Chloride by Autotitrator: User Guide

...

(from CI_Autotitrator_UG_371T_draft_
V1.1 (C. Bennight 10/2/2011; Approved by DJH 10/23/2013)
V371T (D. Houpt, July 2017)

...

...

Introduction

This manual describes the operational methods, standard maintenance, and systems integration of the Metrohm 785 DMP Chloride Autotitration System (Figure 1). Methods describes in this manual are based on the Metrohm manual, Metrohm Application Bulletin 130/2 e (Chloride titrations with potentiometric indication), and email correspondence with Gretchen A. Robertson at the Scripps Institution of Oceanography.
This manual covers the USIO standard method for the determination of chloride via potentiometric titration, Variation A – with Nitric Acid.


Image Modified
Figure 1. Chloride Autotitration System.

...

...

Theory of Operation

...

The

...

Cl

...

autotitrator

...

uses

...

a

...

potentiometric

...

endpoint

...

to

...

determine

...

the

...

amount

...

of

...

free

...

(unbound)

...

chloride

...

in

...

a

...

solution.

...

A

...

silver

...

nitrate

...

reagent

...

is

...

aliquotted

...

into

...

an

...

unknown

...

solution

...

with

...

some

...

amount

...

of

...

dissolved

...

chloride

...

in

...

solution.

...

As

...

the

...

silver

...

nitrate

...

is

...

added

...

the

...

following

...

reaction

...

occurs.

...

Ag

...

+

...

(aq)

...

 

...

+

...

Cl(aq)

...

 ® AgCl(s)

...

The

...

product,

...

AgCl,

...

is

...

insoluble

...

in

...

water

...

and

...

precipitates.

...

A

...

silver-specific

...

electrode

...

in

...

the

...

solution

...

measures

...

a

...

value

...

proportional

...

to

...

the

...

amount

...

of

...

free

...

silver

...

in

...

solution.

...

 

As

...

the

...

silver

...

chloride

...

reagent

...

is

...

added,

...

a

...

measurement

...

(in

...

mV)

...

is

...

taken

...

from

...

the

...

silver-specific

...

electrode.

...

If

...

we

...

consider

...

the

...

function

...

of

...

volume

...

added

...

versus

...

the

...

electrode

...

reading

...

and

...

find

...

the

...

volume

...

where

...

the

...

second

...

derivative

...

of

...

the

...

function

...

is

...

0

...

(or

...

the

...

inflection

...

point),

...

we

...

have

...

found

...

our

...

equivalence

...

point

...

(Figure 2).

Image Added

Figure 2. Potentiometric Equivalence Point.

For most purposes, this first equivalence point corresponds to the volume where all free chloride has been converted to AgCl. Knowing the volume of reagent added, the concentration of reagent, and the stoichiometric ratio between reagent and chloride (1:1),

...

we

...

can

...

determine

...

the

...

amount

...

of

...

chloride

...

contained

...

in

...

the

...

solution:

...


[Cl

...

] = ([AgCl

...

]

...

×

...

Volume

...

(AgNO3))

...

/Volume

...

(Sample)

...


(Note:

...

A

...

value

...

in

...

brackets

...

(

...

[

...

])

...

indicates

...

the

...

concentration

...

of

...

the

...

item

...

enclosed

...

in

...

brackets.)

...

Volume

...

(AgNO3)

...

is

...

simply

...

the

...

volume

...

of

...

reagent

...

added

...

when

...

the

...

first

...

equivalence

...

point

...

is

...

reached.

Assumptions

It is assumed that the first equivalence point corresponds to that of the AgCl reaction. Silver forms sparingly soluble precipitates with other anions as well as chloride. Solutions with high concentrations of bromide or iodide may cause an erroneous first equivalence point (as they titrate before chloride). In these cases, the method on the instrument will need to be adjusted to run past the first endpoint and to determine the Volume(AgNO3) value. The reagent volumes for the non-chloride endpoints must be subtracted from the reagent volume for the chloride endpoint. It should be fairly obvious which the chloride endpoint is, and chloride will typically be many times more concentrated in situ than other anions.

...

...

Apparatus, Reagents, and Materials

Hardware

785 DMP Titrino

The 785 DMP is the primary piece of the instrument. All other components connect to it and it provides the logic and outputs to run the titration (Figure 3).


Image Modified
Figure 3. Titrino Autotitrator.

