GC3-Natural Gas Analysis User Guide: Drilling Safety Monitoring
Manual Information
Author(s): | C. Bennight |
Reviewer(s): | L. Brandt, C. Neal, K. Marsaglia |
Editor(s): | K. Graber, L. Peters |
Management Approval (Name, Title, Date): | D.J. Houpt, Supervisor of Analytical Systems, 9/24/2010 |
Audience: | Scientists, Laboratory Technicians |
Origination date: | 5/12/2008 |
Current version: | Version 1.0 9/24/2010 |
Revised: |
|
Domain: | Chemistry |
System: | Gas Chromatography |
Keywords: | Hydrocarbons, Natural Gas, Headspace |
Changes to User Guide
Summarize requested modifications to this user guide in an e-mail and/or annotate the PDF file and e-mail change requests to techdoc@iodp.tamu.edu.
User Guide Contents
Topic | See page… |
Introduction | |
Apparatus, Reagents, & Materials | |
Instrument Calibration/Calibration Verification | |
Sample Preparation & Analysis | |
Quality Assurance/Quality Control | |
LIMS Integration | |
Health, Safety, & Environment | |
Maintenance & Troubleshooting (HP6890GC) |
...
...
...
Introduction
Overview
Natural gas analysis for hydrocarbons and hydrogen sulfide (H2S) is required to avoid natural gas and oil escaping from the hole and is part of the ship's standard drilling safety plan.
The absolute quantity of hydrocarbons is the primary safety risk during shipboard operations. Gas monitoring via gas chromatography is a means of quantifying the hydrocarbon risk. H2S is another significant risk factor for individuals working in the area. Emergency monitors on the drill floor provide early detection of H2S, while later quantification is performed on the natural gas analyzer (NGA).
...
Sulfate-reducing bacteria produce hydrogen sulfide in euxinic sediments. This may occur in a relatively shallow part of the sediment. Thermochemical sulfate reduction of sulfate by hydrocarbons in reservoirs occurs under high temperature (>127°C ~ 140°C).
Theory of Method
Two instruments monitor gases in core headspace and core void samples:
...
The TCD flow path travels through a 6 ft x 2.0 mm ID stainless steel (SS) column packed with Poropak T (50/80 mesh), a 3 ft x 2.0 mm ID SS column packed with molecular sieve 13x (60/80 mesh), and 6 ft x 2.0 mm ID SS column packed with 80/100 mesh HayeSep R (acid washed).
The FID flow path traverses a 60 m x 0.25 mm ID capillary column with 0.25 µm DB-1 film.
Anchor | ||||
---|---|---|---|---|
|
...
Instrument Calibration/Calibration Verification
Overview
Before unknown samples can be analyzed for headspace gases, each GC system must have a valid calibration curve and the calibration curve must have been verified using a calibration verification standard.
...
1 | Create/refresh calibration curve (start at least 1 day before reaching site) (see Creating a Calibration Curve). |
2 | Verify calibration (Running a Calibration Verification Standard). |
3 | Perform a work flow test (Running a Gas Sample). |
...
...
Creating a Calibration Curve
...
1 | Prepare 5–7 registered standard gases. |
2 | Activate GC3/NGA LIMS uploader located at Start > Program Files > IODP > MegaUploadaTron. The uploader must be activated before the calibration is run. |
3 | In the ChemStation Main menu, click Run Control > Sample Info. |
4 | Fill in the specific fields on the screen as follows:
|
5 | Slowly inject 5000 µL of the first standard gas and observe the floating ball in the flow meter move upward. |
6 | When the ball in the flow meter indicates flow has fallen to just above 0 (is about to hit 0), press the Start button on the control panel of the GC. |
7 | When the run has finished, open the Data Analysis screen in ChemStation and click Calibration. |
8 | On the Main ChemStation menu, select Calibration > Recalibrate. |
9 | On the Recalibration screen, select Level # and Replace (or Average) as applicable for that level. |
10 | Repeat Steps 5–9 for 3 replicate standards (CH4: A 25%, B = 50%, C 75%, D = 99%). |
11 | Click OK to change the calibration value. For NGA calibration, the same standard can be applied to both the appropriate TCD and FID level; you do not need separate standards for |
...
