and pressure gage.\]*

                                                                                Flow should be 150 mL/min, and pressure should be ~450 kPa. Adjust flow or pressure to target values using the Carrier gas pressure controller and Carrier gas flow controller next to the flow meter (see Fig 4).

                                                                                  On the TOC Control screen, select View > Background Monitor (#6 on Fig 11) to open a background window.

                                                                                    Wait ~2 min and then make sure all parameters read OK on screen (exception: oven temperature will take ~30 min to reach 680°C).

                                                                                    Wiki Markup
                                                                                    *\[Author: Need a screen shot of the Background window.\]*

                                                                                    Close the Background window.

                                                                                      On the TOC Control screen select Options > Instrument Conditions (#7 on Fig 1) to set up and/or verify TOC, ASI, and RS232 settings as follows:

                                                                                      • TOC
                                                                                      • Furnace = on
                                                                                      • Auto ranging = on
                                                                                      • Inj volume = on
                                                                                      • Auto regenerating = on
                                                                                      • Syringe size = 250 µm
                                                                                      • ASI
                                                                                      • ASI used = on
                                                                                      • High sensitivity vials = on
                                                                                      • Rinse = on
                                                                                      • No. of washes = 4
                                                                                      • RS232
                                                                                      • RS232 = COM1
                                                                                      • Baud rate = 2400
                                                                                      • Parity = none
                                                                                      • Stop bits = 1

                                                                                      Anchor
                                                                                      RTF350034003400330032003a00
                                                                                      RTF350034003400330032003a00
                                                                                      Preparing the Autosampler

                                                                                      Step

                                                                                      Action

                                                                                        Fill autosampler rinse bottle (Fig 5) with fresh zero water.

                                                                                          Fill high-sensitivity 50 mL autosampler glass vial with zero water and place it to S1 tray position. Remove tray cover or make sure it's in the correct position.

                                                                                          • Each vial in the autosampler tray must be filled to the top and covered immediately with wax paper or sealed with special caps to prevent CO2 penetration. After the sealed surface is pierced by the autosampler needle, the vials will remain usable for only a limited time (usually a few hours).

                                                                                            Remove bubbles from the syringe (Detect Zero point): on the TOC Control screen select Options > Maintenance > Zero point detection.
                                                                                            The ASI arm and tray will start moving and the syringe/4-port valve will perform the bubble removing procedure.

                                                                                            Anchor
                                                                                            RTF380036003500350035003a00
                                                                                            RTF380036003500350035003a00
                                                                                            Setting up a Sample Table and Selecting a Calibration Curve

                                                                                            Step

                                                                                            Action

                                                                                              From the TOC Control screen select File > Open Sample Table to create a new Sample Table (#8 in Fig 11) or open an old Sample Table (#9 in Fig 8).

                                                                                                Fill in the Sample Table with sample type, name, ID, and analysis.

                                                                                                  To specify a method on the Sample Table, select View > Methods on the TOC Control screen to open a saved method or create a new method.

                                                                                                    Fill in information on the General and Method tabs of the Method screen (Fig 12).

                                                                                                      To specify a calibration curve to apply to the samples, select View > Calibration on the TOC Control screen to open a dialog box showing a list of calibration curves saved in the CALIBR folder (Fig 9).

                                                                                                        On the Calibration Curve List dialog box, click Open to select a saved calibration curve.

                                                                                                          Save the Sample Table.

                                                                                                          • Anchor
                                                                                                            RTF380032003700380034003a00
                                                                                                            RTF380032003700380034003a00
                                                                                                            IC Method Tabs.

                                                                                                          Anchor
                                                                                                          RTF320032003000360032003a00
                                                                                                          RTF320032003000360032003a00
                                                                                                          Sample Analysis
                                                                                                          Anchor
                                                                                                          RTF350031003400330038003a00
                                                                                                          RTF350031003400330038003a00
                                                                                                          Running Samples

                                                                                                          Step

                                                                                                          Action

                                                                                                            Load the sample vials into the autosampler in positions that correspond to the Sample Table.