Reagent Exchange Unit

This interchangeable unit contains the following:
–500

...

ml reagent bottle
–Exchange burette (5

...

ml)
–Reagent tubing
–Reagent dispensing tip

Titrino Keypad

This auxiliary data entry device can be used to control the Titrino manually. It is not needed for any of the methods described in this manual.

Silver Electrode

Specifically, a Metrohm 6.0430.100, their "Ag Titrode," this electrode combines a standard silver ring electrode with a pH membrane to serve as the reference. It is the only electrode required for this method. It measures the concentration of silver ions in solution.

...

Automatic Stirrer

The metal ring on the tip of the silver titrode may be periodically cleaned by scrubbing gently with toothpaste. Cleaning in this manner will reduce the titrode efficacy over time. The titrode should be stored in DI water long term. Trichloroethylene may be used to degrease the probe. See Storage of Metal Electrodes for more info.

Automatic Stirrer

This electronic stirrer is controlled by the 785 DMP Titrino. Note that the stir bar will not activate until the titration has started.

Reaction Vessel

This is a non-water-cooled narrow bottom reaction vessel. The electrode and reagent dispensing tip are inserted down into the narrow portion of the vessel to reduce the volume required for the overall reaction.

Reagents

...

2.2 M nitric acid solution

...

Dilution Solution

  • Add about 500 mL of 18 MΩ deionized (DI) water to a 1L volumetric flask. Pipette in 6.08 mL of concentrated trace-metal grade nitric acid

...

  • . Make up to 1 L

...

  • with 18

...

  • deionized (DI) water (reagent water). Note: Always add acid to water

...

0.10 M silver nitrate solution

...

  • .

Note: Make this solution fresh for each run (typically keep no longer than 1 day). Adjust proportions of solution to make the desired final volume.

0.015 M silver nitrate solution

Dilute 150 ml of 0.1 M silver nitrate stock solution with 1 l reagent water, or dissolve 2.548 g silver nitrate in 1

...

l reagent water.

Dilution Solution

To make 500 mL of acid-based dilution solution, add the following to a 500 mL volumetric flask:

Solution

Volume (mL)

*18 m¿ DI water (reagent water)*

430

2.2 M trace metal grade nitric acid in reagent water

20

0.2% polyvinyl acid (PVA) in reagent water

50

...

Setting Up The Instrument

To install the instrument for the first time:

1.

Remove the 785 DMP unit from the container, locate the following cords, and plug them into the indicated slot (Figure 4):
–Connection 11: Power cord
–Connection 9: Silver Titrode cord (connected to titrode)
–Connection 12: Electronic Stirrer cord (connected to electronic stirrer assembly)
–Connection 7 (bottom): Serial Cord (straight-through, Female/Female) (A2 on back of device)
–Connection 10: Keyboard cord (connected to the keyboard device)

2.

Connect the other end of the serial cable to the back of the computer host that will control this device.

3.

Power on the instrument by pressing the power button on the front panel of the 785 DMP unit.

4.

Flip the power switch up on the front of the electronic stirrer (stirring will not start until the run actually begins – controlled by the 785DMP unit).

Image Removed
Figure 4. 785DMP Connector Slots.

Operation

Purging the lines

The lines connecting the reagent bottle to the reagent dispensing tip must be purged before analysis can begin. Wear gloves and safety glasses for this operation.

...

1.

...

Place the acid dispensing tip in a Ziploc bag—but make sure it's not completely air tight.

...

2.

...

Tap, clamp, or have an individual hold the bag and tip up so it is the highest portion of the system.

...

3.

...

Turn the dispensing speed knob in the bottom right of the device fully clockwise to its maximum setting (if it isn't there already).

...

4.

...


General Equipment Setup and Operation

Setting Up The Instrument

To install the instrument for the first time:

1.

Remove the 785 DMP unit from the container, locate the following cords, and plug them into the indicated slot (Figure 4):
–Connection 11: Power cord
–Connection 9: Silver Titrode cord (connected to titrode)
–Connection 12: Electronic Stirrer cord (connected to electronic stirrer assembly)
–Connection 7 (bottom): Serial Cord (straight-through, Female/Female) (A2 on back of device)
–Connection 10: Keyboard cord (connected to the keyboard device)

2.

Connect the other end of the serial cable to the back of the computer host that will control this device.

3.

Power on the instrument by pressing the power button on the front panel of the 785 DMP unit.

4.