TCD and FID. |
Running a Calibration Verification Standard
1 | Ensure the uploader is activated and the CV standard is registered in LIMS. |
2 | Click Run Control in the main menu of ChemStation and select Sample Info. |
3 | Fill in the specific section on the window as follows:
|
4 | Prepare the CV standard at approximately the mid-point concentration of the curve. |
5 | Slowly inject 5000 µL of the standard gas, keeping the outflow rate <80 mL/min. |
6 | Press Start on the GC control panel when the flow meter is just above 0. |
7 | When the run is finished, the report will automatically display the values. Click Upload in the uploader to submit the data to LIMS. |
...
...
Running a Blank
1 | To run a blank, in the Main menu click RunControl > Sample Info. |
2 | Fill in the following fields:
|
3 | Prepare laboratory air (5000 µL) and inject it into the GC in the same fashion as the standards above when the ChemStation software shows Ready. |
4 | Press the Start button on the GC control panel to start the run. |
5 | Confirm the chromatogram on the screen shows no peaks. If peaks are present, the system contamination must be found (injector, detector, sample loop, etc.). |
...
...
...
Running a Gas Sample
1 | Ensure the uploader has been activated. |
2 | Click Run Control in the main menu of ChemStation and select Sample Info. |
3 | Fill in the specific section on the window as follows:
|
4 | Prepare a headspace or void gas sample. |
5 | Slowly inject 5000 µL of the gas sample, keeping the maximum gas outflow <80 mL/min. |
6 | Press Start on the GC control panel when the ball on the flow meter is just above 0. |
...
...
Sample Preparation & Analysis
Overview
There are two primary sample types used for natural gas analysis.
...
Occasionally, cores that come on deck have voids with large amounts of free gas. Free gas must be sampled using a sampling device that penetrates the liner and provides a channel for the gas to be drawn into a gas-tight syringe, vacuum vial, or gas sampling bag.
Sampling Tools
- Sample coring tool (metal cylinder)
- Sample coring plunger
- Puncture tool (to penetrate plastic liner)
- Headspace vial
- Headspace gasket with crimp top
- Crimping tool
- Permanent marker for labeling
...
Collecting a Headspace Gas Sample
1 | Locate a freshly sectioned core (consult with the curator). |
2 | Gently push the sample coring tool into the core section slightly inward of the edge. |
3 | Gently pull out the tool. If the sample recovery (% of coring tool with sample) is >80% (~5–7 cm3), proceed; otherwise repeat Steps 1 and 2. |
4 | Place the open end of the sample coring tool over a clean headspace gas vial and use the plunger to push the sediment into the vial. |
5 | Immediately place a gasket with a crimp top over the vial and crimp shut. |
6 | After sealing the vial, immediately write down the sampling interval, location, and any other information for the sample that was just taken. Generate a proper label and apply it to the vial as soon as possible. |
7 | Place the vial with the sample in a 70°C oven for 30 min to degas the sediment (use timer). |
8 | Inject extracted gas sample into the GC using syringe (see |
Collecting a Void Gas Sample
1 | Use the puncture tool to make a hole in the core liner to make a channel for the gas. |
2 | Quickly collect a free gas sample from the small hole with a syringe. |
3 | Immediately introduce the gas sample into the GC instruments in the same manner as the headspace samples. |
...
...
Running a Sample
...
1 | Start GC and operation system at least 1 day before reaching site (the system should be fully calibrated and ready for analysis) (see Advanced User Guide). |
2 | Ensure LIMS uploader is running. |
3 | Inject 5 mL of headspace gas after the sample has heated in the oven for 30 |
4
Click Upload if the uploader is not in automatic mode.
