Both, whole-round sections and section halves could can be measured.
Procedures
Preparing Samples
...
- Equilibrate core sections to room temperature for at least (equilibration can take up to 4 hr) before measurement.
- Select measuring points in the core and record offset in cm (measured from the top - blue endcap).
– On intact cores, select the middle of section (section 3 or 4, if 7 sections).
– On cracked cores, choose a place with no fractures, just above/below the middle. - If measuring a whole-round section, drill a ~2 mm hole into the core liner on the splitting line between the working and archive halves. If the sediment is semi-consolidated, drill a small hole in the sediment for the needle probe as well.
- Optional - Apply thermal joint compound to the probe unless the sample is very soft and/or moist.
- Carefully insert a clean full-space needle into the sediment. Avoid twisting the needle into the core.
- If measuring a section-half, place the puck or mini-puck on a smooth place, avoiding cracks. Ensure puck to be is completely in contact with the sample.
- Secure the puck to the sample with velcro or a rubber band.
...
- Place the section-half or the whole-round section on in the TCON insulating box and close it. NOTE: If you are using a needle be sure that it is on in a vertical position to prevent it from being crushed when the door is close.
...
View and Verify Data
TCON data could can be view viewed and verify verified using the following programs:
- LIMS: After uploading data. LIVE is LIMS viewer. LORE allowed allows you to review data and download themdata.
- Thermcon File Converter (Figure 5): This program helps to convert TK04 format files to ANSI format. REVIEW
...