Flip the power switch up on the front of the electronic stirrer (stirring will not start until the run actually begins – controlled by the 785DMP unit).

Image Added
Figure 4. 785DMP Connector Slots.

Operation

The settings for the titrator can be accessed by connecting the titrino's detachable keypad to the instrument. The titrator should be set with the following settings using the keypad.

  • Mode
    • Set mode to DET
  • Config
    • COM1 and COM2
      • set Handshake to none
    • Peripheral Controls
      • set Stirrer Control to on
  • Parameters
    • Titration Parameters
      • set Meas. Pt. Density to 0
      • set Signal Drift to 10
    • Stop Condition
      • set Stop EP to 1

Purging the lines

The lines connecting the reagent bottle to the reagent dispensing tip must be purged before analysis can begin. Wear gloves and safety glasses for this operation.

1.

Place the acid dispensing tip in a Ziploc bag—but make sure it's not completely air tight.

2.

Tap, clamp, or have an individual hold the bag and tip up so it is the highest portion of the system.

3.

Turn the dispensing speed knob in the bottom right of the device fully clockwise to its maximum setting (if it isn't there already).

4.

Press and hold the DOS button on the front of the 785 Titrino. While the unit is dispensing flick the reagent lines with your fingers to dislodge any stuck air bubbles.
Note that in DOS mode you will only be moving fluid (and thus purging) lines between the exchange burette and the reagent dispensing tip.

5.

When the burette has exhausted its capacity (5

...

ml), press the

...

FILL button on the front of the 785 Titrino. This will cause the burette to aspirate reagent from the bottle. While this is occurring, flick the exchange lines between the burette and the reagent bottle to dislodge air bubbles.

6.

Repeat steps 4 and 5 until no air bubbles are visible in any of the exchange lines.

Shutting down the instrument

1.

Rinse the dispensing tip in DI water and store in DI water (add a small amount to the housing in the exchange unit where the tip sits).

2.

Rinse the electrode in a 3 M sodium nitrate solution or 2 M nitric acid solution followed by DI water.

3.

Store the electrode in

...

DI water (add to the housing in the exchange unit where the electrode sits).

Analyzing Samples

1.

Place stir bar in bottom of reaction vessel.

2.

Pipette 100 µL of sample into the bottom of the reaction vessel.

3.

Pipette 4 mL of dilution solution into the bottom of the reaction vessel.

4.

Insert the electrode and reagent dispensing tip into the reaction vessel.
Note: The bottom of the electrode and reagent tip should be slightly above the stir bar—they should not come into contact with it.

5.

Start the software – ClAutotitrator.

6.

Log in using LIMS username/password.

7.

Select the sample to be measured by filling in Expedition, Site-Hole, Sample Name, and Sample (Figure 5).

8.

Click the measure button (Figure 6).

9.

While the instrument is running, the plot displays the current data set as well as instantaneous values for volume of reagent added, current value, run time, and current titrator status. No interaction is required.

10.

When the run is complete a sound will play and a window will pop up telling you that the run is finished (Figure 7). Click Save to upload the data to LIMS.

Figure 5. Enter Sample Information.

...

QAQC

IAPSO is used a standard for this analysis. The certified value, typically ~559 mM, is printed on the outside of each bottle. For quality control purposes, IAPSO shall be run as a sample as follows:
–Before every run
–As the last sample of any run
–Every 5 samples or 2 hours, whichever comes first.

...

Component

...

Alias

...

Unit

...

Description

...

chloride

...

Chloride

...

mM

...

Concentration of chloride

...

Titrant amount

...

Amount of titrant added to solution to achieve endpoint

Software

There is one piece of software associated with this instrument, the in-house: Cl Autotitrator piece. It is a WPF application written in C# against the 4.0 .NET runtime. The following table describes its characteristics.

...

Characteristic

...

Description

...

Supported Operating Systems

...

Windows 7 – 64 bit

...

Source Code

...

https://build.ship.iodp.tamu.edu/svn/wapps/Apps/ClTitrator/

...

The project is deployed as a Click Once application installable without administrative permissions. The install page is at: http://web.ship.iodp.tamu.edu/tasapps/cltitrator/

...

Web Services

...

This project makes calls to the Web Services (WAPPS). Specifically, it:

  • Obtains a list of new expeditions, sites, holes, sample names, and samples based on user-entered criteria
  • Creates new tests and results based on measured parameters

Health, Safety, and Environment

The primary safety concerns for this instrument and method relate to chemical handling and disposal.