...
min. | |
4 | Click Upload if the uploader is not in automatic mode. |
Quality Assurance/Quality Control
Overview
QA/QC for GC3/NGA analysis consists of instrument calibration and continuing calibration verification using check standards, instrument blanks, and replicate samples.
...
Analytical Batch
An analytical batch is a method-defined number of samples with which QC samples including calibration verification, blank, and replicate samples are run. Samples are implicitly grouped into batches based on the spacing between CV samples.
...
...
QC Samples
...
...
Blank
- The blank determines the level of contamination originating from the laboratory environment (air) and sample path in the GC (injection port with screen, sample loop, and separation column).
- Run a blank with each batch of samples by injecting 5 mL of ambient laboratory air into the GC using the same syringe used to inject headspace gas samples.
- All calibrated values other than O2 and N2 should be nondetectable in the blank. If aberrant peaks appear, bake the column for 8 hr and repeat the blank analysis.
...
- Five to seven levels of calibration samples (standard gases) are used to create a calibration curve, which is saved with the measurement data (see Instrument Calibration/Calibration Verification).
- Correlation coefficient values for calibration curves should be 0.99 or better, except O2 and N2, which should be 0.95 or better.
...
...
Calibration Verification (CV) Sample
- Select one of the 5–7 calibration samples from the calibration curve for the calibration verification sample.
- Run a CV sample at least every shift that samples are taken (see Instrument Calibration/Calibration Verification).
- The CV should fall within 3% of the calibrated value; O2 and N2 should be within 10% of the calibrated value.
...
Control Limits
For a system to be considered in control, all QA/QC samples (blank and calibration verification) must be in control.
...
A QA/QC sample is in control when the sample analysis result is within a certain tolerance of acceptable limits (see above). Calibration verification samples should be within acceptable limits of the actual value calculated against the calibration curve (see Calibration Verification (CV) Sample) and blanks should be within acceptable limits of background levels of headspace hydrocarbons and gases (see Blank). When the system is in control, as indicated by acceptable results on QA/QC samples, analytical results for unknown samples are considered to be reliable.
...
If the control limits are exceeded, the instrument system is considered out of control and all samples in the current analytical batch are invalid and must be rerun after the system is proved to be in control.
...
LIMS Integration
LIMS Components
Analysis | Component | Unit | Description |
---|---|---|---|
GC3 | dat_asman_id | — | Serial number of chromatographic data file in digital asset management database (ASMAN) |
| dat_filename | — | File name of chromatographic data file containing measurements |
| run_test | — | Test number of related calibration or QA/QC test |
| propene | ppmv | Relative concentration of propene in the sample |
| propane | ppmv | Relative concentration of propane in the sample |
| ethene | ppmv | Relative concentration of ethene in the sample |
| ethane | ppmv | Relative concentration of ethane in the sample |
| methane | ppmv | Relative concentration of methane in the sample |
GC3_QAQC | dat_asman_id | — | Serial number of chromatographic data file in ASMAN |
| dat_filename | — | File name of chromatographic data file containing measurements |
| run_test | — | Test number of related calibration or QA/QC test |
| propene | ppmv | Relative concentration of propene in the sample |
| propane | ppmv | Relative concentration of propane in the sample |
| ethene | ppmv | Relative concentration of ethene in the sample |
| ethane | ppmv | Relative concentration of ethane in the sample |
| methane | ppmv | Relative concentration of methane in the sample |
GC3_QCAL | mtd_asman_id | — | Serial number of chromatographic method in ASMAN |
