Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

CHNS Elemental Analysis Advanced User Guide

Manual Information


Author(s):

Kazu Fujine

Reviewer(s):

L. Brandt

Revised by:

D. Houpt

Management Approval (Name, Title, Date)

D.J. Houpt, Supervisor of Analytical Systems

Audience:

Research Specialists/Marine Laboratory Technicians

Origination Date:

5/6/08

Current Version:

Version 1.0

6/07/11

Revised:

371T

July 2017

Domain:

Chemistry

Analysis:

CHNS Elemental Analysis

Contents


Topic

See page…

Introduction

Apparatus, Reagents, & Materials

Sample Flow

Starting the Instrument

Shutting Down the Instrument

Maintenance and Troubleshooting

Environment, Health, & Safety

Software User Guide

Anchor
_Ref376954122
_Ref376954122
Introduction

...

Before the cruise (during transit), confirm the existence and the condition of consumables, preparation tools, and chemicals for CHNS analysis on solid matrix.

Gases

–Helium (He) carrier gas: chromatographic grade
–Oxygen (O2) for sample oxidation: minimum 999.995%

...

–Agate mortar and pestle: well washed, dried
–Chamber for freeze drier: well washed, dried
–Desiccator and silica gel: use well-dried silica gel
–Kimwipes
–Weighing paper
–Mill (for hard rock sample)
–Sample holder (2): stainless steel, well washed, dried
–Spatula (laboratory spoon, 2): stainless steel, well washed, dried
–Spatula (micro spatula, 2)
–Spring tweezers (2)
–Tin containers (set of 100): Thermo Electron PN 240 064 00
–Tin discs (30 mm) (set of 100): Thermo Electron PN 252 080 15, optional

Chemicals

Material Name

Description

Purpose

Aspartic acid (L-Aspartic acid)

C4H7NO4: white fine crystals

Standard reference material

2.5-Bis(5-tert-butyl-benzoxazol-2- yl)thiophene (BBOT)

C26H26N2O2S: pale green crystals

Standard reference material

Copper oxide

Pre-packed

Filling material

Electrolytic copper

Pre-packed

Filling material

Ethyl alcohol (ethanol)

C2H5OH

Solvent for sample preparation

Quartz wool

SiO2

Filling material

Soil Reference Material

Light brown powder

Reference for nitrogen and carbon

Sulfanilamide

C6H8N2SO2: white odorless crystals

Calibration standard for CHNS/CNS

L-Cystine

C6H12N2O4S2: white odorless crystals

Calibration standard

Vanadium pentoxide

V2O5: yellow to red crystalline powder

Catalyst

Magnesium perchlorate

Mg(ClO4)2: white granulate

Filling material for water trap

Other Consumables and Tools

...

C, H, N, and S are analyzed on a Thermo Electron Corporation FlashEA 1112 CHNS elemental analyzer (Figure 1, below) with the following specifications:
–Supplier: Thermo Electron Corporation
–Model: FlashEA 1112
–Detector: Thermal conductivity detector (TCD)
–Instrument Control: Eager 300 software for MS Windows
–Power Supply: 230 V AC; 60 Hz (US); 140 VA
–Dimensions: 500 mm H × 590 mm W × 558 mm D
–Weight: 65 kg
–Measurement Range: 100 ppm to 100% for C, N, H, and S
–Nominal Measurement Time: CNS: 8 min, CHNS: 10 min
–Sample Weight: 0.01–100 mg according to nature of the sample
–Temperature range:15°–35°C Maximum rel. humidity:30%–85%
The basic CHNS instrument configuration contains the following components.