Nitric Acid

2.2 M nitric acid is required by this method; if it is made in house then, handling of concentrated nitric acid is required. Nitric acid is an oxidizer and should not be brought into contact with any organic materials.

Nitric Acid, Concentrated

–Open and handle concentrated nitric acid in a working fume hood.
–When diluting, always add acid to water.
–Wear gloves and safety glasses at all times when handling concentrated nitric acid.
–Body contact: rinse immediately with copious amounts of water.
–Clothing contact: remove the clothing and rinse with water and/or acid neutralization solution.
–If a spill occurs apply an acid neutralization solution to the spill area.

Nitric Acid, Dilute (2.2 M)

–Store this solution in the non-organic acid section of a ventilated fume hood storage cabinet.
–This acid can be aliquotted outside of a fume hood.
–Always wear gloves worn when using this solution.
–Neutralize or flush with copious amounts of water for disposal of this solution or waste products.
–Follow the recommendations for the concentrated solution above if contact with skin or clothing.

Silver Nitrate Solution (0.10 M)

...

Calibration

It is recommended to ‘re-calibrate’ the system whenever the silver nitrate solution is replaced, if the electrode needs replacing, if you replace the dilution solution and your checks are not acceptable or during normal operation when once again your checks are not within acceptable limits.

1.Click on the SETTINGS button in the software window. Change the eluent molarity parameter to 0.015 (or whatever your concentration is thought to be).
2.Measure IAPSO five times and average the measurements.
3.

Update the eluent molarity field using: 

  • eluent molarity = (0.015 * 559) / averaged measured value

Note: 0.015 is the original AgNO3 concentration in this example. In practice, use the value listed in the "Eluent Molarity" field of the "Settings" dropdown menu when the five IAPSO measurements were performed. The concentrations of those measurements are proportional to the molarity the user initially input into the software.

4.Re-measure IAPSO. The values should be close to 559.

NOTE: when changing the AgNO3 solution, make sure that you flush all lines, with the new solution, before attempting to re-calibrate.

Analyzing Samples

1.

Place stir bar in bottom of reaction vessel.

2.

Pipette 100 µl of sample into the bottom of the reaction vessel.

3.

Pipette 30 ml of dilution solution into the bottom of the reaction vessel (the actual amount isn't important as long as the electrode is submerged).

4.

Insert the electrode and reagent dispensing tip into the reaction vessel.
Note: The bottom of the electrode and reagent tip should be slightly above the stir bar—they should not come into contact with it.

5.

Start the software – ClAutotitrator.

6.

Log in using LIMS username/password.

7.

Select the sample to be measured by filling in Expedition, Site-Hole, Sample Name, and Sample (Figure 5).

8.

Click the measure button (Figure 6).

9.

While the instrument is running, the plot displays the current data set as well as instantaneous values for volume of reagent added, current value, run time, and current titrator status. No interaction is required.

10.

When the run is complete a sound will play and a window will pop up telling you that the run is finished (Figure 7). Click Save to upload the data to LIMS.


Image Added

Figure 5. Enter Sample Information.

Image Added

Figure 6. Measure Sample.
Image Added
Figure 7. Run Complete.

QAQC

IAPSO is used a standard for this analysis. The certified value, typically ~559 mM, is printed on the outside of each bottle. For quality control purposes, IAPSO shall be run as a sample as follows:
–Before every run
–As the last sample of any run
–Every 5 samples or 2 hours, whichever comes first.

Data Available in LORE

Interstitial Waters Standard Report 

  • Exp: expedition number
  • Site: site number
  • Hole: hole number
  • Core: core number
  • Type: type indicates the coring tool used to recover the core (typical types are F, H, R, X).
  • Sect: section number
  • A/W: archive (A) or working (W) section half.
  • Top offset on section (cm): position of the upper edge of the sample, measured relative to the top of the section.
  • Bottom offset on section (cm): position of the lower edge of the sample, measured relative to the top of the section.
  • Top depth CSF-A (m): position of observation expressed relative to the top of the hole.
  • Top depth [other] (m): position of observation expressed relative to the top of the hole. The location is presented in a scale selected by the science party or the report user.
  • Sampling tool: tool used to collect sample
  • Data columns: header lists parameter measured and concentration units, followed by wavelength (for ICP-AES) and then analysis method.