| mtd_filename | — | File name of the chromatographic method file containing measurements |
| ethene_corr2 | R2 | Ethene calibration coefficient |
| ethene_intercept | — | Intercept of ethene calibration curve |
| ethene_slope | — | Slope of ethene calibration curve |
| ethane_corr2 | R2 | Ethane calibration coefficient |
| ethane_intercept | — | Intercept of ethane calibration curve |
| ethane_slope | — | Slope of ethane calibration curve |
| propene_corr2 | R2 | Propene calibration coefficient |
| propene_intercept | — | Intercept of propene calibration curve |
| propene_slope | — | Slope of propene calibration curve |
| propane_corr2 | R2 | Propane calibration coefficient |
| propane_intercept | — | Intercept of propane calibration curve |
| propane_slope | — | Slope of propene calibration curve |
| methane_corr2 | R2 | Methane calibration coefficient |
| methane_intercept | — | Intercept of methane calibration curve |
| methane_slope | — | Slope of methane calibration curve |
NGAFID | dat_asman_id | — | Serial number of chromatographic data file in ASMAN |
| dat_filename | — | File name of chromatographic data file containing measurements |
| run_test | — | Test number of related calibration or QA/QC test |
| iso_butane | ppmv | Concentration of iso_butane in a sample |
| iso_heptane | ppmv | Concentration of iso_heptane in a sample |
| iso_hexane | ppmv | Concentration of iso_hexane in a sample |
| iso_pentane | ppmv | Concentration of iso_pentane in a sample |
| n_butane | ppmv | Concentration of n_butane in a sample |
| n_heptane | ppmv | Concentration of n_heptane in a sample |
| n_hexane | ppmv | Concentration of n_hexane in a sample |
| n_pentane | ppmv | Concentration of n_pentane in a sample |
| ethane_ethene | ppmv | Concentration of ethane + ethene in a sample |
| propane_propene | ppmv | Concentration of propane + propene in a sample |
| methane | ppmv | Concentration of methane in a sample |
NGAFID_QA | dat_asman_id | — | Serial number of chromatographic data file in ASMAN |
| dat_filename | — | File name of chromatographic data file containing measurements |
| run_test | — | Test number of related calibration or QA/QC test |
| iso_butane | ppmv | Concentration of iso_butane in a sample |
| iso_heptane | ppmv | Concentration of iso_heptane in a sample |
| iso_hexane | ppmv | Concentration of iso_hexane in a sample |
| iso_pentane | ppmv | Concentration of iso_pentane in a sample |
| n_butane | ppmv | Concentration of n_butane in a sample |
| n_heptane | ppmv | Concentration of n_heptane in a sample |
| n_hexane | ppmv | Concentration of n_hexane in a sample |
| n_pentane | ppmv | Concentration of n_pentane in a sample |
| ethane_ethene | ppmv | Concentration of ethane + ethene in a sample |
| propane_propene | ppmv | Concentration of propane + propene in a sample |
| methane | ppmv | Concentration of methane in a sample |
NGAFID_QC | mtd_asman_id | — | Serial number of chromatographic method in ASMAN |
| mtd_filename | — | File name of the chromatographic method file containing measurements |
| iso_butane_corr2 | R2 | Iso-butane calibration coefficient |
| iso_butane_intercept | — | Intercept of iso-butane calibration curve |
| iso_butane_slope | — | Slope of iso-butane calibration curve |
| iso_heptane_corr2 | R2 | Iso-heptane calibration coefficient |
| iso_heptane_intercept | — | Intercept of iso-heptane calibration curve |
| iso_heptane_slope | — | Slope of iso-heptane calibration curve |
| iso_hexane_corr2 | R2 | Iso-hexane calibration coefficient |
| iso_hexane_intercept | — | Intercept of iso-hexane calibration curve |
| iso_hexane_slope | — | Slope of iso-hexane calibration curve |
| iso_pentane_corr2 | R2 | Iso-pentane calibration coefficient |
| iso_pentane_intercept | — | Intercept of iso-pentane calibration curve |
| iso_pentane_slope | — | Slope of iso-pentane calibration curve |
| n_butane_corr2 | R2 | n-butane calibration coefficient |
| n_butane_intercept | — | Intercept of n-butane calibration curve |