                                                                                                              Verify status of instrument and baseline: View > Background Monitor, then close Background Monitor screen.

                                                                                                                Select Measure > Start or click Start on the instrument control bar. When the Standby screen opens, select Keep Running and click OK.

                                                                                                                  To monitor the status of the measurements, select View > Realtime Window or select Realtime Window on the toolbar.

                                                                                                                    When the sample runs are complete the instrument status bar will indicate Ready.

                                                                                                                      Sample results calculated from the loaded calibration file will display in the Sample Table Mean Area and Result columns.

                                                                                                                      Anchor
                                                                                                                      RTF340030003400320032003a00
                                                                                                                      RTF340030003400320032003a00
                                                                                                                      Quality Assurance/Quality Control
                                                                                                                      Analytical Batch
                                                                                                                      Blanks
                                                                                                                      The TOC analyzer has an automated TC Blank Check Program that runs zero water blanks from 4 different vials a total of 10 times.

                                                                                                                      Step

                                                                                                                      Action

                                                                                                                        Place vials filled with zero water in autosampler positions S1–S4.

                                                                                                                          On the autosampler control panel, open the Maintenance screen and select TC Blank Check.

                                                                                                                            The autosampler injects S1 twice, S2 three times, S3 three times, and S4 twice and then returns to Home.


                                                                                                                            Calibration
                                                                                                                            Calibration Verification
                                                                                                                            Precision
                                                                                                                            Accuracy

                                                                                                                            Anchor
                                                                                                                            RTF310038003400310036003a00
                                                                                                                            RTF310038003400310036003a00
                                                                                                                            LIMS Integration
                                                                                                                            Overview
                                                                                                                            TC/IC/TOC analysis results are manually entered into the LabWare LIMS client by the technical staff once the scientist(s) is (are) finished reducing the data.
                                                                                                                            LIMS Components
                                                                                                                            Results are stored in the LIMS database associated with an analysis code and an analysis component. Analysis codes and their components, definitions, and units are listed below.

                                                                                                                            Analysis

                                                                                                                            Component

                                                                                                                            Definition

                                                                                                                            Unit

                                                                                                                            TOC

                                                                                                                            carbon_total

                                                                                                                            Concentration

                                                                                                                            ppm

                                                                                                                             

                                                                                                                            inorganic_carbon_total

                                                                                                                            Concentration

                                                                                                                            ppm

                                                                                                                             

                                                                                                                            organic_carbon_total

                                                                                                                            Concentration

                                                                                                                            ppm

                                                                                                                             

                                                                                                                            run_test

                                                                                                                            link to calibration test

                                                                                                                             

                                                                                                                            TOC_QAQC

                                                                                                                            carbon_total

                                                                                                                            Concentration

                                                                                                                            ppm

                                                                                                                             

                                                                                                                            inorganic_carbon_total

                                                                                                                            Concentration

                                                                                                                            ppm

                                                                                                                             

                                                                                                                            organic_carbon_total

                                                                                                                            Concentration

                                                                                                                            ppm

                                                                                                                             

                                                                                                                            run_test

                                                                                                                            link to calibration test

                                                                                                                             

                                                                                                                             

                                                                                                                            cal_filename

                                                                                                                            calibration file

                                                                                                                             

                                                                                                                             

                                                                                                                            method_filename

                                                                                                                             

                                                                                                                             

                                                                                                                             

                                                                                                                            run_filename

                                                                                                                             

                                                                                                                             


                                                                                                                            Anchor
                                                                                                                            RTF340033003900350038003a00
                                                                                                                            RTF340033003900350038003a00
                                                                                                                            Health, Safety, & Environment
                                                                                                                            Safety
                                                                                                                            The list of TOC operation safety rules, issued by general chemistry laboratory safety, oven safety, handling of concentrated H3PO4 and automatic machine operation safeties as well, should be as following:

