  • Introduction
  • Making a Measurement
  • Uploading Data to LIMS
  • QE Pro Calibration

...

  1. Place section in the core tray. Make sure the section is oriented blue endcap up, and is flush against the benchmark (Figure 2). Figure 2.- Benchmark
  2. Adjust the section so that it is parallel to the benchmark.
  3. Click START. The Section Information screen is displayed (Figure 3).  Figure 3.- Section Information Screen
  4. Place the cursor in the SCAN field and scan the section label. SAMPLE ID and LIMS ID automatically fills. These ID tags contain information on expedition, site, hole, core, section number and length.

...

  1. All instrument data (laser profile, MSP, color reflactance, and standards) are stored in C:\DATA\IN.
  2. Data is automatically uploaded to the LIMS database, and is available to view on LORE and LIVE. If data is not instantly visible, please contact the PP technician.
  3. Upon upload, data is moved to C:\DATA\Archive. If upload is unsuccessful, data is automatically moved to C:\DATA\Error. Please contact the PP technician if this occurs.




Uploading Data to LIMS

...

  1. Click the MUT Icon (Figure 10). Log in using database credentials.

    Figure 10.- MUT Icon

...

  1. Once activated, the list of files from the C:\DATA\IN directory is displayed. Files are marked ready for upload by a green check mark (Figure 11).

    Figure 11.- Main MUT screen

...

  1. To manually upload files, check each file individually and clip upload. To automatically upload files, click on the Automatic Upload checkbox

...

  1. .
  2. If files are marked by a purple question mark or red and white X icons, please contact a technician.
    Purple question mark: cannot identify the file. Red and white X icons: contains file errors.