Liquid Scintillation Counter User Guide
Wallac 1414 Scintillation Counter
Setup
Computer/Laptop is connected through comp port #2 on the counter
Startup
Verify "Wallac 1414 WinSpecial Ins.Prog "3.5" disc is inserted in the front port
Log on to the computer using daq/daq
Open "1414 Winspecial" software on the computer
Turn on the scintillation counter, the power switch is on the back
Winspecial main window status line should show "Loading" > "counter is busy"> "counter online"
Now the counter is ready
Sample Loading
Two types of sample racks, can be used in combination
Locate the barcode on the side of the sample rack. This should be your first rack, so the instrument knows where to "start" the measurement
Barcode should be facing the right hand side, when operator is facing the counter. The first rack should go in the right side of the instrument closer to the back
Place an empty rack or a "stop" rack after your samples, the counter will stop all the measurement after it detect the empty/stop rack
To Run Sample/Standard
5 Sub-windows show up in the Winspecial main window in minimized windows. To open the window, select "restore"
Easy GLP Protocol for standard reference measurement
Control Live To view live signal
User Biolab Easy Count is default method for unknown sample
Latest result show latest result
FineTuning Select spectra to plot or store
Standards used in Perkin Elmer Unquenched LSC-std, Loading order is 3H, 14C, and Blank.
Our standards have the following strength. This information is entered in Easy GLP
3H 196600 2001/7/01 half life 12.35yrs
14C 101000 2001/7/01 half life 5730 yrs
Instrument will calculate base on the half-life and give you the current strength in comparison to the measurement result.
Easy counter cam measure 3H, 14C, 125I, 32P, 35S, and 45Ca. We only carry the first 2 standards.
Easy counter test value, please refer to software manual P67.
Result
Once the measurement is finished, software will ask if user want to print result. The Laptop is setup to print directly to the Fdeck copier (via TCPIP). In 1414 Winspecial software, select file/print/setting and select Fdeck hall-color printer. (iprint connection resulted in error message and tend to shut down the software. see MCS if need to use iprint)
User can also export the data to a USB dri
Safety
Store all the standards in the grey plastic boxes when they are not being used.
Wash hands with soap after touching the standards and other potential contaminated samples.
Always follow Microbiology Van user protocol.
Credits
This document originated from Word document Liquid Scintillation Counter Quick User Guide (see Archived Versions below for a pdf copy) that was written by Chieh Peng (Sept - Oct 2010) and edited by Chieh Peng (2018-12-23). Credits for subsequent changes to this document are given in the page history.
Archives
Liquid Scintillation Counter Quick User Guide
Liquid Scintillation Counter User Guide 23022020