Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

...

...

...

...

Table of Contents

Introduction


The P-wave velocity gantry measures the speed at which ultrasonic sound waves pass through materials that are placed between its transducers; this includes section halfshalves, discrete samples, and mini-cores.
The three orthogonal sets of piezoelectric transducers allow the velocity to be determined in the X-, Y-, and Z-directions on working-half split-core sections. The P-wave bayonets (PWBs) measure the velocity along the core (Z-direction) and across the split-core face (Y-direction), and a P-wave caliper (PWC) measures the velocity perpendicular to the split-core face (X-direction) (Figure 1). 
For discrete sample cubes, the velocity is measured along each of the three axes separately using the PWC; whereas, mini. Mini-cores are measured along the axis of the cylinder (X-direction). For discrete samples cubes and mini-cores, all sample information, including offset, is entered by the user???.


Figure 1. Core section-half liner showing top of section and relative orientations.

...


   CAUTION: DO NOT place anything, including samples, on the track below the instruments before opening the program. The program will close lower all transducers upon startup, and CAN CAUSE SEVERE INJURY.

  1. Double-click the MUT icon on the desktop (Figure 1a2a) and login using ship credentials. For more information on data uploading see the "Uploading Data to LIMS" section below.
  2. Double-click the IMS icon (Figure 1b2b). IMS initialize's initializes the instrument and checks the movement of the the three sets of instruments. Once initialized, the logger is ready to measure the first sample.

Figure 12. - (a) MUT Icon. (b) IMS Icon.

...

  1. Make sure the sensors are clean.
  2. For section halfshalves, place the desired measurement location underneath the desired sensor. Make sure that the section is oriented with the blue end-cap towards the laser (Figure 23).


Figure 23. - Correct core position.

3. Place a small piece of Glad Wrap at the measurement location to keep the sensors caliper sensor clean and the core free of contamination (including water).

4. Place one or two drops of deionized (DI) water at any relevant at  sample/sensor interface. It will improve the signal transmission.

...

  1. Select the measurement type. Seven different measurement configurations are available (Figure 34). Currently measures to whole round sections are not available.


Image RemovedImage Added

Figure 34. Measurement Configuration Options


2. Depending on which measurement, insert the bayonets into or close the caliper onto the sample. Click the instrument motion buttons to slowly lower the transducers to the sample (Figure 45). The autoclose Alternatively, use auto close. Auto close button functions for either the caliper or the bayonets.


Figure 45. - GANTRY motion buttons.


Note: Insert the bayonets until the entire sensor (black dot) is within the sample. Stop before the bayonets make contact with the liner.

3. The program continually updates the velocity calculation and displays the value (Average of the sets of stacksstack set) in the Velocity-Auto box (Figure 56).


Figure 56. - Data display on Velocity-Auto box.


It Program also displays a signal graph with a red vertical auto-pick location line (Figure 67), where the program's algorithm has placed the first arrival wave.
Note: Typical slide bar settings are 100 for the stack bar and10mV for the threshold bar.

Figure 67. - Clean waveform indicating a strong/good signal.

...


Note: If the velocity is out of the expected range, or the pick location is not near to or at the first arrival wave, contact the PP technician.

...

6. Place the cursor in the SCAN field and scan the section label. SAMPLE ID and LIMS ID automatically fills. These ID tags contain information on expeditionThis ID tag contains sample information, site, hole, core, section number and length.

...

10. MUT is used to upload data to the LIMS database. See section "Uploading Data to LIMS" for more information on data upload.

11. Once uploaded, data is available to view on LORE and LIVE. If data is not instantly visible, please contact the PP technician.

...


The program will automatically calculate the velocity from its auto-picked first arrival wave, display the value (Figure 56) and the auto-pick location line (Figure 78). If either of these are not expected results, follow steps A, B and C.


Figure 78. - Raw Stacked signal display with red line indicating auto-pick location.

...

The signal graph has several different displaysdisplay options, all accessible by clicking on different tabs at the top. To zoom on the a graph, pause the program and click on the magnifying glass in the "Graph Tools" box, Figure 128.

To evaluate the 'auto-pick velocity', you need to look at two graphs.Auto-pick location line
The signal graph has several different displays, all accessible by clicking on different tabs at the top. Tip, you can zoom by pausing the program and clicking on the magnifying glass in the Graph Tools box, Figure 12. The first graph you need to examine is the stacked raw value ("RAW STACKED" tab), (Figure 11). Here you see the stacked measurements display with a vertical red line showing the location of the programs auto-pick of the first arrival wave. If the red line appears very near the first significant deviation from 0 on the Y-Axis, then the value is very likely reliable.
Figure 12. Raw Stacked signal display with red line indicating auto-pick location.

Uploading Data to LIMS

  1. Click the MUT Icon. Log in using database credentials.
  2. Once activated, the list of files from the C:\DATA\IN directory is displayed. Files are marked ready for upload by a green check mark (Figure 7).

