Manual Information
...
Manual Information
Author(s): | C. Bennight | ||
Reviewer(s): | D. Houpt, K. Fujine, L. Brandt | ||
Revised by: | V.Percuoco, D. Houpt |
| |
Manager Approval (Name, Title, Date): | D.J. Houpt (Supervisor of Analytical Services) | ||
Audience: | Scientists, Laboratory Technicians | ||
Origination date: | 3/1/2009 | ||
Current version: | V 1.0 | 9/08/2010V3741T | July 2017March 2018 |
Previous revisions:Revised: | V1.1 | 1/6/2014 (IODP-II), V 1.0 | 9/08/2010 |
Domain: | Chemistry | ||
System: | Coulometry |
User Guide Contents
Topic | See page… |
Apparatus, Reagents, & Materials | |
Sample Preparation | |
Sample Analysis | |
Quality Assurance/Quality Control | |
LIMS Integration | |
Health, Safety, & Environment | |
Maintenance/Troubleshooting | |
Installation Guide.0 | 9/08/2010 |
Domain: | Chemistry |
System: | Coulometry |
User Guide Contents
Anchor | ||||
---|---|---|---|---|
|
...
IODP's UIC Coulometrics CM5011 CM5015 coulometer provides absolute determination of the concentration of carbon dioxide (CO2) evolved from an acidification process. The coulometer cell is filled with a proprietary solution containing monoethanolamine and a colorimetric pH indicator. A platinum cathode and silver anode are positioned in the cell, and the assembly is located between a light source and a photodetector. When a gas stream passes through the solution, CO2 is quantitatively absorbed, reacting with the monoethanolamine to form a titratable acid. This acid causes the color indicator to fade. A spectrophotometer monitors the change in the solution's percent transmittance (%T). As %T increases, the titration current is automatically adjusted to generate a base at a rate proportional to the reduction of %T. When the solution returns to its original color (original %T), the current stops. The amount of CO2 evolved is quantitated calculated from the duration and magnitude of the current required to balance the acid by CO2 evolution. Based on the principle of Faraday's Law of Electrolysis (the quantity of a substance produced by electrolysis is proportional to the quantity of electricity used), each mole of electrons added to the solution is equivalent to 1 mole of CO2 titrated.
Chemical reactions occurring in the coulometer cell follow:
Absorption of CO2 by the cathode solution (cathode reaction):
CO2 + HOCH2CH2NH2 —> HOCH2CH2NHCOOH
Electrochemical generation of OH– (cathode reaction):
2H2O + 2e– —> H2 (g) + 2OH–
Neutralization of absorbed CO2 reaction product by electrochemically generated OH–:
HOCH2CH2NHCOOH + OH– —> HOCH2CH2NHCOO– + H2O
Anode reaction:
AgO —> Ag+ + e–
Interferences
A variety of carrier gases can be used for coulometry (O2, N2, He, and dry air). The JRSO uses N2 for the measurement. Interferences caused by compounds such as SO2, SO3, H2S, HCl, HBr, HI, and Cl2 are removed with KOH and AgNO3 scrubbers.
...
- Coulometer unit (UIC CM5011CM5015) with titration cell (Figure 1)
- Acidification module (similar to UIC CM5030) (Figure 2)
- Dual balance system, motion-compensated, with control software
Figure 1. Model CM5011 CM5015 Coulometer. | Figure 2. Acidification Module. |
Dual Balance System Hardware
A Cahn balance and 2 Mettler Toledo XS204 analytical balances with motion compensation software are used to measure the mass of samples and chemicals. The Cahn balance (Figure 3) measures samples for the Coulometer.
...
Motion compensation software developed in house allows the user to weigh the mass of chemicals and samples at sea. Reagents must be measured on the Mettler-Toledo XS204 balance using the Balance Master program (see Balance User Guide)(Figure 4). Sample material must be measured on the Cahn balance (unless the sample is larger than ~1 gram) (Figure 5Figure 4).
Figure 4. Mettler-Toledo Dual Balance Control Software.
...
- Nitrogen (99.995% or better) is used as carrier gas to minimize the amount of CO2 the scrubber (KOH) must absorb
Reagent Solutions
- 45% KOH ([%w/v]: add 90 g KOH pellets to water and make up to 200 mL once fully dissolved. Warning!This procedure liberates caustic fumes and heat. Perform in a fume hood.)
