SHMSL Quick Start Guide
...
...
Table of Contents |
---|
...
Introduction
The SHMSL (Section Half Multisensor Logger) makes point-source, is a multisensor logger that makes contact measurements of magnetic susceptibility and color reflectance ; as well as a surface laser profile. Split core surfaces, predominantly from the archive half, are measuredon the split surface of cores. It also archives the laser profile of each section's surface for use in other applications such as routines to cull data from other loggers collected from voids.
...
Getting Started
Open
...
the SHMSL application by double-clicking the IMS icon on the desktop (Figure 1). IMS
...
will begin by initializing the Acuity AR700 laser, the Bartington MS2K point source susceptibility probe, and the Ocean Optics QE Pro spectrometer. Please ask the PP Tech for assistance in changing the instrument's measurement parameters. Once initialized, the logger is ready to measure a section.
Figure 1
...
: IMS Icon
Making a Measurement
NOTE: If you would like to change measurement parameters, please contact the PP technician.
- Place Place the archive section in the core tray . Make sure the section is oriented with the blue endcap up , and is flush against the benchmark (Figure 2).
Figure 2. - Benchmark - Adjust the section so that it is parallel to as flat as possible with respect to the plane of the benchmark.
- Click START. The the Start button to open the Section Information screen is displayed (Figure 3). .
Figure 3. - Section Information Screen - Place the cursor in the SCAN text field (center of Figure 3) and scan the section label. SAMPLE ID and LIMS ID automatically fills. These ID tags contain information on expedition, site, hole, core, section number and length.
...
- , the information automatically fills for Expedition, Site, Hole, and so on.
- Enter Comments (field is optional).
- Missing Top: If a sample taken at top of section (material is missing), position the section half all the way up to the benchmark. Enter missing length, software will automatically adjust.
- Exclude Intervals:
...
- enter section intervals to be excluded
...
- . Specify Interval: enter specific intervals to be measured.
...
- Measure Control Set: Measurement control for
...
- Click MEASURE
- The sensor assemby moves to the bottom of section. A surface profile, taken by the laser, will be acquired the QE Pro.
- When ready, click the Start Measurement button. Double check to make sure the Label Length matches the LIMS Length. If the QE Pro does not require calibration (see QE Pro Calibration), proceed to the next step.
- The sensor assembly moves down the track while the laser acquires a profile of the split surface of the section (Figure 4).
Figure 4.
...
Laser Profile
...
- After profile completion, a pop-up is displayed, containing After the surface profile is completed, the measured section length is displayed. Confirm this length visually (Figure 6) and adjust 5, below). Adjust if needed.
- Using the pre-cut GLAD wrap, cover the section surface (Figure 7). Click >>>GO to begin MS and color reflectance measurements.
Figure 6.- Confirmation length window
NOTE: Wrap MUST have a proper seal. Any bubbles or wrinkles WILL affect measurements. Sediment contamination within the sensors WILL cause DAMAGE. Figure 7.- Final Core Preparation
- MS and Color reflectance data graphs are displayed in real time during acquisition (Figure 8.)
NOTE: the QE Pro has indindividual display tabs for each component of the color reflectance. Display tabs can be toggled back and forth.
Figure 8.- Main window with measurement profiles
- After data acquisition, measurements of MSP (Magnetic Susceptibility Point) and white standards are performed. Placement of the sensors to their respective standard is extremely important. For the MSP measurement, this corresponds to the insertition of the sensor into the standard.
...
- All instrument data (laser profile, MSP, color reflactance, and standards) are stored in C:\DATA\IN.
- Data is automatically uploaded to the LIMS database, and is available to view on LORE and LIVE. If data is not instantly visible, please contact the PP technician.
- Upon upload, data is moved to C:\DATA\Archive. If upload is unsuccessful, data is automatically moved to C:\DATA\Error. Please contact the PP technician if this occurs.Apply Glad Wrap to the surface of sediment cores. Click GO!!!
IMPORTANT!!! Glad rap must have a good seal to the surface of the core; any bubbles or wrinkles will affect the measurement. Any mud that gets inside the integration sphere will ruin it!!! - Click GO!
Figure 5. Final Sample Preparation - The measurement sequence begins. For efficiency, measurements begin at the bottom of the section and move upcore. Results are displayed during acquisition on instrument graphs on screen as shown in Figure 6, below.
Note that the QE Pro has several tabs that display different views of the color reflectance data. Tabs can be changed during the measurement sequence. - After the section has been measured, the logger takes measurements of the MS2K and white standards as verification.
Figure 6. Data Acquisition Screen
Data Storage and Format
C:\data\IN
- Section half L* a* b* and XYZ data
- MS2K check standard result
- MS2K (MSPOINT) data file
- QE Pro white check standard result
- Laser profile pointer file
C:\AUX_DATA\PROFILE
- Laser profile data
C:\AUX_DATA\RSC\CALIB
- White and Dark calibration spectral data per wavelength (380–900 in 2 nm steps)
C:\AUX_DATA\RSC\CNTRL
- Raw spectra and normalized (% color reflectance) spectral data per wavelength (380–900 in 2 nm steps) for the white check standard at the end of a run
C:\AUX_DATA\RSC\RUN
- Raw spectra and normalized (% color reflectance) spectral data per wavelength (380–900 in 2 nm steps) for the measured section
Uploading Data to LIMS
1. | Uploading SHMSL data to the LIMS database is executed via the MUT application. Click the MegaUploadaTron (MUT) Icon (Figure |
Figure 10.- MUT Icon
7, below). | |
2. | Once the application is activated, |
it displays a table-like list of files |
in the C:\DATA\IN directory |
. Files are |
ready to be uploaded if they have a green check mark |
Figure 11.- Main MUT screen
next to them as shown in Figure 8, below. Log into MUT using database credentials. |
3. |
Click Automatic Upload if you want files to upload automatically, or check each file individually and |
click upload. |
4. |
If files are marked by a purple question mark or red and white X icons, please contact a technician.
Purple question mark: cannot identify the file. Red and white X icons: contains file errors.
...
Please ask a technician for help if you see these indications: |
Figure 8. MUT Uploader
QE Pro Calibration
1. |
If the QE Pro is out of calibration, the Instrument Calibration screen will appear. | |
2. | Click the Calibrate Instruments button to begin the calibration process |
. The logger moves the QE Pro integrating sphere |
over the |
white calibration standard |
Figure 12.- White Calibration
Note: If the integration time increases, contact the PP technician
and lowers it until it touches. It will take some preliminary measurements to establish the integrating time based on an 80% saturation level of the total response. These measurements take about 0.30–0.55 seconds for the current light and detector configuration. (Figure 9, below) | |
3. | Click Accept for the suggested integration time. The logger will then take 20 measurements of the white standard at the automatic integration time. Figure 9. White Calibration |
4. | After the |
white measurements are completed, the |
When finished click Accept. Figure 13.-
logger will raise the integrating sphere, position it over the dark standard, and lower it on the dark standard. The shutters on the light sources will close and the dark measurement will acquire 20 measurements, as above. When finished, Click Accept! |
5. |
Figure 14.- Normalize spectra
Figure 12 shows the normalization of spectra that will occur with the just-acquired calibration. |
Figure 12. Calibration Normalization Factor
Archived Versions
View file | ||||
---|---|---|---|---|
|