  1. Select NEW SPECIMEN and a new specimen window will open.
    1. Scan or enter the sample name
    2. Enter the volume of the sample in cubic centimeters. Plastic cubes with rounded corners are 7cc, hard rock cubes are 8cc, cylinders are generally 10cc.
    3. You have the option to add a note
    4. Enter the number of measurement repetitions in the Auto Repeat box
    5. Select OK
      Anchor
      _GoBack
      _GoBack
  2. Insert the sample into the specimen holder in the desired orientation.
  3. Select START
    1. The instrument will first take a measurement of the empty coils to zero the instrument.
    2. If the up/down manipulator is enabled the holder will automatically be moved into the pick up coil
    3. If the up/down manipulator is disabled, wait for a long beep then insert the sample into the pick up coil until it rests on the white plastic cylinder. When you hear a short beep indicating that the measurement is complete, remove the sample from the pickup coil.
  4. Select SAVE

...