| n_butane_slope | — | Slope of n-butane calibration curve |
| n_heptane_corr2 | R2 | n-heptane calibration coefficient |
| n_heptane_intercept | — | Intercept of n-heptane calibration curve |
| n_heptane_slope | — | Slope of n-heptane calibration curve |
| n_hexane_corr2 | R2 | n-hexane calibration coefficient |
| n_hexane_intercept | — | Intercept of n-hexane calibration curve |
| n_hexane_slope | — | Slope of n-hexane calibration curve |
| n_pentane_corr2 | R2 | n-pentane calibration coefficient |
| n_pentane_intercept | — | Intercept of n-pentane calibration curve |
| n_pentane_slope | — | Slope of n-pentane calibration curve |
| ethane_ethene_corr2 | R2 | Ethane + ethene calibration coefficient |
| ethane_ethene_intercept | — | Intercept of ethane + ethene calibration curve |
| ethane_ethene_slope | — | Slope of ethane + ethene calibration curve |
| propane_propene_corr2 | R2 | Propane + propene calibration coefficient |
| propane_propene_intercept | — | Intercept of propane + propene calibration curve |
| propane_propene_slope | — | Slope of propane + propene calibration curve |
NGAFID_QC | methane_corr2 | R2 | Methane calibration coefficient |
| methane_intercept | — | Intercept of methane calibration curve |
| methane_slope | — | Slope of methane calibration curve |
NGATCD | dat_asman_id | — | Serial number of chromatographic data file in ASMAN |
| dat_filename | — | File name of chromatographic data file containing measurements |
| run_test | — | Test number of related calibration or QA/QC test |
| carbon_dioxide | ppmv | Concentration of carbon dioxide in a sample |
| ethane | ppmv | Concentration of ethane in a sample |
| ethene | ppmv | Concentration of ethene in a sample |
| hydrogen_sulfide | ppmv | Concentration of hydrogen sulfide in a sample |
| methane | ppmv | Concentration of methane in a sample |
| nitrogen | ppmv | Concentration of nitrogen in a sample |
| oxygen | ppmv | Concentration of oxygen in a sample |
| propane | ppmv | Concentration of propane in a sample |
| propene | ppmv | Concentration of propene in a sample |
Uploading Data to LIMS
Data are uploaded to LIMS automatically using a process explained in the GC3-NGA Advanced User Guide. If the data do not upload to LIMS, contact the laboratory technician.
...
...
Health, Safety, & Environment
Safety
- The following parts are dangerously hot. Avoid touching these areas and cool completely to room temperature before servicing them:
- Inlets
- Oven
- Detectors
- Column nuts
- Be careful when working behind the instrument; during cooldown cycle the oven emits hot exhaust that can cause burns.
- Do not place temperature-sensitive items (e.g., gas cylinders, chemicals, regulators, and plastic tubing) in the path of the heated exhaust.
- Insulation around inlets, detectors, and valve box contains refractory ceramic fibers. Avoid inhaling particles and wear personal protective equipment including gloves, safety glasses, and dust/mist respirator when working in these areas.
- Do not leave flammable gas flows on if GC will be unmonitored for long periods of time (however, leave carrier gas on for column flow).
- Always operate the instrument with the cover properly installed.
...
Maintenance & Troubleshooting (HP6890GC)
Troubleshooting
Faults
- Beeping instrument (cancel beep by pressing Clear on the instrument keyboard)
- One beep: instrument fault, warning, or shutdown
- Series of beeps: gas flow cannot reach setpoint and flow will be shut down after 1–2 min
- Continuous beep: thermal shutdown
- Blinking setpoint on GC display
- Control table setpoint blinking: gas flow, valve, or oven shutdown
- Detector On/Off line blinking: pneumatics or detector failure
- Instrument screen messages (press Clear to remove message)
- Caution: configuration problems
- Error: setpoint out of range or incorrect hardware
- Popup: shutdown, fault, or warning (see error table)
- FID will not stay lit
- Make sure the dessicant in the hydrogen generator is not saturated with water (replace/recharge as necessary).
- Check water level in hydrogen generator
...