Expanded TITRA_MAN Report

  • Exp: expedition number
  • Site: site number
  • Hole: hole number
  • Core: core number
  • Type: type indicates the coring tool used to recover the core (typical types are F, H, R, X).
  • Sect: section number
  • A/W: archive (A) or working (W) section half.
  • text_id: full text ID of sample
  • sample_number: sample number of sample. text ID with sample type prefix removed.
  • label_id: id combining exp, site, hole, core, type, sect, A/W, parent sample name (if any), sample name
  • sample_name: name of sample
  • x_sample_state:
  • x_project: expedition project the sample is uploaded under. typically the same as Exp.
  • x_capt_loc: 
  • location: location sample was taken
  • x_sampling_tool: tool used to collect sample
  • changed_by: name of person who uploaded sample
  • changed_on: date and time sample was uploaded
  • sample_type: type of sample. typically LIQ, for liquid.
  • x_offset: top offset of parent sample where sample was taken in m
  • x_offset_cm: top offset of parent sample where sample was taken in cm
  • x_bottom_offset_cm: bottom offset of parent sample where sample was taken in cm
  • x_diameter: 
  • x_idmp: 
  • x_orig_len: 
  • x_length: length of sample in m
  • x_lengeth_cm:  length of sample in cm
  • status:
  • old_status:
  • original_sample: 
  • parent_sample:
  • standard:
  • login_by: name of person logged into LIMS application used for this test
  • sampled_date: 
  • legacy:
  • test changed_on: date of last edit of analysis
  • test date_started: date analysis was started
  • test group_name: 
  • test status:
  • test old_status:
  • test test_number: number assigned to performed analysis on sample
  • test date_received: date analysis was uploaded to LIMS
  • test instrument: instrument used to perform analysis
  • test analysis: analysis type
  • test x_project: project test was assigned to
  • test version: 
  • test order_number:
  • test replicate_test:
  • test replicate_count:
  • rest sample_number: sample number for sample the analysis was performed on
  • Top depth CSF-A (m): position of observation expressed relative to the top of the hole.

  • Bottom depth CSF-A (m): position of observation expressed relative to the top of the hole.

  • Top depth CSF-B (m):

  • Bottom depth CSF-B (m):

  • ammonium (mM): concentration of ammonium 
  • calcium (mM): concentration of calcium
  • chloride (mM): concentration of chloride in mM
  • dilution_factor: the amount the sample was diluted by
  • magnesium (mM): concentration of magnesium
  • phosphate (uM): concentration of phosphate
  • ssup_asman_id:
  • ssup_filename:
  • sulfide (uM): concentration of sulfide
  • titrant_amount (mL): 
  • sample description: description/comment of sample

  • test test_comment: comment added for performed analysis

  • result comments: comment added for analysis results


Software

There is one piece of software associated with this instrument, the in-house: Cl Autotitrator piece. It is a WPF application written in C# against the 4.0 .NET runtime. The following table describes its characteristics.

Characteristic

Description

Supported Operating Systems

Windows 7 – 64 bit

Source Code

https://build.ship.iodp.tamu.edu/svn/wapps/Apps/ClTitrator/

Deployment

The project is deployed as a Click Once application installable without administrative permissions. The install page is at: http://web.ship.iodp.tamu.edu/tasapps/cltitrator/

Web Services

This project makes calls to the Web Services (WAPPS). Specifically, it:

  • Obtains a list of new expeditions, sites, holes, sample names, and samples based on user-entered criteria
  • Creates new tests and results based on measured parameters


Health, Safety, and Environment

The primary safety concerns for this instrument and method relate to chemical handling and disposal.

Nitric Acid

2.2 M nitric acid is required by this method; if it is made in house then, handling of concentrated nitric acid is required. Nitric acid is an oxidizer and should not be brought into contact with any organic materials.

Nitric Acid, Concentrated

–Open and handle concentrated nitric acid in a working fume hood.
–When diluting, always add acid to water.
–Wear gloves and safety glasses at all times when handling concentrated nitric acid.
–Body contact: rinse immediately with copious amounts of water.
–Clothing contact: remove the clothing and rinse with water and/or acid neutralization solution.
–If a spill occurs apply an acid neutralization solution to the spill area.

Nitric Acid, Dilute (2.2 M)

–Store this solution in the non-organic acid section of a ventilated fume hood storage cabinet.
–This acid can be aliquotted outside of a fume hood.
–Always wear gloves worn when using this solution.
–Neutralize or flush with copious amounts of water for disposal of this solution or waste products.
–Follow the recommendations for the concentrated solution above if contact with skin or clothing.

