Microscope property numberMicroscope TypeDiagnosis
90871 Axioscope A1 POLStage counterweights had to be cut down which was accomplished for all three specimen holder sets of that type on board
90869 Axioscope A1 POLXY stage faulty (inner pulley system problem), Serco can’t fix this, needs to go back to Zeiss – tested 3 other stages, 2 out of 3 are broken and will be send to College Station to be forwarded to Zeiss
91187 Axio Imager A2

- known issue with PCBA board, which has missing parts on order,

- scope is set up for fluorescence, unsure why in that spot


91188 Axio Imager A2 (MBIO)

No issue reported.

90229 Axioplan POL

Bertrand lens is de-laminated, attachment not useful, so it was removed.

90207 Axioscope POL (Thin Section)

No issues reported.

90257 Axiophot

Stage couldn’t be moved far enough to focus on specimen, it was discovered that gear inside wasn’t lined up, Serco had to remove focus knob covers and align 15 O-rings back to where they are needed in order to move stage to correct positions.

Recommendation from Serco – don’t tinker with inner workings of scopes if not absolutely necessary, send to Zeiss or have Zeiss rep service call.

91060 Stereoscope

Magnification knob doesn’t stop at 1x or 8x, no way to fix it as body is glued shut. Stored in Imaging Lab as fall back for high usage expeditions.

90812 Stereoscope

No issue reported.

91105 Stereoscope

No issue reported.

90811 Stereoscope

No issue reported.

90812 Stereoscope

No issue reported.

90228 Stereoscope SV11 (reserve in Imaging Lab)

No issue reported.

ODP00584 Old grey stereoscope (reserve in Imaging Lab)

No issue reported.


At the end of the tie-up, AF arrived and the missing parts were installed and tested on the Axio Imager 2 (91187):

Image Added

Microscope cameras (SPOT)

...

  • Printer ink for Epson ecotank and Stylus
  • Phone case/screen protector for Foldio iPhone
  • Light bulbs for Imaging Office
  • Tripod head that holds more weight (for remote setup for group pictures)
  • Light bulbs for Imaging Lab overhead light fixture
  • Missing pieces for Microscope Imager A2 - push button and cable (AF @ end of expedition)

Items ordered

  • Small tripod for Foldio phone
  • Misssing pieces for Microscope Imager A2 - push button and cable (beginning of expedition)

Items shipped to shore

  • Prepared shipment with all camcorders as the technology has not been used in a decade (see tech reports) and is outdated at this point
  • Old grey Zeiss microscope and accessories
  • Several light box transformers (we have too many here)
  • Two Maerzenhauser XY stages that only work in one direction (for repair at Zeiss)

...

One additional task was to prepare for the new SEM/EDS system install during 397T. A tracking confluence page has been created, where all information, manuals, progress etc. are reported. It would be nice to keep using this tracking page until the project is finished and approved by all stake holders. The page is located on the front page of the Laboratory Notebook and linked: 379PT 379P/T SEM EDS install. All tasks associated with 397P have been accomplished.

...