CHNS Elemental Analysis Advanced User Guide
Author(s): | Kazu Fujine | |
Reviewer(s): | L. Brandt | |
Revised by: | D. Houpt | |
Management Approval (Name, Title, Date) | D.J. Houpt, Supervisor of Analytical Systems | |
Audience: | Research Specialists/Marine Laboratory Technicians | |
Origination Date: | 5/6/08 | |
Current Version: | Version 1.0 | 6/07/11 |
Revised: | 371T | July 2017 |
Domain: | Chemistry | |
Analysis: | CHNS Elemental Analysis |
Topic | See page… |
Introduction | |
Apparatus, Reagents, & Materials | |
Sample Flow | |
Starting the Instrument | |
Shutting Down the Instrument | |
Maintenance and Troubleshooting | |
Environment, Health, & Safety | |
Software User Guide |
CHNS is fundamental element analyses performed in the shipboard laboratories for IODP cruises. Fluctuations in the concentrations and/or content ratios of C, N, and S define the origin, depositional environment, and diagenetic alteration of source materials. Analysts have a few choices for sample preparation, instrument settings, and measurement methodologies. Specific analytical methodology may be required based on the nature of certain sample materials. When this circumstance arises, laboratory specialists must authorize and document methodology deviations.
Before the cruise (during transit), confirm the existence and the condition of consumables, preparation tools, and chemicals for CHNS analysis on solid matrix.
–Helium (He) carrier gas: chromatographic grade
–Oxygen (O2) for sample oxidation: minimum 999.995%
–Agate mortar and pestle: well washed, dried
–Chamber for freeze drier: well washed, dried
–Desiccator and silica gel: use well-dried silica gel
–Kimwipes
–Weighing paper
–Mill (for hard rock sample)
–Sample holder (2): stainless steel, well washed, dried
–Spatula (laboratory spoon, 2): stainless steel, well washed, dried
–Spatula (micro spatula, 2)
–Spring tweezers (2)
–Tin containers (set of 100): Thermo Electron PN 240 064 00
–Tin discs (30 mm) (set of 100): Thermo Electron PN 252 080 15, optional
Material Name | Description | Purpose |
Aspartic acid (L-Aspartic acid) | C4H7NO4: white fine crystals | Standard reference material |
2.5-Bis(5-tert-butyl-benzoxazol-2- yl)thiophene (BBOT) | C26H26N2O2S: pale green crystals | Standard reference material |
Copper oxide | Pre-packed | Filling material |
Electrolytic copper | Pre-packed | Filling material |
Ethyl alcohol (ethanol) | C2H5OH | Solvent for sample preparation |
Quartz wool | SiO2 | Filling material |
Soil Reference Material | Light brown powder | Reference for nitrogen and carbon |
Sulfanilamide | C6H8N2SO2: white odorless crystals | Calibration standard for CHNS/CNS |
L-Cystine | C6H12N2O4S2: white odorless crystals | Calibration standard |
Vanadium pentoxide | V2O5: yellow to red crystalline powder | Catalyst |
Magnesium perchlorate | Mg(ClO4)2: white granulate | Filling material for water trap |
–Pre-packed reactor: Thermo Electron PN 468 020 15; also CE Elantech 061110
–Cleaning devices: Thermo Electron PN 205 006 25 and 276 060 10
–HPAR reactor extractor
–Crucible extractor
An expedition-specific science strategy or sample materials may require additional apparatus to handle liquid sample analysis.
–Automatic sampling system with injection assembly
–Water container and vials for solvent
–Syringe (0.5 µL): PN 36504045
–Syringe (10 µL): PN 36500525
–Syringe (250 µL): PN 013680 (36504042)
C, H, N, and S are analyzed on a Thermo Electron Corporation FlashEA 1112 CHNS elemental analyzer (Figure 1, below) with the following specifications:
–Supplier: Thermo Electron Corporation
–Model: FlashEA 1112
–Detector: Thermal conductivity detector (TCD)
–Instrument Control: Eager 300 software for MS Windows
–Power Supply: 230 V AC; 60 Hz (US); 140 VA
–Dimensions: 500 mm H × 590 mm W × 558 mm D
–Weight: 65 kg
–Measurement Range: 100 ppm to 100% for C, N, H, and S
–Nominal Measurement Time: CNS: 8 min, CHNS: 10 min
–Sample Weight: 0.01–100 mg according to nature of the sample
–Temperature range:15°–35°C Maximum rel. humidity:30%–85%
The basic CHNS instrument configuration contains the following components.
