Author: | E. Fisher |
Reviewer(s): | W. Mills, TOP Team Tethys |
Editor(s): | |
Management Approval (Name, Title, Date): | W. Mills (Lab Officer) |
Audience: | Scientists, Technicians |
Previous Versions: | 1/22/2014, 8/15/2013 |
Current version: | V375 | V371T | July 2017 |
Revised: | T. Cobb, L. Crowder, C. Peng |
Domain: | |
System: | Section Half Image Logger |
Keywords: | Image, Section Half |
This guide is designed to help the scientists use the Section Half Image Logger (SHIL) to image core sections, upload the image data to LIMS, and retrieve image data from LIMS. The SHIL takes digital images of the flat face of a split section or the surface of whole round hard rock cores using a line scan camera.
***Please have a technician give you an overview before using this instrument***
*PLEASE DO NOT TOUCH THE CAMERA or LEAN ON THE TRACK*
Click the IMS-SHIL icon on the desktop (Figure 1) and wait for the program to Initialize. The process will include initializing the camera and moving the track into the "Home" position.
Figure 1. SHIL Desktop Icon
Click the START button on the main SHIL user interface (Figure 2).
Figure 2. Main SHIL User Interface Window
The Sample Information screen (Figure 3) will appear.
Figure 3. SHIL Section Information window
Make sure that you have set the images attributes in the panel on the left side:
Once you set these values they will default to the previous values until changed.
Scan section (half) barcode label, make sure cursor is in the SCAN box.
No scanner or label? Select the tab marked LIMS and select you section's ID from the list boxes or use the MANUAL tab and enter the values by hand (not recommended).
Confirm Label, LIMS and physical length are within 2 cm of each other. If not, ask WHY?
Maybe there was some section expansion, whole round sample taken, or a mistake??
By default the instrument is set for imaging the archive half and will not allow you to scan a working label. If you want to take a picture of a working half you need to go to the MANUAL tab and select W (working) into the Section Half label (Figure 4). Once you have selected W (working), you will not be able to scan an archive half; in order to do that you need to go back into the MANUAL tab and re-select A (archive).
Figure 4. SHIL MANUAL Section Information window
If a whole round sample was taken (IW or MBIO), place whole round spacer where this sample was taken.
Click or touch the TAKE A PICTURE (Figure 5) button to start imaging the section.
Figure 5. TAKE A PICTURE button
During the imaging process the Image DAQ screen (Figure 6) will display. If everything is going right, there is nothing for you to do. Only use ABORT to stop the imaging process for non-emergencies. If a body part has been caught in the track and you face imminent personal injury, use the EMERGENCY STOP button just above the track.
Figure 6. Image DAQ window
The window at the top shows the current image frame and the construction of the full image, and the graph at the bottom displays the RGB data taken from a stripe down the middle of the image.
If the DROP LINES indicator is lit (orange), notify the technical staff. The Acquisition Error indicator usually lights just at the end of the image and that’s OK but if it lights before then, we have a problem. Again, notify the technical staff.
Crop and Save
When the imaging process is done, the lights will go off and the track will return to the start position, so stand back.
The Image CROP widow (Figure 7) will open and you will have some work to do.
Figure 7. Image CROP window
Check the position of the GREEN crop lines by clicking the Show Top of Image and then Show Bottom of Image buttons. If you used the RED stops and the crop parameters have been set correctly, then the image should be perfectly boxed by the GREEN lines. If not, then use the mouse (or finger) and touch the GREEN line or corner and drag the lines into position. The goal is to create an image of just the core as one of three images that will be created (the other images are un-cropped). The cropped image is used in LimSpeak and for printing the VCD scratch sheets.
When happy with the crop box location, click/touch CROP Image (Figure 8). The display will change to show the cropped image and the Save Image button (Figure 9) will be enabled.
Figure 8. Crop Image button
Click Save Image and you are finished! Image files are stored in 'C:\Data\in', and the Section Information window will be displayed to repeat the process for the next section.
Figure 9. Save Image button
To rescan, click Discard Image and confirm when prompted to discard.
If the VCD auto print is enabled (Figure 10), a sheet with the current image will be automatically print. Core Describers can use this sheet as a base for their descriptions.
Figure 10. Set VCDS Parameters window
To upload the images into the database MegaUploadaTron must be running in background. If not already started do the following:
Figure 11. MUT icon
2. The following window (Figure 12) will appear:
Figure 12. LIMS Uploader Window
3. Check Automatic Upload in the lower right hand corner and then click the minimize window and MUT will run in background. The uploaded data will be moved from 'C:\Data\in' to 'C:\Data\archive'
Retrieving and Viewing Images from LIMS
The SHIL also can be used to image the external surface of whole round hard rock cores in order to assemble a 360° composite image of the whole round. Using the custom whole round scanning tray, the whole round core surface is imaged four times at the 0°, 90°, 180°, and 270° orientation from the splitting line. The Imaging Specialist will download the images and assemble a composite image from the four scans and upload the composite separately.
Figure 13. Alignment blocks position for whole round imaging