Table 1 contains the values of the standards as they were measured in California before receipt by IODP. Note that the units of intensity in the provided values was in emu and the volume used was 1cc. Intensity has been converted to A/m as this is the unit used by IODP instruments and the intensity has also been listed if you assume a 7cc volume.. The conversion used was 10-1 A/m = 10-3 emu. It is extremely important that these standards do not get demagnetized. Be sure to stress this to anyone who may use the standards.

Table 1 Paleomagnetic Standards Values

Sample Name

Declination

Inclination

Intensity (A/m) as provided (assumes 1cc)

Intensity (A/m) assuming 7cc volume

csd

IODPSTD01

217.4

33.9

3.17E-01

4.529E-02

0.9

IODPSTD02

134.2

35.5

8.48E-02

1.211E-02

0.8

IODPSTD03

143.5

-29.1

9.40E-02

1.343E-02

1.3

IODPSTD04

227.5

-22

6.36E-02

9.086E-03

0.1

IODPSTD05

18.9

-24.9

9.09E-02

1.299E-02

0.4

IODPSTD06

137.6

28.2

4.20E-01

6.00E-01

0.5

IODPSTD07

326

34

5.86E-01

8.371E-02

0.7

IODPSTD08

221.8

-12.2

7.82E-02

1.117E-02

0.6

IODPSTD09

34.7

-28.4

1.13E-01

1.614E-02

1

IODPSTD10

316.2

-19.8

6.39E-02

9.129E-03

0.4

IODPSTD11

333.3

24.3

9.78E-02

1.40E-02

1

IODPSTD12

28.4

40.2

1.96E-01

2.80E-02

0.4


Standard Orientation

The Samples are oriented as if they were taken from an archive half. This can be confusing as most of the time, discrete samples would be collected from the working half. Be sure to keep this orientation in mind when loading into the instruments.  See Figure 3 for sample orientation arrows

...