Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

  1. Place hard rock samples in an ambient temperature seawater bath to equilibrate and saturate (4–12 hr). Keep sample saturated until measurement. A bell jar/vacuum pump can aid in saturation.

    NoteNOTE: Seawater bath is not necessary when using the mini-puck.

  2. If the surface of the split core is excessively rough, use a lap plate and grit from the Thin Section Lab to prepare a smooth surface on a split-core piece. Pieces must be at least 10 cm long.
  3. Equilibrate the sample and sensor together in an insulated seawater bath for at least 15 min prior to measurement. NoteNOTE: Do not submerge the puck – let the water level rise to half the depth of the puck or less.
  4. Secure the puck to the sample with velcro or a rubber band (Figure 2).
  5. Optional - Apply thermal joint compound to the side of the probe where the line source is located.

...

Figure 2.- Correct sensor position for measuring hard rock samples.


  • Making a Measurement

  1. Place the section-half or the whole-round section on the TCON insulating box and close it (Figure 3). NOTE: If you are using a needle be sure that it is on vertical position to prevent it from being crushed when the door is close.


Add picture

Figure 3.- Correct position of a whole round section on the insulating box.


A) ThermCon Program


1. Double-click the ThermCon icon (Figure

...

4a) on the desktop and login using ship credentials. ThermCon window will open (

...

Figure 4b). Ensure that the Text_ID field is blank.


Add picture

Figure 4.- (a) ThermCon icon.


2. Scan the section label. Sample_ID automatically fills.

3. Write down the measurement offset.

4. Click Verify Sample.

5. Text_ID of the sample will be generated.

NOTE: The folder path is shown on the screen. Do not close this window during measurement.





Measuring Sampl

Configuring the Measurement Program

...