Total Organic Carbon Analyzer User Guide (from TOC_UG_20100429)
This file is for editing content only! Cross-references and hyperlinks do not work. Use track changes and return file to peters@iodp.tamu.edu with copy to techdoc@iodp.tamu.edu.
Manual Information
...
Topic | See page… |
Introduction | 2 |
Apparatus, Reagents, & Materials | 3 |
Instrument Calibration/Verification | 9 |
Sample Preparation | 12 |
Sample Analysis | 17 |
Quality Assurance/Quality Control | 18 |
LIMS Integration | 19 |
Health, Safety, & Environment | 20 |
Maintenance & Troubleshooting | 21 |
Software Guide | 25 |
Introduction
Theory of Method
The Total Organic Carbon (TOC) instrument analyzes liquid samples by physically combusting the samples at 680°C for TOC or by chemically combusting the samples in 25% phosphoric acid for inorganic carbon (IC). The instrument measures the amount of carbon degraded to CO2 using nondispersive infrared gas analysis at 4.3 µm wavelength, calculated based on Lambert-Beer's law.
Total Carbon (TC)
To analyze for Total Carbon (TC) a sample is delivered to the TC combustion tube filled with platinum catalyst and heated to 680°C. Carrier gas (purified air or oxygen) , along with the CO2 generated by combustion, flows through the reaction vessel to the dehumidifier and scrubber, where it is cooled, dried, and purified. It then it passes into the nondispersive infrared detector, which generates an analog signal that is digitized and processed by the TOC Control software. The peak area is proportional to the concentration of total carbon in the sample.
Inorganic Carbon (IC)
During inorganic carbon analysis the liquid sample is introduced into the reaction vessel, which is half filled with 25% phosphoric acid. The sample is purged with carrier gas (high purity air) through the acid solution. The phosphoric acid reacts only with inorganic carbon components (CO32- and HCO3), transforming them to CO2. The CO2 is carried fromthe IC vessel through the dehumidifier and scrubber to the NDIR detector.
Total Organic Carbon
TOC concentration can be obtained by subtracting the IC concentration from the TC concentration.
Note: Before operating the TOC system every operator must get either additional training from a senior technician or carefully study all three manufacturer TOC/ASI/Software manuals.
Apparatus, Reagents, & Materials
Laboratory Apparatus
...
Gas Generator
The Parker Balston (Whatman)78-40 TOC Gas Generator (Figure 1-7)converts compressed air into high purity carrier/combustion gas for the TOC analyzer with hydrocarbon levels < 0.1 ppm and CO2 level < 1 ppm.
The Gas Generator should be fed from the air compressor, not from an air tank, because of the high rate of air consumption. Incoming air should be relatively free of dust and water (typical air compressor produced air should meet these requirements).
...
Figure 1-7. TOC Gas Generator.
Instrument Calibration/Verification
Preparing Calibration Standards
Total Carbon
In a 100 mL volumetric flask, make up the following calibration levels to bracket general expected TC levels from the 100 ppm intermediate stock solution:
- 100 ppb: 0.1 mL of 100 ppm + 99.9 mL zero water
- 500 ppb: 0.5 mL of 100 ppm + 99.5 mL zero water
- 1 ppm: 1.0 mL of 100 ppm + 99 mL zero water
- 2 ppm: 2.0 mL of 100 ppm + 98 mL zero water
- 5 ppm: 5.0 mL of 100 ppm + 95 mL zero water
...
Step | Action |
| Run a midpoint curve standard as an "unknown" sample in the Sample Table. |
| Apply the calibration curve to calculate the sample results. |
| Open View > Injection Table or click the Injections button on the toolbar to examine the sample results in the Concentration column and the result of the Outlier Test. For acceptable calibration verification, CV should be >0.90. |
| Close the Injection Table screen. |
Sample Preparation
Overview
Samples are prepared by squeezing interstitial waters from sediment cores. Total organic carbon is run on unfiltered interstitial water samples; dissolved organic carbon (DOC) is run on interstitial water samples filtered through a 0.45 µm membrane.
Interstitial waters are diluted ~1:5 before running on the TOC analyzer and the analyzer is calibrated for quantitation of sample results. Running samples involves the following steps:
...
