- Two types of sample racks, can be used in combination
- Locate the barcode on the side of the sample rack. This should be your first rack, so the instrument knows where to "start" the measurement
- Barcode should be facing the right hand side, when operator is facing the counter. The first rack should go in the right side of the instrument closer to the back
Power switch
sSample rack direction
barcode
Sample loading direction
To Run Sample/Standard
- 5 Sub-windows show up in the Winspecial main window
Easy GLPProtocol GLP Protocol for standard reference measurement
Control Live To Live To view Live signal
User BiolabEasy Biolab Easy count is default method for unknown sample
Latest Resultshow Result Show latest result
Fine TuningSelect Tuning Select spectra to plot or store
- Standards used in Perkin Elmer Unquenched LSC-std, Loading order is 3H, 14C, and blank
- Out standard have the following strength. This information is entered in Easy GLP
3H196600 2001/7/01 halfH 196600 2001/7/01 half-life is 12.35 yrs
14C101000 2001/7/01 halfC 101000 2001/7/01 half-life is 5730 yrs
- Instrument will calculate base on the half-life and give you the current strength in comparison to the measurement result.
- Easy counter cam measure 3H, 14C, 125I, 32P, 35S, and 45Ca. We only carry the first 2 standards.
- Easy counter test value, please refer to software manual P67.
...