Figure 1. CHNS Components.
1. Autosampler (highlighted in red in Figure 2 and shown in detail in Figure 3)
2. Gas tubing for helium and oxygen
3. Synoptic panel
4. Furnace (furnace cover plate shown in Figure 4; opened plate in Figure 5)
5. Reactor
6. Oven
7. Water adsorption column (shown in Figure 6)
7. Gas chromatographic column (also in Figure 6)
8. TCD detector
Figure 2. Top and Front Instrument Panels. (Dry Sample Autosampler is located on top of furnace.)
Figure 3. Dry Sample Autosampler.
Figure 4. Furnace Compartment Access with Protection Plate.
WARNING! Do not open the furnace compartment during operation.
Figure 5. Interior Furnace Compartment.
WARNING: Do not touch either furnace when the furnace temperature is >40°C. (Select View > View Elemental Analyzer Status).
Figure 6. Detector Access: Protection Plate, GC Column (inside), Adsorption Filter.
WARNING: The GC column may only be serviced by the shipboard technicians.
Sample flow in the onboard laboratory normally proceeds as described in the following steps.
5. Upload sample list to Eager 300 software.
6. Edit sequence table.
7. Run test analysis and samples.
8. Output data summary (Excel file).
9. Upload data summary and files to LIMS.
Perform any required maintenance before starting the system (see Instrument Maintenance section).
Proceed according to the following operating sequence. Early start up (6 hr before measurement) is recommended for this instrument.
The sequence of data files must be saved in the folder with the method file. Create new folders in the following directory: My Computer\Local Disk\Program files\Thermo Electron\Eager 300 for EA1112\Expedition No.& abbreviationdata 001-200.
If there are more than 200 expected samples, create two or more data folders (e.g., data 201-400).
Figure 7. CHNS Screen
1. Select Edit > Edit Elemental Analyzer parameter.
2. Select the Flow / Timing tab (Figure 8).
3. Select (check) the Carrier, Oxygen, and Reference parameters.
4. Click Send to transfer the parameters.
5. Click OK to close window.
6. Click NO to exit without saving the method.
7. Select View > View Elemental Analyzer Status.
8. Select the Special Functions tab (Figure 9).
9. Click Leak Test to open the Leak Check screen.
10. Click Start to begin the operation.
11. Click Yes for the Eager 300 to automatically test for leaks.
12. After a few seconds, the reference gas flow must be 0.
Note: After ~560 sec, carrier flow will decrease. Carrier flow must be between 0 and 3 mL/min. Higher values indicate the system is not leak free. Leaks in the system are generally due to incorrect closure of the reactor and filter locking nut. Rarely, leaks may be due to the autosampler.
Figure 8. Flow/Timing Tab.
Figure 9. Special Functions Tab and Leak Test Screen.
If the analytical configuration changes from CHNS to CNS, select File > Instrument Configuration on the main screen before editing the method (see instructions in the FlashEA Operating Manual, p. 217–221).
Figure 10. Copy Method from Screen.
If a reactor is changed, the maintenance program must be reset. The maintenance program shown in Figure 11 gives an indication of reactor life. Reactor deterioration can also be detected on chromatograms.
Figure 11. Maintenance Program Screen
Reactor life is also influenced by the nature of samples, especially water content. So you may have to check for anomalies on the chromatogram on a regular basis, as follows:
4. Click Edit > Manual. Based on onshore tests, change the settings to
5. Click OK to save changes.
Select Edit > Edit Element Analyzer parameters. The following parameters will display:
–Temperature tab:
–Flow / Timing tab:
–Detector tab:
Note: Peak retention times are confirmed based on the chromatogram of the Bypass sample obtained using standard materials, which is then used to validate the calibration curve.