                                                                                                                            • Handle TOC oven with caution and carefully watch all warning and safety signs around and inside instrument.
                                                                                                                            • Carefully read all three Shimatzu TOC/ASI/Software manuals, which are provided both in hard and soft copies about issues related to personal and instrument safety.
                                                                                                                            • Prepare 25% phosphoric acid from concentrated phosphoric acid under while wearing acid gloves, closed-toed shoes, protective clothing, and safety glasses.
                                                                                                                            • Label interim and stock solutions with the chemical name, concentration, operator name, and date.
                                                                                                                            • Ensure proper operation with gas tanks:
                                                                                                                            • Prevent the tank from dropping by tightening it with strong belt to the stable support
                                                                                                                            • Never attempt to disassemble the high pressure regulator on the top of the tank
                                                                                                                            • Do not disassemble gas tubing until inside pressure drops to close to atmospheric pressure; uptight fitting slightly and let pressure release from gas lines before further operations.
                                                                                                                            • Do not place your fingers or other parts of your body under autosampler needle working area, especially during autosampler operation.
                                                                                                                            • Follow the main safety rules for operation with electronics:
                                                                                                                            • check proper cable setting
                                                                                                                            • Avoid contact with electronic parts inside instrument during "power on" condition

                                                                                                                            Pollution Prevention
                                                                                                                            Waste Management

                                                                                                                            • The contents of the waste bottle (when it's full) need to be checked for neutral pH, corrected if necessary to neutral point with sodium carbonate, and disposed to the sink.

                                                                                                                            Anchor
                                                                                                                            RTF370038003600330039003a00
                                                                                                                            RTF370038003600330039003a00
                                                                                                                            Maintenance & Troubleshooting
                                                                                                                            Scheduled Maintenance
                                                                                                                            Daily

                                                                                                                            • Check water level in humidifier and dehumidifier pot
                                                                                                                            • Check acid level in IC reaction vessel
                                                                                                                            • Inspect drain line positions and empty drain jar, if necessary
                                                                                                                            • Refill autosampler rinse bottle with distilled water
                                                                                                                            • Keep a lint-free environment when cleaning; even small sized particles can clog the sample needle, tubing, or 4-port valve!
                                                                                                                            • Use only distilled water for cleaning procedures--no other solvents. Wipe surfaces with a water-dampened cloth.
                                                                                                                            • Always make sure the oven has cooled down (~30 min) before the TOC ventilators go off.
                                                                                                                            • Proper care must be taken at all times when the TOC is switched off. Usually the TOC combustion oven is working when instrument power with ventilator is going down.
                                                                                                                            • In case of an emergency power black out, remove the TOC cover panels to let heat escape from the oven without significant warming-up of electronic and plastic TOC parts.

                                                                                                                            Weekly/Monthly

                                                                                                                            • Check the gas lines for leaks by comparing flow-meter readings at incoming and outgoing (L-line for CO2 absorber) tubing connections.
                                                                                                                            • Note: All connections in the TOC are finger-tight, so NO TOOLS should be applied to avoid improper pressure!

                                                                                                                            Annual

                                                                                                                            • Replace CO2 absorber (PN 630-00999).

                                                                                                                            When Needed

                                                                                                                            • Replace halogen scrubber (PN 630-00992) when it turns blue, green, or black. After replacement, wait about an hour for baseline to stabilize.
                                                                                                                            • Replace or regenerate high-sensitivity TC catalyst when sensitivity and repeatablility become unacceptable.
                                                                                                                            • Wash TC combustion tube interior with 1+2 HCl solution, rinse in water, and dry when washing or replacing catalyst.
                                                                                                                            • Replace TC combustion tube when it becomes broken or cracked or leaks gas. Regardless of the color or transparency of the glass tube, it can be used until it begins to leak.

                                                                                                                            Hardware Troubleshooting
                                                                                                                            Unstable Baseline

                                                                                                                            • Leak in carrier gas
                                                                                                                            • TC catalyst needs to be replaced
                                                                                                                            • Dehumidifier not operating correctly
                                                                                                                            • Environmental: vitration, ambient temperature, power supply
                                                                                                                            • NDIR defective

                                                                                                                            Bubbles in Sample/Syringe Tubing
                                                                                                                            Bubbles in the sample tubing affect the accuracy of an analysis. If bubbles are noted in the sample tubing, they need to be removed before beginning an analysis.