Silver Nitrate Solution (0.10 M)

The primary hazard from silver nitrate solution is contact with body parts. It forms a silver compound and turns black, staining the skin and other membranes until they are naturally replaced by the body (typically 1–3 weeks). Certain individuals may also exhibit redness, swelling, and rash-like symptoms in the afflicted area. Of particular concern is getting the solution in the eye. Care should be taken to avoid this.
–Use gloves and eye protection when transferring the stock solution to the reagent bottle.
–Use gloves when handling the waste solution—it contains a small amount of unreacted silver.

LIMS Component Table

Although the titrator could theoretically be used to titrate a number of constituents (e.g., Ca2+, S2-), it is in practice only used for chlorinity. Nonetheless, the LIMS component table includes those other analytes should they ever be done by titration.

ANALYSISTABLENAMEABOUT TEXT
TITRA_MANSAMPLEExpExp: expedition number
TITRA_MANSAMPLESiteSite: site number
TITRA_MANSAMPLEHoleHole: hole number
TITRA_MANSAMPLECoreCore: core number
TITRA_MANSAMPLETypeType: type indicates the coring tool used to recover the core (typical types are F, H, R, X).
TITRA_MANSAMPLESectSect: section number
TITRA_MANSAMPLEA/WA/W: archive (A) or working (W) section half.
TITRA_MANSAMPLEtext_idText_ID: automatically generated database identifier for a sample, also carried on the printed labels. This identifier is guaranteed to be unique across all samples.
TITRA_MANSAMPLEsample_numberSample Number: automatically generated database identifier for a sample. This is the primary key of the SAMPLE table.
TITRA_MANSAMPLElabel_idLabel identifier: automatically generated, human readable name for a sample that is printed on labels. This name is not guaranteed unique across all samples.
TITRA_MANSAMPLEsample_nameSample name: short name that may be specified for a sample. You can use an advanced filter to narrow your search by this parameter.
TITRA_MANSAMPLEx_sample_stateSample state: Single-character identifier always set to "W" for samples; standards can vary.
TITRA_MANSAMPLEx_projectProject: similar in scope to the expedition number, the difference being that the project is the current cruise, whereas expedition could refer to material/results obtained on previous cruises
TITRA_MANSAMPLEx_capt_locCaptured location: "captured location," this field is usually null and is unnecessary because any sample captured on the JR has a sample_number ending in 1, and GCR ending in 2
TITRA_MANSAMPLElocationLocation: location that sample was taken; this field is usually null and is unnecessary because any sample captured on the JR has a sample_number ending in 1, and GCR ending in 2
TITRA_MANSAMPLEx_sampling_toolSampling tool: sampling tool used to take the sample (e.g., syringe, spatula)
TITRA_MANSAMPLEchanged_byChanged by: username of account used to make a change to a sample record
TITRA_MANSAMPLEchanged_onChanged on: date/time stamp for change made to a sample record
TITRA_MANSAMPLEsample_typeSample type: type of sample from a predefined list (e.g., HOLE, CORE, LIQ)
TITRA_MANSAMPLEx_offsetOffset (m): top offset of sample from top of parent sample, expressed in meters.
TITRA_MANSAMPLEx_offset_cmOffset (cm): top offset of sample from top of parent sample, expressed in centimeters. This is a calculated field (offset, converted to cm)
TITRA_MANSAMPLEx_bottom_offset_cmBottom offset (cm): bottom offset of sample from top of parent sample, expressed in centimeters. This is a calculated field (offset + length, converted to cm)
TITRA_MANSAMPLEx_diameterDiameter (cm): diameter of sample, usually applied only to CORE, SECT, SHLF, and WRND samples; however this field is null on both Exp. 390 and 393, so it is no longer populated by Sample Master
TITRA_MANSAMPLEx_orig_lenOriginal length (m): field for the original length of a sample; not always (or reliably) populated
TITRA_MANSAMPLEx_lengthLength (m): field for the length of a sample [as entered upon creation]
TITRA_MANSAMPLEx_length_cmLength (cm): field for the length of a sample. This is a calculated field (length, converted to cm).
TITRA_MANSAMPLEstatusStatus: single-character code for the current status of a sample (e.g., active, canceled)
