...
Figure 3. QE Pro White Calibration.
Note: If the integration time increases (i.e., reaches 0.55 and beyond), please contact the PP technician.
3. Click Accept.
4. After the White Calibration is complete, the Dark Calibration will begin (Figure 4). The difference between the calibration values of Dark Pixels and the Spectral Mean should be less than 100. If this difference is higher or if negative values appear on the spectrum, please contact the PP technician.
When finished click Accept.
...
9. MSP and Color reflectance data graphs are displayed in real time during acquisition (Figure 10).
NOTE: the QE Pro has indindividual individual display tabs for each component of the color reflectance. Display tabs can be toggled back and forth.
...