Figure 1-11. Software TOC Control Screen with Blank Sample Table.
Preparing TOC Software
Step | Action |
| Click on the TOC Control icon on the computer desktop (see #1 on Fig 11). Type DAQ for username and password to open the TOC Control screen. |
| From the Toolbar on the TOC Control screen, select Measure > Connect (#3 on Fig 11), and at the bottom of the Control screen click the TOC button (#4 on Fig 11). |
| Check the carrier gas flow meter and pressure gage on the front panel of the TOC analyzer. [Author: Need a photo of this meter and pressure gage.]
|
4 | On the TOC Control screen, select View > Background Monitor (#6 on Fig 11) to open a background window. |
5 | Wait ~2 min and then make sure all parameters read OK on screen (exception: oven temperature will take ~30 min to reach 680°C). [Author: Need a screen shot of the Background window.]
|
6 | On the TOC Control screen select Options > Instrument Conditions (#7 on Fig 1) to set up and/or verify TOC, ASI, and RS232 settings as follows: TOC
ASI
RS232
|
Preparing the Autosampler
Step | Action |
| Fill autosampler rinse bottle (Fig 5) with fresh zero water. |
2 | Fill high-sensitivity 50 mL autosampler glass vial with zero water and place it to S1 tray position. Remove tray cover or make sure it's in the correct position.
|
| Remove bubbles from the syringe (Detect Zero point): on the TOC Control screen select Options > Maintenance > Zero point detection. |
Setting up a Sample Table and Selecting a Calibration Curve
Step | Action |
| From the TOC Control screen select File > Open Sample Table to create a new Sample Table (#8 in Fig 11) or open an old Sample Table (#9 in Fig 8). |
| Fill in the Sample Table with sample type, name, ID, and analysis. |
| To specify a method on the Sample Table, select View > Methods on the TOC Control screen to open a saved method or create a new method. |
| Fill in information on the General and Method tabs of the Method screen (Fig 12). |
| To specify a calibration curve to apply to the samples, select View > Calibration on the TOC Control screen to open a dialog box showing a list of calibration curves saved in the CALIBR folder (Fig 9). |
| On the Calibration Curve List dialog box, click Open to select a saved calibration curve. |
| Save the Sample Table. |
...
Figure 1-12. IC Method Tabs.
Sample Analysis
Running Samples
Step | Action |
| Load the sample vials into the autosampler in positions that correspond to the Sample Table. |
| Verify status of instrument and baseline: View > Background Monitor, then close Background Monitor screen. |
| Select Measure > Start or click Start on the instrument control bar. When the Standby screen opens, select Keep Running and click OK. |
| To monitor the status of the measurements, select View > Realtime Window or select Realtime Window on the toolbar. |
| When the sample runs are complete the instrument status bar will indicate Ready. |
| Sample results calculated from the loaded calibration file will display in the Sample Table Mean Area and Result columns. |
Quality Assurance/Quality Control
Analytical Batch
Blanks
The TOC analyzer has an automated TC Blank Check Program that runs zero water blanks from 4 different vials a total of 10 times.
...
Step | Action |
| Place vials filled with zero water in autosampler positions S1–S4. |
| On the autosampler control panel, open the Maintenance screen and select TC Blank Check. |
| The autosampler injects S1 twice, S2 three times, S3 three times, and S4 twice and then returns to Home. |
Calibration
Calibration Verification
Precision
Accuracy
LIMS Integration
Overview
TC/IC/TOC analysis results are manually entered into the LabWare LIMS client by the technical staff once the scientist(s) is (are) finished reducing the data.
LIMS Components
Results are stored in the LIMS database associated with an analysis code and an analysis component. Analysis codes and their components, definitions, and units are listed below.
...
Analysis | Component | Definition | Unit |
TOC | carbon_total | Concentration | ppm |
| inorganic_carbon_total | Concentration | ppm |
| organic_carbon_total | Concentration | ppm |
| run_test | link to calibration test |
|
TOC_QAQC | carbon_total | Concentration | ppm |
| inorganic_carbon_total | Concentration | ppm |
| organic_carbon_total | Concentration | ppm |
| run_test | link to calibration test |
|
| cal_filename | calibration file |
|
| method_filename |
|
|
| run_filename |
|
|
Health, Safety, & Environment
Safety
The list of TOC operation safety rules, issued by general chemistry laboratory safety, oven safety, handling of concentrated H3PO4 and automatic machine operation safeties as well, should be as following:
...