TITRA_MANSAMPLEold_statusOld status: single-character code for the previous status of a sample; used by the LIME program to restore a canceled sample
TITRA_MANSAMPLEoriginal_sampleOriginal sample: field tying a sample below the CORE level to its parent HOLE sample
TITRA_MANSAMPLEparent_sampleParent sample: the sample from which this sample was taken (e.g., for PWDR samples, this might be a SHLF or possibly another PWDR)
TITRA_MANSAMPLEstandardStandard: T/F field to differentiate between samples (standard=F) and QAQC standards (standard=T)
TITRA_MANSAMPLElogin_byLogin by: username of account used to create the sample (can be the LIMS itself [e.g., SHLFs created when a SECT is created])
TITRA_MANSAMPLElogin_dateLogin date: creation date of the sample
TITRA_MANSAMPLElegacyLegacy flag: T/F indicator for when a sample is from a previous expedition and is locked/uneditable on this expedition
TITRA_MANTESTtest changed_onTEST changed on: date/time stamp for a change to a test record.
TITRA_MANTESTtest statusTEST status: single-character code for the current status of a test (e.g., active, in process, canceled)
TITRA_MANTESTtest old_statusTEST old status: single-character code for the previous status of a test; used by the LIME program to restore a canceled test
TITRA_MANTESTtest test_numberTEST test number: automatically generated database identifier for a test record. This is the primary key of the TEST table.
TITRA_MANTESTtest date_receivedTEST date received: date/time stamp for the creation of the test record.
TITRA_MANTESTtest instrumentTEST instrument [instrument group]: field that describes the instrument group (most often this applies to loggers with multiple sensors); often obscure (e.g., user_input)
TITRA_MANTESTtest analysisTEST analysis: analysis code associated with this test (foreign key to the ANALYSIS table)
TITRA_MANTESTtest x_projectTEST project: similar in scope to the expedition number, the difference being that the project is the current cruise, whereas expedition could refer to material/results obtained on previous cruises
TITRA_MANTESTtest sample_numberTEST sample number: the sample_number of the sample to which this test record is attached; a foreign key to the SAMPLE table
TITRA_MANCALCULATEDTop depth CSF-A (m)Top depth CSF-A (m): position of observation expressed relative to the top of the hole.
TITRA_MANCALCULATEDBottom depth CSF-A (m)Bottom depth CSF-A (m): position of observation expressed relative to the top of the hole.
TITRA_MANCALCULATEDTop depth CSF-B (m)Top depth [other] (m): position of observation expressed relative to the top of the hole. The location is presented in a scale selected by the science party or the report user.
TITRA_MANCALCULATEDBottom depth CSF-B (m)Bottom depth [other] (m): position of observation expressed relative to the top of the hole. The location is presented in a scale selected by the science party or the report user.
TITRA_MANRESULTammonium (mM)RESULT ammonium (mM): ammonium (NH4+) concentration by titration
TITRA_MANRESULTcalcium (mM)RESULT calcium (mM): calcium (Ca2+) concentration by titration
TITRA_MANRESULTchloride (mM)RESULT chloride (mM): chloride (Cl-) concentration by titration
TITRA_MANRESULTdilution_factorRESULT dilution factor: dilution factor for the IW sample (e.g., 50 would mean 1:50)
TITRA_MANRESULTmagnesium (mM)RESULT magnesium (mM): magnesium (Mg2+) concentration by titration
TITRA_MANRESULTphosphate (µM)RESULT phosphare (uM): phosphate (PO43-) concentration by titration
TITRA_MANRESULTssup_asman_idRESULT spreadsheet uploader ASMAN_ID: serial number of ASMAN link for the spreadsheet uploader file
TITRA_MANRESULTssup_filenameRESULT spreadsheet uploader filename: file name for the spreadsheet uploader file
TITRA_MANRESULTsulfide (µM)RESULT sulfide (uM): sulfide (S2-) concentration by titration
TITRA_MANRESULTtitrant_amount (mL)RESULT titrant amount (mL): number of milliliters of the titrant used in the experiment
TITRA_MANSAMPLEsample descriptionSAMPLE comment: contents of the SAMPLE.description field, usually shown on reports as "Sample comments"
TITRA_MANTESTtest test_commentTEST comment: contents of the TEST.comment field, usually shown on reports as "Test comments"
TITRA_MANRESULTresult commentsRESULT comment: contents of a result parameter with name = "comment," usually shown on reports as "Result comments"