                                                                                                                            Step

                                                                                                                            Action

                                                                                                                              Select Options > Maintenance > Mechanical Check > TOC.

                                                                                                                                Move the syringe plunger up and down using the plunger height control on the left while changing the position of the 4-port valve to Sample, TC port, or IC port.

                                                                                                                                  When the sample goes to the TC/IC waste line, do not bring the syringe plunger all the way up.

                                                                                                                                  Peaks Are Too Broad or Are Tailing
                                                                                                                                  Tailing should be considered a significant issue. Tailing is identifed by a T in the real-time window in the Notes column and means the integration was cut before the peaks curve reached the baseline. Tailing peaks can be cause by

                                                                                                                                  • Contaminated tubing: check a blank analysis for a contamination peak
                                                                                                                                  • Clogged or leaking tubing:
                                                                                                                                  • Leaking and restricted tubing affect the outlet (CO2 absorber L-line) flow rate. Check both the flowmeter inlet and outlet openings because the TOC and laboratory flow meters may be different. Lower flow at the outlet suggests and clog or leak that must be isolated and fixed.
                                                                                                                                  • Connect both L and S tubes of the CO2 absorber in one line. In this configuration, the operator should see immediate bubbling in the dehumidifier drain container, meaning the line has no major leak and all gas is escaping through the drain. If no bubbling occurs that means the system has a leak or is clogged. The most common places to check for the flow rate inside the instrument are the connections before the combustion tube and the cooling coil outlet.

                                                                                                                                  No Peaks

                                                                                                                                  • The most common reason for an absent peak is a TOC flow line leak. Check the air supply line/carrier gas flow reading and make sure the humidifier stopper is tightly plugged.

                                                                                                                                  TOC/ASI Normal Analysis Routine Stopped in Middle of Run without Operator Command
                                                                                                                                  This may happen because of a hardware malfunction (e.g., bent needle or 4-port valve sticking.

                                                                                                                                  • If the needle is bent, unscrew it and remove it from the autosampler arm. Usually the needle can be straightened. If not, then replace the needle. When replacing the needle in the autosampler arm, ensure the needle height and alignment above the rinse/sample vials are positioned correctly before tightening down the arm.
                                                                                                                                  • If the 4-port valve is stuck, take it apart and check for breakage or particle contamination, as those can cause rotation failure.

                                                                                                                                  Autosampler Troubleshooting
                                                                                                                                  Regenerating TC Catalyst

                                                                                                                                  Step

                                                                                                                                  Action

                                                                                                                                    Place a vial containing 2 mM HCl in position S1.

                                                                                                                                      Open the autosampler Maintenance screen and select Regeneration of TC Catalyst.

                                                                                                                                        The sampling needle moves to the S1 vial then to the rinse water receptacle and rinses the injection needle 3 times.

                                                                                                                                        Replacing Rinse Water Pump Head

                                                                                                                                        Step

                                                                                                                                        Action

                                                                                                                                          After ~300 hr running time, replace the pump head assembly (PN 042-00405-11).

                                                                                                                                            Take off the small cover attached to the side of the autosampler.

                                                                                                                                              Pull out the pump head (white plastic) from the pump body (black plastic) by picking the claws of both sides with fingers.

                                                                                                                                                Replace the pump head by disconnecting the elastic tubes from the Teflon tubes and connecting new ones.

                                                                                                                                                  Remount the pump head and cover.

                                                                                                                                                  Removing the Syringe

                                                                                                                                                  Step

                                                                                                                                                  Action

                                                                                                                                                    Unscrew the plastic plunger holder on the bottom of the syringe (may use pliers).

                                                                                                                                                      Unscrew the syringe body from the 4-port valve stand.


