-The Torvane shear device (ELE International model 26-2261

...

; Figure

...

13) enables the user to quickly determine the shear strength of cohesive soils.
Three sizes of vanes are supplied with torvane

...

. The standard vane

...

has a stress from 0 to 1 kg/cm2, this is also the approximate range of torque that can be easily applied by finger pressure.
It is used for fully saturated cohesive soils with undrained strength independent of normal pressure. The stress range permits it to be used for clays varying in consistency from very soft to stiff.

...

A large vane is provided for use with remolded samples

...

, and has a ratio of 0

...

.2

...

. A smaller vane with a ratio of

...

2.

...

5 is used with stiffer clays

...


Image Removed

...

-

...

The pocket penetrometer was develop by Soiltest Inc. for use for field engineers to check the visual classification of soils. The readings of the penetrometer were compiled from several thousand unconfined compression tests of silty and clay-like soils

...


...







Measuring the Sample

Torvane

  1. Select the appropriate vane size and fit to vane driver (Figure 2). 

Image Removed

Figure 2.- Vane driver and the 3 kinds of vanes.

...

  1. 0.2, 1.0, 0.2 ratio)

0.22.51.0 Image Added

  1. Make sure zero on dial is aligned with index mark on knob

...

  1. , rotating a counter-clockwise the dial face

...

Image Modified

Figure 3.- Dial aligned.

...

  1. Test the surface

...

  1. should be reasonably flat and least two inches in diameter.

...

  1. Press torvane into soil to depth of blades and maintain a constant vertical pressure while turning knob

...

  1. . A rate of rotation such that failure develops in five to ten seconds is recommended

...


Image Modified

...

...

  1. After failure develops, release the remaining spring tension slowly and the index mark on the knob will indicate the maximum shear value

...



Image Modified

...



Penetrometer

...

The red

...

ring

...

on the barrel of penetrometer

...

hold the maximum reading of the scale

...

for

...

esy reading

...

Image Removed

Figure 6.- Red-ring in position to start measurement.

...

  1. Slide the ring down against the instrument handle Image Added
  2. Grip the knurled portion of the handle and push the tip of the penetrometer into the soil to the

...

  1. grove located ¼'' from the tip

...


Image Added




  1. Read the unconfined compression strength directly in tons per square foot (TSF) or kilograms per square centimeter (kg/cm2) from the scale. The reading is located in the red ring, side closest to the knurled handle

...


Image Modified

...

Uploading Data to LIMS

...






Uploading the data

  1. Open the spreadsheet for to enter the data (Figure 9).

Image Modified

...

...

  1. Scan the label of

...

Image Removed

Figure 10.- Spreadsheet with sample information.

...

  1. your sample

Image Added

  1. Select the instrument on cell (torvane or penetrometer), and write

...

  1. the

...

  1. value

Image Added

  1. Write the offset and select the

...

  1. adapter number

...

  1. (torvane

...

  1. )

...

  1. or

...

  1. if you use it or no (penetrometer)

Image Modified

Figure 11.- Measurement information on the spreadsheet.

...

  1. Select the

...

  1. direction in which you realize the measurement

...

Image Modified

...










  1. Select the "copy me" tab, down on spreadsheet and copy the rows with your sample data

...

Image ModifiedImage Modified

Figure 13.- Copy Me sheet.

7. Double-click the SpreadUploader icon on the desktop (Figure 14) and login using ship credentials.

Image Removed

Figure 14.- SpreadUploader icon.

...

  1. Image Added Open the SpreadUploader software and login into the app window. Image Added


  1. Paste the copied rows data and validate the sheet

...

  1. . A message will appear at the bottom if

...

  1. everything is fine.

Image Modified
Image Modified

Figure 15.- Successful data validation.

  1. Anchor
    _GoBack
    _GoBack

...

  1. Select Upload in the Lims button. A message will appear at the bottom

...

  1. when it is done

Image ModifiedImage Modified

...

That is all !!