- 3% AgNO3 ([%w/v]: dissolve 3 g silver nitrate in water and make up to 100 mL when fully dissolved.)
- Caution! Concentrated acids should always be added to water, and may generate significant heat.
- 2N H2SO4: add 55.5 mL concentrated sulfuric acid to water and make up to 1L
- 2N HCl: add 166 mL concentrated hydrochloric acid to water and make up to 1L
...
- A scientist or staff member logs wet sample information into SampleMaster at the sampling table. The sample is given the name CARB to ensure proper routing.
- Freeze-dry the sample (Freeze-Drying the Sample).
- Homogenize (grind) the sample (Grinding the Sample).
- Weigh the sample, assign a container and code, and upload the mass data to LIMS (Weighing the Sample).
- Prepare the coulometer acidification for analysis (Preparing Acidification Module and Coulometer Cell).
Freeze-Drying the Sample
...
- Log into the Dual Balance system for the Cahn Balance. Answer Yes or OK on all prompts that appear during the log-in process. The user's log-in ID must be same as the LIMS database ID.
- Click Test Option, and enter a number (usually >100 based on sea state; see the technician for guidance). Click Save/Exit to return to the main window.
- Fold a small piece of wax paper (~0.5 cm x 0.5 cm) on opposite edges to create a U-shaped wax paper sample boat. Place the wax paper boat on the left weighing pan. Place a similar size of paper on the tare pan (right). Close the door, click Tare, and then Start on the plot screen.
- Once the measurement is finished and the value is acceptable, click Get Mass. The tare value will be changed and the display will clear.
- Put the sample on the weighing pan (~7–13 mg) using the scoop.
- Press Weigh on the screen and then Start on the plot panel. The Weigh measurement will not begin if you do not press Start.
- Once the measurement is done and the value is acceptable, click Get Mass. Final mass value (under the weigh button) will be changed and the display will clear.
- Select COULOMETER from the Objective from the list, then and enter a part of the text ID or label ID of the sample, then click Search.
- Select a the appropriate sample from the list, then click Assign to return to the main window.
- Enter a container number, and click Save to save the mass value into the LIMS. Write down on a piece of paper the mass, container number, and text_id. Keeping a good logbook of your experiments is highly recommended!
...
- Add granular KI to the empty small section of the Carbon Coulometer Cell (the anode cell) to a depth of 5 mm from the bottom of the cell (Figure 6, far right).
- Fill the large section of the Carbon Coulometer Cell with cathode solution to a mark 4 cm from the base.
- Fill the small section of the Carbon Coulometer Cell with anode solution to a mark 4 cm from the base.
- Important! Do this Add the anode solution quickly (within 1 min) after filling the cathode cell, or else the cathode solution will start filtering through the junction between the cells and contaminate the anode solution.
- Fill the KOH pre-scrubber trap 1/2 full of 45% KOH solution.
- Fill the AgNO3 post-scrubber trap 1/2 full of 3% AgNO3 solution.
- Add 3 drops of 2N H2SO4 to the AgNO3 trap.
- Attach the input gas tube (carrier gas inlet) to the KOH trap.
- Turn on the gas flow and set to 100 cm3/min.
- Connect the KOH trap to the reaction flask.
- Connect the reaction flask to the horizontal fitting on the AgNO3 trap.
- Connect the top of the AgNO3 trap to the Carbon Coulometer Cell.
- Connect the anode/cathode to the titration cell ports next to the titration cell.
...
- Turn on the heating unit and power to the main coulometer unit.
- With titration cell in place, adjust transmittance to 100%.
- Turn on power to the cell.
- Calibrate the instrument (see Calibration) or verify calibration (Calibration Verification), as applicable.
- Login to the Coulometer software using a LIMS login.
- Highlight a sample to be measured. Replicates of a sample (same TEXTID) are stored within the same line of the sample list. A dropdown option appears over the sample name allowing the user to select the desired replicate.
- Connect the sample vial to jacketed condenser component of the sample introduction system (Figure 6). Ensure the connection is airtight. Then slowly add 5 mL of 2N HCl using the connected repeater dispensette.