                                                                                                                                                      Shimadzu Technical Support

                                                                                                                                                      • 1-800-477-1227, ext 1442
                                                                                                                                                      • Leonardo Ponds

                                                                                                                                                      Anchor
                                                                                                                                                      RTF53006f006600740077006100
                                                                                                                                                      RTF53006f006600740077006100
                                                                                                                                                      Software Troubleshooting
                                                                                                                                                      Toolbar options don't respond properly

                                                                                                                                                      • Make sure previous windows are closed; software version is old and can operate only with one window open in the TOC Control screen
                                                                                                                                                      • Close/reopen TOC Control window
                                                                                                                                                      • Restart software/computer

                                                                                                                                                      Gas Generator Troubleshooting

                                                                                                                                                      • Normally the Gas Generator should be connected to the TOC through blue plastic tubing. In this case, the outlet flow rate of the Gas Generator is controlled by the TOC flow controller (150 mL/min). If the Gas Generator outlet flow line is disconnected and open to the atmosphere, the flow rate will vary, depending on the inlet pressure. When outlet flow is higher than 1250 mL/min, the Gas Generator front panel overflow indication light will turn yellow. In this situation, the technician should correct the flow rate, because overflow conditions will cause the Gas Generator catalyst module to cool and will not produce appropriate quality air.
                                                                                                                                                      • The outlet flow reading may drop below 150 mL/min or have no flow at all. This will significantly affect the TOC flow rate. The most common reason for such a malfunction is the final filter or other filters are clogged. Replace the filters to fix this problem.
                                                                                                                                                      • If the CHECK SYSTEM light located on the bottom of the front panel starts flashing, and especially when it turns to a permanent yellow light, the catalyst unit needs to be exchanged for a new one. Normally happens once every 2–3 yr of constant Gas Generator use.
                                                                                                                                                      • Every 4–5 min a pneumatic switch triggers audibly between the dryer and co-remover towers on the back of the Gas Generator. If the interval time between switches is ~10 min, there may be problems with the Gas Generator.

                                                                                                                                                      Anchor
                                                                                                                                                      RTF310037003200300034003a00
                                                                                                                                                      RTF310037003200300034003a00
                                                                                                                                                      Software Guide
                                                                                                                                                      TOC-Control Screen Toolbar Functions

                                                                                                                                                      • New File: Closes the current Sample Table and brings up a blank Sample Table.
                                                                                                                                                      • Open File: Opens a saved Sample Table, which replaces the current Sample Table. Select a .toc file in the dialog box.
                                                                                                                                                      • Save File: Saves the data in the current Sample Table.
                                                                                                                                                      • Cut: Removes data from highlighted cells in a table and saves them to the clipboard.
                                                                                                                                                      • Copy: Copies data from highlighted cells in a table and saves them to the clipboard.
                                                                                                                                                      • Paste: Copies the contents of the clipboard to a selected location in a table.
                                                                                                                                                      • Print: Sends data from currently displayed window to the printer.
                                                                                                                                                      • Run Time Report: Toggles Run Time Report on and off.
                                                                                                                                                      • Configure by selecting Report > Run Time Report.
                                                                                                                                                      • Recalculate: Recalculates the currently displayed sample run after applying a new calibration curve or excluding injections from the Injection Table.
                                                                                                                                                      • Show Injection Table: Displays the Injection Table.
                                                                                                                                                      • Highlight a row in a Sample Table to indicate which sample run to show in the Injection Table.
                                                                                                                                                      • Show Peak Profile: Displays a Peak Profile graph.
                                                                                                                                                      • Highlight a row in a Sample Table to indicate which sample run to display.
                                                                                                                                                      • To display a peak profile for an individual injection, highlight a row in the Injection Table.
                                                                                                                                                      • Show Calibration Curve: Displays calibration curve data for a specific sample run.
                                                                                                                                                      • Highlight a row in the Sample Table to indicate which calibration curve to display.
                                                                                                                                                      • Exclude: Excludes the selected sample or injection from a calculation or re-includes a previously excluded sample.
                                                                                                                                                      • Real Time Window: Opens the Real Time analysis window.
                                                                                                                                                      • Connect/Disconnect Instrument: Opens or closes the elctronic connection between the instrument and the computer.

