Figure 2.- Correct sensor position for measuring hard rock samples.
Preparing Samples for Analysis
...
Making a Measurement
- Double-click the ThermCon icon (Figure 3) on the desktop and login using ship credentials. ThermCon window will open. Ensure that the Text_ID field is blank.
Measuring Sampl
Configuring the Measurement Program
1. | Load ThermCon software in offline mode. Ensure that the Text_ID field is blank. |
2. | Scan the core label using a scanner, then click Verify Sample. |
3. | If login is requested, enter UserName and Password and then click OK. |
4. | Run TK04 program and choose Measuring > Configuration. |
5. | Set configuration parameters as follows: –Probe Number: serial number of probe to be used in the measurement (Note: results may be wrong by several percent if the wrong serial number is entered or by a factor of ~2 if the wrong type of probe is entered). For the HLQ (large and mini puck), set equal to the estimated thermal conductivity value of the measured sediment. The mini puck often gets better results without the thermal compound when measuring the standard. Click Expert Options to configure Drift Control and Pause in Minutes (see Step 6). Enter comments. |
6. | To configure Drift Control and Pause in Minutes in Expert Options: –Pause in Minutes: insert a pause between single measurements; recommended = 10 min. This parameter does not apply when conducting a single measurement on each core or each core section. |
Measuring Samples
1. | Confirm configuration settings shown in the lower part of the TK04 screen, insert the probe into the hole drilled into the sample, and click Start Measuring. |
2. | Drift control (DCL) repeats until the criterion for drift (set in the Expert Options window) is met. Each drift control measurement takes 0.5 min. |
3. | After satisfying drift control, sample heating and measuring begins, and temperature values are corrected automatically for the drift effect predicted from the last drift series. Elapsed measurement time is controlled by the value entered in Configuring Measurement Program > Step 5. |
4. | Best solutions calculated by the TeKa SAM algorithm are shown on the screen. Note the sample result on the hand-written log. |
...