Image Removed

Figure 7.- Main MUT screen

3. To manually upload files, check each file individually and clip upload. To automatically upload files, click on the Automatic Upload checkbox.

4. If files are marked by a purple question mark or red and white X icons, please contact a technician.
    Purple question mark: Cannot identify the file.

    Red and white X icons: Contains file errors.

5. Upon upload, data is moved to C:\DATA\Archive. If upload is unsuccessful, data is automatically moved to C:\DATA\Error. Please contact the PP technician is this occurs.

...

This document originated from Word document Gantry Velocity: Quick Start Guide Version 1.0; Version 372 written by A. de Loach (2017-12-16). Credits for subsequent changes to this document are given in the page history.

All improvements to the Quick Start Guides and User Guides are a communal effort, with honorable mention to the group of LOs, ALOs, and technicians who have helped.

...

LMUG-P-WaveQuickStartGuide: An exported PDF version of this wiki page as of 2020-10-28.

...

check the 'Autopick location line' (red line):

1. Go to RAW STACKED tab (Figure 8): Here you see the stacked measurements display. If the red line appears near the first significant deviation from 0 on the Y-Axis, then the value is

...

likely reliable.

...

 However, if the 'auto-pick location' is off, or the velocities don't seem right, you may need to adjust the

...

Threshold, which corresponds to amplitude on the Y-Axis.

...

It will move the 'auto-pick' location.

2. Go to

...

ABS ZOOM

...

tab (Figure

...

9)

...

: Here you see a zoomed window of the absolute (ABS) signal value.

...

 Verify that the 'auto-pick location' is at the zero crossing after the first significant wave above noise levels

...

(not at

...

the first

...

arrival peak), use the

...

Threshold slide bar

...

to adjust its location

...

. By changing the threshold value the 'auto-pick' location will change in steps and

...

the velocity will be recalculated.

...

 

3. Check on RAW STACKED or other raw displays to verify the position of the red line at the first arrival wave. If satisfied, save your data. If not, you can manually pick the first arrival.


Note: The program must be running for changes to apply.


Image Modified

Figure

...

9. ABS Zoom display with threshold slide bar and uncorrected auto-pick location.

...


B) Manual pick

...

  1. Pause the program.

...

  1. Select MANUAL PICK tab (Figure 10).
  2. Click on the crosshair tool on the far left side of the toolbox bar

...

  1. . Hover over the dotted yellow-green vertical

...

  1. pink line. When the icon changes, click and drag the pick line to the desired location and release.

...

  1. The Velocity –Manual

...

  1. box

...

  1. will be updated with the manually picked velocity. Both the 'auto-pick' and 'manual pick velocities' are saved simultaneously to the database

...

  1. upload.

...


Image Added

Figure

...

10. Manual Pick display with

...

pink manual pick location.


C) Running a Standard as a QAQC Check

...

You can run a standard to verify if the instrument is measuring correctly. There are three materials available as a standard

...

: Aluminum and acrylic are

...

used on the caliper, and a core liner with D.I. water

...

on the bayonets.

...

Expected velocities for each

...

standard: 
Aluminum 6295

...

m/

...

s (+/- 63 m/s)
Acrylic 2730

...

m/

...

s (+/- 27 m/s)
Water 1480

...

m/

...

s (+/- 7 m/s)

...

Typical allowable deviation is 1% for the caliper and 0.5% for the bayonets. There will also be differences based on temperature, especially for water and aluminum. If the standard values are out of this range, ask a technician to help calibrate the instrument.

...

Uploading Data

...

to LIMS

  1. Click the MUT Icon. Log in using database credentials.
  2. Once activated, the list of files from the C:\DATA\IN directory is displayed. Files are marked ready for upload by a green check mark (Figure 11).

Image Added

Figure 11. Main MUT screen

3. To manually upload files, check each file individually and clip upload. To automatically upload files, click on the Automatic Upload checkbox.

4. If files are marked by a purple question mark or red and white X icons, please contact a technician.
    Purple question mark: Cannot identify the file.

    Red and white X icons: Contains file errors.

5. Upon upload, data is moved to C:\DATA\Archive. If upload is unsuccessful, data is automatically moved to C:\DATA\Error. Please contact the PP technician is this occurs.


Credits

This document originated from Word document Gantry Velocity: Quick Start Guide Version 1.0; Version 372 written by A. de Loach (2017-12-16). Credits for subsequent changes to this document are given in the page history.

All improvements to the Quick Start Guides and User Guides are a communal effort, with honorable mention to the group of LOs, ALOs, and technicians who have helped.


Archived Versions

P-Wave Quick Start Guide (*.pdf) - 27Sept2022

LMUG-P-WaveQuickStartGuide: An exported PDF version of this wiki page as of 2020-10-28.

P-Wave Quick Start Guide v2020
PWV_QSG_372.docx: A Word document of the Gantry Velocity: Quick Start Guide