–Row 1:
–Row 2: Insert one blank line to stop measurement automatically after the blank.
–Row 3:
Note: Confirm chromatogram from the Blank to check the condition of the elemental analyzer. If the chromatogram has unexpected peaks and/or an unacceptable (disturbed) baseline, identify the cause of problem before running the Bypass sample. If the Blank is fine, continue measurement of Bypass.
10. Click Open to view the chromatogram.
11. Select Time of any component on the table.
12. Move the cursor near the red-gray arrow on the screen and right-click.
13. Adjust the red line on the screen to intersect the top of the peak.
14. Repeat Steps 11–13 for each peak on the chromatogram, and click OK to save changes.
Warning: If peak retention times have changed a lot, measure an additional Bypass sample to confirm the new peak positions. Possible causes of peak retention time shifts include (1) gas leaks (include outside of the instrument), (2) gas flow changes, (3) atmospheric pressure changes, and (4) others.
Figure 12. Sample Table.
Figure 13. Component Table.
Figure 14. Edit Method Screen.
Figure 15. Peak Detect Screen.
Select the Calculation parameters 4 tab on the Edit Method screen to select the calibration method. Eager 300 has three calibration methods, details of which are described in the FlashEA 1112 Operating Manual, p. 232–234. Generally, we use Linear fit for sediment samples. K-Factor is sometimes useful for samples that are characterized by low component concentrations.
Proceed to measuring samples, see the CHNS Analyzer User Guide.
At the end of the cruise, if needed, shut down the system following the procedure below:
Warning: The reactor must be installed with the furnaces at room temperature or <50°C with gloves. Do NOT use any mechanical tools to screw or unscrew the fixing nut.
Figure 16. Autosampler Removal.
See the FlashEA 1112 Operating Manual, p. 126–129 and 134. Pages 306–308 are also useful when replacing the O-ring of the coupling unions.
Figure 17. Furnace Reactor.
The autosampler does not normally require maintenance except for the sample tray. However, when the instrument is extensively used, it is a good practice to occasionally check the shaft housed in the autosampler.
Review the FlashEA 1112 Operating Manual, p. 309–313. Once per week or whenever the autosampler is dirty, clean the sample tray as follows:
If you detect a gas leak in the autosampler (see "Test for Leaks"), clean the shaft of the MAS 200 R. The cleaning method is shown in the FlashEA 1112 Operating Manual, p. 314–318.
Note: If you replace a shaft, you must send the old shaft back to vendor, CE Elantech, Inc., to maintain a shaft.
Warning: Wear gloves and face protector when preparing the adsorption filter. Work with quartz wool under the draft chamber because quartz wool is an inhalation hazard.
Offshore laboratory specialists must complete all mandatory training before sailing. In addition to the general laboratory safety training, offshore laboratory specialists must complete general user training at the onshore facility. The CHNS User Guide has more information on safety related to this system.
High-pressure gas lines, particularly the O2 line, can be hazardous if there is leak. Check for gas leaks at gas fittings at the instrument regularly.
The FlashEA 1112 has a safety cut-off function to prevent hazards. When an alarm condition is detected, this device cuts off the power to the heating elements of the oxidation and reduction furnaces. Instrument malfunctioning, because of component failure or abnormal operating conditions, is identified by the red safety cut-off LED on the synoptic panel. Reference: FlashEA1112 Operating Manual, p. 23–24.
Wear gloves and face protector when working with quartz wool. Also work under the draft chamber because quartz wool is an inhalation hazard. Vanadium pentoxide and magnesium persulfate are also significant hazards; be sure to read the material safety data sheet (MSDS) for each and every reagent and chemical, including the standards, used in this procedure!
The CHNS screen of the Eager 300 software application is the starting point to enter all functions.
–File: instrument operation functions:
–Run: start, stop, or abort sequences:
–Edit: instrument setup and analytical parameters:
–View: analysis and chromatogram functions
–Recalculation: cancel or recalculate calibration or results
–Tools: reactor maintenance functions
–Help