- Quickly press Measure in the sample list page of the coulometer software. If the measurement is delayed the results may underestimate the calcium carbonate percentage. A measurement screen will appear displaying real time data acquisition, the options to abort or stop the measurement, and to save/not save the results. The slope threshold is a measurement of the µg carbon with respect to time, and may be adjusted to specify the stopping point of the titration. Setting the slope threshold too low increases measurement times with the possibility of including circuit noise in the results, whereas setting the threshold too high will cause the measurement to prematurely terminate. The default slope threshold is 0.1.
- The cell solution will fade upon dissolution of carbon dioxide and will return to a blue color (i.e., the start point) during titration.
- After the measurement is complete, press Save or Don't Save to keep or disregard the data. A few reasons to not save data:
- Sample powder coated the sides of the vial and was not dissolved by the acid.
- The amount of calcium carbonate was so low its signal is greatly influenced by instrument noise.
- The slope threshold was set incorrectly.
- There may be constituent siderite in the sample that confounds the results. Siderite tends to react with the acid less quickly than calcium carbonate
- After saving the data the measurement screen will revert to the sample list screen.
...
- Sample tubes: rinse sample tubes with DI water and place into the oven to dry. They do not need to be acid washed.
- Cell: clean the cathode/anode cell in a fume hood by adding acetone to the anode cell. The acetone will leach through the bridge between the cells and clean it. Follow the acetone rinse by placing DI water in the anode cell and letting that leach through.
- Platinum electrodes: Electrodes can acquire surface coatings from the solutions. Remove this coating by placing the electrode in a solution of 1:1 concentrated nitric acid solution for 1 hr: water for 20 seconds. Rinse with DI water immediately.
Data Handling
Weight percent calcium carbonate is calculated from µg carbon measured during the titration as follows:
%CaCO3 = µg C x 8.333/sample mass
Sample mass is stored in LIMS associated with the container ID that the coulometer analysis is associated with.
...
–Incompatible materials: alkaloid salts, chloral hydrate, potassium chlorate, metallic salts, tartaric and other acids, bromine trifluoride, fluorine perchlorate
Waste Management
Waste may be washed down drain with flowing waterof cathode and anode solutions should be collected in a bottle until it can be removed during the next port call. The potassium hydroxide and silver nitrate solutions may be disposed of in the sink.
Maintenance/Troubleshooting
...
Potential explanation | Solution |
Lamp brightness has deteriorated with age | Replace lamp (CM140-005) |
Path to detector is blockedLight path blocked | Check for physical blocking of the light path; you will need to run a new Cell Setup once the cell is moved |
Lamp voltage is incorrect | Measure lamp voltage (see Measure Lamp Voltage) |
Detector and/or filter are cloudedDefective photodiode | Replace filter (CM140-001) or photodiode (CM140-002). It is best to replace entire photodiode subassembly (CM101-178).Contact UIC for repair |
Detector is defectiveDefective amplifier circuit | See Evaluate Electronics Contact UIC for repair |
Loose connection on front end board | Locate the front end board (CM110-020). Ensure all connectors to the board are plugged in securely; reset connectors by pushing on them. |
Electronic problem on circuit board | Run electronics checks (see Evaluate Electronics) |
...
The calculated % Difference for any of the Analysis Types should be below ± 0.15%. If any of the values are > 0.15%, contact UIC for a bench calibration of the instrument.
Thorough Cleaning
At times, component parts may require a more thorough cleaning. To clean the frit, fill the cell with enough 1:1 concentrated nitric acid to water solution to cover the frit and allow the acid to clean the frit overnight. Dispose of the acid and rinse the cell and frit completely with water before re-use. If the potassium iodide solution turns brown after refilling the anode compartment, the frit has not been sufficiently rinsed.
- With no cell in coulometer, install a shorting strap and turn on current.
- Set coulometer as follows:
...
- Clean compressed air (oil-free; zero grade preferred) ¿ 40 psi
- Two 110 V outlets; 8 A peak
- Vent for reaction effluent (preferred as effluent smells bad, releasing amine derivatives)
- Counter area ¿ 2 ft x 2 ft
- Cooling capacity = 800 btu
Hardware Setup
Be certain that the CM5015 is running in CM5011 emulation mode for proper interface with the JRSO software. Also note that the "latch" commands used with the CM5011 are not applicable to the CM5015.
Coulometer Serial Port Jumper Configuration
...