- The contents of the waste bottle (when it's full) need to be checked for neutral pH, corrected if necessary to neutral point with sodium carbonate, and disposed to the sink.
Maintenance & Troubleshooting
Scheduled Maintenance
Daily
...
- Replace halogen scrubber (PN 630-00992) when it turns blue, green, or black. After replacement, wait about an hour for baseline to stabilize.
- Replace or regenerate high-sensitivity TC catalyst when sensitivity and repeatablility become unacceptable.
- Wash TC combustion tube interior with 1+2 HCl solution, rinse in water, and dry when washing or replacing catalyst.
- Replace TC combustion tube when it becomes broken or cracked or leaks gas. Regardless of the color or transparency of the glass tube, it can be used until it begins to leak.
Hardware Troubleshooting
Unstable Baseline
- Leak in carrier gas
- TC catalyst needs to be replaced
- Dehumidifier not operating correctly
- Environmental: vitration, ambient temperature, power supply
- NDIR defective
...
- If the needle is bent, unscrew it and remove it from the autosampler arm. Usually the needle can be straightened. If not, then replace the needle. When replacing the needle in the autosampler arm, ensure the needle height and alignment above the rinse/sample vials are positioned correctly before tightening down the arm.
- If the 4-port valve is stuck, take it apart and check for breakage or particle contamination, as those can cause rotation failure.
Autosampler Troubleshooting
Regenerating TC Catalyst
Step | Action |
| Place a vial containing 2 mM HCl in position S1. |
| Open the autosampler Maintenance screen and select Regeneration of TC Catalyst. |
| The sampling needle moves to the S1 vial then to the rinse water receptacle and rinses the injection needle 3 times. |
...
- 1-800-477-1227, ext 1442
- Leonardo Ponds
Software Troubleshooting
Toolbar options don't respond properly
...
- Normally the Gas Generator should be connected to the TOC through blue plastic tubing. In this case, the outlet flow rate of the Gas Generator is controlled by the TOC flow controller (150 mL/min). If the Gas Generator outlet flow line is disconnected and open to the atmosphere, the flow rate will vary, depending on the inlet pressure. When outlet flow is higher than 1250 mL/min, the Gas Generator front panel overflow indication light will turn yellow. In this situation, the technician should correct the flow rate, because overflow conditions will cause the Gas Generator catalyst module to cool and will not produce appropriate quality air.
- The outlet flow reading may drop below 150 mL/min or have no flow at all. This will significantly affect the TOC flow rate. The most common reason for such a malfunction is the final filter or other filters are clogged. Replace the filters to fix this problem.
- If the CHECK SYSTEM light located on the bottom of the front panel starts flashing, and especially when it turns to a permanent yellow light, the catalyst unit needs to be exchanged for a new one. Normally happens once every 2–3 yr of constant Gas Generator use.
- Every 4–5 min a pneumatic switch triggers audibly between the dryer and co-remover towers on the back of the Gas Generator. If the interval time between switches is ~10 min, there may be problems with the Gas Generator.
Software Guide
TOC-Control Screen Toolbar Functions
...
Message | Procedure |
Sparging | ASI sparging on |
Washing | Rinsing sample syringe |
Sampling | Aspirating sample into syringe |
Injecting | Injecting sample into furnace or reaction vessel |
Measuring | Integrating peak profile data |
Draining | Draining the sample |
Waiting: Interval setting | Waiting for interval setting |
Waiting: Ready status | Waiting: ready status is NG (no good) |
Waiting: POC | Waiting for POC bubbler |
Waiting: ESU | Waiting for ESU |
Waiting: Other | Waiting for other |
ASI moving | ASI moving |
IC regenerating | Automatic regeneration of IC fluid |
Bubbling | Bubbling (POC bubbler) |
Instrument Status Bar
Instrument Control Bar Buttons
...