Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

Table of Contents
maxLevel3

...

Thermal conductivity is measured by transient heating of an isotropic material with a known heating power generated from a source of known geometry and measuring the temperature change with time. The needle probe contains a heater wire and calibrated thermistor. It is assumed to be a perfect conductor because it is much more conductive than unconsolidated sediments. With this assumption, the temperature of the probe has a linear relationship with the natural logarithm of the time after initiation of heating:

T(t) = (q/4pk4πk) ln(t) + C,

where:

T = temperature,

...

Process

Time (min)

Comments

1. Obtain a whole-round core section from the core rack

0.3

See Preparing Sections & Samples


2. Locate the appropriate probe for the sample type

0.5

3. Verify sample identification in software

0.5

See Set Measurement Parameters


4. Configure measurement program

0.3

5. Perform drift control

5

See Making a Measurement




6. Heat and measure sample

2

7. 10 minute pause between measurements

10

8. Repeat steps 5-7 for 2 additional measurements (3 total)

34

9. Upload results to LIMS

0.2

See Uploading Data to LIMS



10. Check results in LIMS

1

11. Remove the section and deliver to splitting room

0.2

Total Time per sample:

54 (max)


...

  1. Load ThermCon software in offline mode. Ensure that the Text_ID field is blank.
  2. Scan the core label using a scanner, then click Verify Sample.
  3. If login is requested, enter UserName and Password and then click OK.
  4. The folder path is shown on the screen. Do not close this window during measurement. 
  5. Run TK04 program and choose Measuring > Configuration
  6. Set configuration parameters as follows (see figure below):
    1. Probe Number: serial number of probe to be used in the measurement (Note: results may be wrong by several percent if the wrong serial number is entered or by a factor of ~2 if the wrong type of probe is entered),
    2. Root Name: six characters or less; suggest Core-Type-Section (no special characters in the root name).
    3. Serial Number: number of repeat measurements at each measurement point (1–99 single measurements).
    4. Folder: path for saving data results.
    5. Heating Power: for the VLQ (needle probe), set to twice the estimated thermal conductivity value of measured sediment. For example, 2–3 is good for sediment. (See the Appendix: TK04 Recommended Heating Power for power guidance.)
    6. Measuring Time: set to at least 80 s, or for mini HLQ 60 s.
    7. Click Expert Options to configure Drift Control and Pause in Minutes (see Step 7).
    8. Enter comments.

...