                                                                                                                                                      Instrument Status

                                                                                                                                                      Message

                                                                                                                                                      Procedure

                                                                                                                                                      Ready

                                                                                                                                                      Ready to receive command

                                                                                                                                                      Measuring

                                                                                                                                                      Measurement proceeding

                                                                                                                                                      Waiting for Ready State

                                                                                                                                                      Waiting: ready status condition

                                                                                                                                                      Washing syringe with acid

                                                                                                                                                      Aspirating acid

                                                                                                                                                      Injecting acid

                                                                                                                                                      Injecting acid

                                                                                                                                                      Washing syringe with rinse water

                                                                                                                                                      Rinsing syringe

                                                                                                                                                      ASI moving

                                                                                                                                                      Initializing and checking tray type

                                                                                                                                                      ASI pause

                                                                                                                                                      ASI pausing

                                                                                                                                                      Washing ultra pure water trap

                                                                                                                                                      Rinsing ultra pure water trap

                                                                                                                                                      Making ultra pure water

                                                                                                                                                      Generating ultra pure water

                                                                                                                                                      Measuring for blank check

                                                                                                                                                      Measuring for blank check

                                                                                                                                                      Processing for end of blank check

                                                                                                                                                      Processing for end of blank check

                                                                                                                                                      Searching for zero point

                                                                                                                                                      Searching for syringe home position

                                                                                                                                                      Regenerating

                                                                                                                                                      Regenerating catalyst

                                                                                                                                                      Residue Removal

                                                                                                                                                      Removing residue from catalyst

                                                                                                                                                      Flow line wash

                                                                                                                                                      Rinsing flow lines

                                                                                                                                                      Mechanical Check

                                                                                                                                                      Mechanical check processing

                                                                                                                                                      Monitoring

                                                                                                                                                      Background data being sent

                                                                                                                                                      Standby

                                                                                                                                                      Standby option selected; processing

                                                                                                                                                      Shutdown

                                                                                                                                                      Shutdown in processing

                                                                                                                                                      Initializing

                                                                                                                                                      Intializing entire instrument

                                                                                                                                                      TOC mech. initializing

                                                                                                                                                      Mechanical initializing

                                                                                                                                                      Moving

                                                                                                                                                      Syringe moving

                                                                                                                                                      Analysis Status

                                                                                                                                                      Message

                                                                                                                                                      Procedure

                                                                                                                                                      Sparging

                                                                                                                                                      ASI sparging on

                                                                                                                                                      Washing

                                                                                                                                                      Rinsing sample syringe

                                                                                                                                                      Sampling

                                                                                                                                                      Aspirating sample into syringe

                                                                                                                                                      Injecting

                                                                                                                                                      Injecting sample into furnace or reaction vessel

                                                                                                                                                      Measuring

                                                                                                                                                      Integrating peak profile data

                                                                                                                                                      Draining

                                                                                                                                                      Draining the sample

                                                                                                                                                      Waiting: Interval setting

                                                                                                                                                      Waiting for interval setting

                                                                                                                                                      Waiting: Ready status

                                                                                                                                                      Waiting: ready status is NG (no good)

                                                                                                                                                      Waiting: POC

                                                                                                                                                      Waiting for POC bubbler

                                                                                                                                                      Waiting: ESU

                                                                                                                                                      Waiting for ESU

                                                                                                                                                      Waiting: Other

                                                                                                                                                      Waiting for other

                                                                                                                                                      ASI moving

                                                                                                                                                      ASI moving

                                                                                                                                                      IC regenerating

                                                                                                                                                      Automatic regeneration of IC fluid

                                                                                                                                                      Bubbling

                                                                                                                                                      Bubbling (POC bubbler)

                                                                                                                                                      Instrument Status Bar
                                                                                                                                                      Instrument Control Bar Buttons

                                                                                                                                                      • Start: starts measurement
                                                                                                                                                      • Stop: stops the measurement after the current injection is complete
                                                                                                                                                      • Pause: pauses instrument operation after current injection is complete
                                                                                                                                                      • Halt: immediately interrupts measurement
                                                                                                                                                      • Continue: continues interrupted measurement sequence
                                                                                                                                                      • Peak Stop: stops current measurement