START button will allow the user to begin measurements. Section Information window (Figure 4) will pop up.
Figure 4.
...
Section Information window.
B. Instrument Calibration
...
- Select the tab labeled DARK-1 (Figure 5).
- The default exposure values are automatically displayed and are the recommend values, but you may set other values. Be aware that the dark saturation (image intensity saturation) must be 7% or lower and you need to capture at least 100 images for stacking.
- Default Recommended values are Image stack: 10020, kV: 8060, mA: 0.73, Integration Time: 1 ms
- Click the Take DARK Image button.
- The X-Ray Image Acquisition window will open (Figure 7). During acquisition, the image window appears black but you will see values in the histogram and in the image line profile.
...
- Select the tab labeled WHITE-2 (Figure 8).
- The default exposure values are automatically displayed and are the recommend values, but you may set other values. Just be aware that the maximum intensity cannot exceed 99%.
- Default Recommended values are : Image stack : = 20, kV : 112=80, mA: =1.0, Integration Time : 12 ms= 30
- Click the Take WHITE Image button.
- The X-Ray Image Acquisition window will open (Figure 9).
- During acquisition, the image window appears white shading to darker grey on the left and right edges.
Once acquisition is completed, you will be return to the FPD Calibration Utility window.
3. Check the maximum saturation value and confirm that it is in the target range of 94-98%. If not, adjust the exposure values and repeat (Figure 8).
...
Note: All core sections entering the NGR XMSL must be completely dry. Any moisture or liquid present will DAMAGE the instrument.
...
A.2 Maintenance and Troubleshooting
Trouble Shooting
Issue | Solution |
---|---|
An error message occurs when the user opens the software indicating an invalid calibration for the dark and white images, but the dark and white images are available on the computer. | Verify that the .ini files are referring to valid calibration images that exist in the correct locations. If the files exist, then check the USB connections to the computer, particularly the USB for the X-ray source. |
B.1 IMS Program Structure
...
The following table shows the motion profile settings for the XMSL.
Axis→ | X-Axis | ||
---|---|---|---|
Profile↓ | Speed | Accel. | Decel. |
DAQ Move | 5.0 | 20.0 | 30.0 |
Limit Seek | 5.0 | 1.0 | 80.0 |
Home Final | 3.0 | 2.0 | 80.0 |
Load/Unload | 5.0 | 5.0 | 10.0 |
Push-Slow | 5.0 | 5.0 | 100.0 |
Push-Fast | 5.0 | 5.0 | 50.0 |
User Defined | 5.0 | 5.0 | 5.0 |
Table: Motion Profile X-Axis for XMSL
...
C.2 System Settings
System Settings
Hostname | PC53236 |
DNS Name | |
Vendor | HP |
Model | HP Z240 SFF Workstation |
Serial Number | 2UA7121YXV |
Firmware Version | N51 Ver. 01.52 |
Hardware Revision | |
Operating System | Microsoft Windows 10 Enterprise |
System Start Time | 3/4/2019 7:24:09 PM |
Description | PC53236 - 64 bit Workstationt Image |
System Configuration Web Access | Local Only |
System Resources
Total Physical Memory | 31.8 GB |
Free Physical Memory | 26.2 GB |
Total Virtual Memory | 36.5 GB |
Free Virtual Memory | 30.7 GB |
Primary Disk Capacity | 476 GB |
Primary Disk Free Space | 130 GB |
CPU Model | Intel(R) Core(TM) i7-6700 CPU @ 3.40GHz |
Teledyne DALSA GoScan 1510 HR "1727014677" |
Settings
Name | 1727014677 |
Hostname | 169.254.5.141 |
Vendor | Teledyne DALSA |
Model | Teledyne DALSA GoScan 1510 HR |
Serial Number | DC60A1378609 |
Attribute Categories
AcquisitionAttributes | |
Property | Value |
BitsPerPixel | Use hardware value |
ImageDecoderCopyMode | Auto |
IncompleteBufferMode | Ignore |
OutputImageType | Grayscale (U16) |
OverwriteMode | Get Newest |
PacketSize | 9004 |
ReceiveTimestampMode | System Time |
ShiftPixelBits | false |
SwapPixelBytes | false |
Timeout | 5000 |
AcquisitionAttributes AdvancedEthernet BandwidthControl | |
Property | Value |
ActualPeakBandwidth | 1000.00000 |
DesiredPeakBandwidth | 1000.00000 |
AcquisitionAttributes AdvancedEthernet Controller | |
Property | Value |
DestinationMode | Unicast |
DestinationMulticastAddress | 239.192.0.1 |
AcquisitionAttributes AdvancedEthernet EventParameters | |
Property | Value |
MaxOutstandingEvents | 50 |
AcquisitionAttributes AdvancedEthernet FirewallTraversal | |
Property | Value |
Enabled | true |
