The CP120B x-ray source can be operated using IMS. The operator should be familiar with the x-ray source operation and safety procedures before beginning a survey. See the IMS X-ray user guide for further instructions on software and source operation. The survey meters are kept in the Laboratory Officers office on bridge deck.
Figure 1- Survey meter calibration information found on the survey meter
Use the attached forms and location diagrams to record your survey results.
Figure 2-Direction of radiation scatter when standard is centered in the beam path
a. Amber light is lit when the x-ray source is connected to power.
b. Green light is lit when the load and unload doors are closed, FPD access door is closed, and the emergency shut off switch is not engaged.
c. Red light is lit when x-rays are generated. The light will blink quickly as it powers on and then slowly once the kV and mA are stable.
If these lights are not working, do not conduct further tests until the issue has been resolved.
For the following tests, the area must be secured with only the test personnel in the space. Also, make sure that the Load/Unload and Emergency Off tests (steps 2 and 3) are successful before you start the Area Monitor tests because you may need to use the doors or X-Ray off switches to power down the X-Ray source should the Area Monitor fail to shut down the system!
For the following tests, power the X-ray source at minimal levels (80kV, .8mA).
2. Door switches: With the X-rays on, open the Load end door a few millimeters and verify that the X-Rays shut down. The vinyl lead shielding must continue to cover the gap! Repeat above for the other unload end door.
3. Emergency Off: With the X-rays on, press the Emergency X-Ray OFF button (found above the load end) and verify that the X-Rays shut down.
4. Area Monitor Alarm :
Remove the vinyl lead shielding as shown from the Load end (Figure 3), power the X-Rays and move to the unload end of the track and be prepared to open the Unload door.
Figure 3- Lead vinyl shielding removed for testing the area monitor alarm. The image on the left is the load end shielding and the image on the right is the unload end.
Location | Description | µR/hr | notes |
Starboad-Aft Survey Form - Standard should be placed with blue end centered on the detector and the remaining core toward the unload end of the x-ray system | |||
1 | Along the top seam |
|
|
2 | Along the door seam |
|
|
3 | At the end just under the keyboard countertop |
|
|
4 | Along the door seam |
|
|
5 | Along the join between the MS-GRA shield and door |
|
|
6 | Along the outer side corner |
|
|
7 | Center of side panel, 40 cm from the top |
|
|
8 | Along the outer side corner |
|
|
9 | Along the outer side corner |
|
|
10 | Along the outer top corner |
|
|
Port-Aft Survey Form- Standard should be placed with blue end centered on the detector and the remaining core toward the load end of the x-ray system | |||
11 | Along the join of the Source and MS-GRA shields |
|
|
12 | Center of side panel, 40 cm from the top |
|
|
13 | Along the outer top corner |
|
|
14 | Top panel |
|
|
15 | Along the outer side corner |
|
|
16 | Along the seam between the unload shield and the source shield |
|
|
17 | Center of side panel, 40 cm from the top |
|
|
18 | Along the door seam |
| |
19 | Along the top of the door | ||
20 | Center of center panel |
Port-Forward Survey Form- Standard should be place with blue end centered on the detector and the remaining core toward the load end of the x-ray system | ||||
21 | Along the door seam |
|
| |
22 | Along the outer side corner |
|
| |
23 | Center of side panel, 40 cm from the top |
|
| |
24 | Along the join between the MS-GRA shield and door |
|
| |
25 | Along the outer side corner |
|
| |
26 | At the center of the panel |
|
| |
27 | Along the outer side corner |
|
| |
28 | Along the center of the panel |
|
| |
Bottom Survey Form- These measurements are taken below the countertop. | ||||
29 | Center of the MS-GRA Shield |
|
| |
30 | Directly under the detector (Source Shield) |
|
| |
31 | Around the detector (Source Shield) |
|
|
Area Alarm Sensors and Safety Circuit Check
At The Area Detectors | |||
Location | Description | µR/hr | notes |
Strbd | Next to detector |
|
|
Port | Next to dector |
|
|
BKGND | X-Rays Off |
|
|
Location | kcps | notes | |
Area Monitor w/X-Rays OFF |
|
| |
Area Monitor w/X-Rays ON |
|
| |
High Alarm Setting |
|
| |
Low Alarm Setting |
|
|
Switch/Interlock | Passed |
Emergency X-Ray OFF switch |
|
Load Door Interlock |
|
Unload Door Interlock |
|
Area Monitor Interlock |
|
Survey Date:____________________________
Assoicated Expedition (optional):__________________________________________________________
Name of Surveyor:______________________________________________________________________
Surveyor’s Signature:____________________________________________________________________