Table of Contents
...
Table 2: Serial dilution scheme for PFMD. As of Exp395C, we are only using the Low Level calibration.
PFD calibration Level | µL STD A added to 20 mL screw cap GC2 vial | Concentration PFD (ng/mL headspace) |
BLANK | 0 | 0 |
1 | 0.02 | 0.1 |
2 | 0.04 | 0.2 |
3 | 0.06 | 0.3 |
4 | 0.1 | 0.5 |
5 | 0.2 | 1 |
6 | 1 | 5 |
7 | 2 | 10 |
8 | 4 | 20 |
Table 3: Dilution scheme for PFD.
...
Both PFMCH and PFMD are considered to be non-hazardous by US and EU classifications under normal conditions.
Chemical Hazards:
- Incompatible substances: oxidizing agents, strong acids, strong bases
- Emits toxic fumes (HF) under fire conditions
...