KeepAliveTime | 30 |
Supported | true |
AcquisitionAttributes AdvancedEthernet ResendParameters | |
Property | Value |
MaxResendsPerPacket | 25 |
MemoryWindowSize | 1024 |
MissingPacketTimeout | 2 |
NewPacketTimeout | 100 |
ResendBatchingPercentage | 10 |
ResendResponseTimeout | 2 |
ResendThresholdPercentage | 5 |
ResendTimerResolution | 1 |
ResendsEnabled | true |
AcquisitionAttributes AdvancedEthernet TestPacketParameters | |
Property | Value |
MaxTestPacketRetries | 1 |
TestPacketEnabled | true |
TestPacketTimeout | 250 |
AcquisitionAttributes AdvancedGenicam | |
Property | Value |
CommandTimeout | 100 |
EventsEnabled | true |
EventsSupported | true |
IgnoreCameraValidationErrors | false |
AcquisitionAttributes ChunkDataDecoding | |
Property | Value |
ChunkDataDecodingEnabled | true |
MaximumChunkCopySize | 64 |
CameraAttributes deviceInformation | |
Property | Value |
DeviceVendorName | Teledyne DALSA |
DeviceModelName | GoScan 1510 HR |
DeviceID | 1727014677 |
DeviceVersion | 209.21 |
HardwareRevision | IL1.0 |
TimingVersion | 12.119 |
deviceMacAddress | DC:60:A1:37:86:09 |
DeviceUserID | 1727014677 |
UserSetDefaultSelector | Factory Setting |
UserSetSelector | Factory Setting |
UserSetLoad | |
UserSetSave | |
DeviceTemperature | 50.0000000 |
CameraAttributes detectorControl | |
Property | Value |
SettingsRevision | 2 |
Standby | Off |
ReadOutMode | Full Resolution |
ReadOutDirectionY | Bottom to Top |
ReadOutDirectionX | Right to Left |
TriggerMode | Free Running |
SoftwareTrigger | |
SoftwareTrigIntTime | 50000 |
ExtendedExposure | 10000 |
Offset | 1.95% |
FullWell | High Full Well |
ROIStartH | 0 |
ROIStopH | 1031 |
ROIStartV | 0 |
ROIStopV | 1549 |
CameraAttributes Diagnose | |
Property | Value |
ImageSourceSelector | Sensor Data |
SensorDigitalTestPatternValue | 0 |
TestImageIncrement | 1 |
PowerCheck3V5DIG | false |
PowerCheck5V4 | false |
PowerCheck1V2 | false |
PowerCheck1V5 | false |
PowerCheck2V5 | false |
PowerCheck3V3 | false |
PowerCheck3V5ANA | false |
PowerCheck1V8 | false |
ResetHighReadbackAbs | 3.49500000 |
SelectHighReadbackAbs | 3.49375000 |
VrReadbackAbs | 3.19875000 |
NWellReadbackAbs | 3.49375000 |
AdcRefReadbackAbs | 2.79875000 |
AnaT2ReadbackAbs | 0.0387500000 |
PSupReadbackAbs | 11.4541485 |
Opamp5VReadbackAbs | 4.99875000 |
RefreshDiagnoseBits | |
transferQueueCurrentBlockCount | 0 |
CameraAttributes ImageFormatControl | |
Property | Value |
PixelPacking | Mono16 |
PixelFormat | Monochrome 16-Bit Packed |
PixelCoding | Mono |
PixelSize | 16 Bits/Pixel |
Width | 1032 |
Height | 1550 |
Mono12PackedBitSelection | LSB |
CameraAttributes ImageProcessingControl | |
Property | Value |
SUMEnable | Off |
SUMNofFrame | 1 Frame |
CameraAttributes >> TransportLayerControl | |
Property | Value |
deviceUPnPDiscoveryMode | Active |
GevLinkSpeed | 1000 |
devicePacketResendBufferSize | 12.9375000 |
GevIPConfigurationStatus | DHCP |
GevPrimaryApplicationIPAddress | 169.254.116.129 |
deviceCCP | Control Access |
GevDiscoveryAckDelay | |
GevGVCPHeartbeatDisable |
CameraAttributes >> FileAccessControl | |
Property | Value |
FileSelector | Firmware |
FileOperationSelector | Open |
FileOperationExecute | |
FileOpenMode | Write |
FileAccessOffset | |
FileAccessLength | 4 |
FileOperationStatus | Success |
FileOperationResult | 0 |
FileSize | 0 |
CameraInformation | |
Property | Value |
BaseAddress | 0x0 |
BusType | Ethernet |
HostIPAddress | 169.254.116.129 |
HostMACAddress | 00:0A:CD:2F:13:8C |
HostNetworkInterfaceName | {1E2FC758-4010-4217-A057-038CD47B344A} |
IPAddress | 169.254.5.141 |
ModelName | GoScan 1510 HR |
PrimaryURLString | Local:TeledyneDALSA_CD45M21_02_2457f43f_1CA40.0000.xml;b000;31a8e |
SecondaryURLString | http://www.teledynedalsa.com/public/mv/software/genicam/GenICamFile.xml |
SerialNumberHigh | 0xDC60 |
SerialNumberLow | 0xA1378609 |
VendorName | Teledyne DALSA |
StatusInformation | |
Property | Value |
AcqInProgress | false |
HandledEventCount | 0 |
IncompleteBufferCount | 0 |
LastBufferCount | 0 |
LastBufferNumber | 4294967292 |
LostBufferCount | 0 |
LostEventCount | 0 |
LostPacketCount | 0 |
ReceivedResendPackets | 0 |
RequestedResendPacketCount | 0 |
VI. Credits
All improvements to the Quick Start Guides and User Guides are a communal effort, with honorable mention to the group of LOs, ALOs, and technicians who have helped.