Figure 1. CHNS Components.
1. Autosampler (highlighted in red in Figure 2 and shown in detail in Figure 3)
2. Gas tubing for helium and oxygen
3. Synoptic panel
4. Furnace (furnace cover plate shown in Figure 4; opened plate in Figure 5)
5. Reactor
6. Oven
7. Water adsorption column (shown in Figure 6)
7. Gas chromatographic column (also in Figure 6)
8. TCD detector


Image Modified
Figure 2. Top and Front Instrument Panels. (Dry Sample Autosampler is located on top of furnace.)
Image Modified
Figure 3. Dry Sample Autosampler.
Image Modified
Figure 4. Furnace Compartment Access with Protection Plate.
WARNING! Do not open the furnace compartment during operation.
Image Modified
Figure 5. Interior Furnace Compartment.
WARNING: Do not touch either furnace when the furnace temperature is >40°C. (Select View > View Elemental Analyzer Status).
Image Modified
Figure 6. Detector Access: Protection Plate, GC Column (inside), Adsorption Filter.
WARNING: The GC column may only be serviced by the shipboard technicians.

...

Sample Flow

Sample flow in the onboard laboratory normally proceeds as described in the following steps.

...

8. Output data summary (Excel file).
9. Upload data summary and files to LIMS.

...

...

Starting the Instrument

Perform any required maintenance before starting the system (see Instrument Maintenance section).

Powering on the System

...

  1. Double-click on the Eager 300 for EA 1112 icon.
  2. Select CHNS from the list and click OK to answers prompted step by step (Figure 7).

Image Modified
Figure 7. CHNS Screen

Checking for Leaks

1. Select Edit > Edit Elemental Analyzer parameter.
2. Select the Flow / Timing tab (Figure 8).
3. Select (check) the Carrier, Oxygen, and Reference parameters.
4. Click Send to transfer the parameters.
5. Click OK to close window.
6. Click NO to exit without saving the method.
7. Select View > View Elemental Analyzer Status.
8. Select the Special Functions tab (Figure 9).
9. Click Leak Test to open the Leak Check screen.
10. Click Start to begin the operation.
11. Click Yes for the Eager 300 to automatically test for leaks.
12. After a few seconds, the reference gas flow must be 0.
Note: After ~560 sec, carrier flow will decrease. Carrier flow must be between 0 and 3 mL/min. Higher values indicate the system is not leak free. Leaks in the system are generally due to incorrect closure of the reactor and filter locking nut. Rarely, leaks may be due to the autosampler.


Image Modified
Figure 8. Flow/Timing Tab.
Image Modified
Figure 9. Special Functions Tab and Leak Test Screen.

Copying and Loading Method

  1. To copy and load a previous method, select File > Copy method from... (Figure 10).
  2. Uncheck the Sample table check box.
  3. Click the button to the right of Method File… to select the method.
  4. Select the method file IODP_NCHSsystem from the folder that has the previous expedition number and click Open.
  5. Select Edit > Edit Elemental Analyzer parameters.
  6. Click File and select the folder that has last expedition number.
  7. Open file IODP_NCHS.eam.
  8. Click File > Save as to save instrument settings into the created folder. Note: Your created folder now contains three files with extensions MTH, EAM, and CC.
  9. Select File > Save method.
  10. Use same method file that you copied in an earlier step and select the folder that you created.
  11. Click Save to save the copied method in the new folder.
  12. Select File > Load method > Load New Method.
  13. Select the folder that contains the copied method and click Open to load method.
  14. Click Send Parameters to Analyzer to send the method to the instrument.
    Warning: When the furnaces begin to heat, helium flows in the circuit and the Furnace, Oven, and TCD lights illuminate on the synoptic panel.
  15. Select View > View Elemental Analyzer Status to confirm status of instrument.

If the analytical configuration changes from CHNS to CNS, select File > Instrument Configuration on the main screen before editing the method (see instructions in the FlashEA Operating Manual, p. 217–221).


Image Modified
Figure 10. Copy Method from Screen.

Checking Maintenance Status

If a reactor is changed, the maintenance program must be reset. The maintenance program shown in Figure 11 gives an indication of reactor life. Reactor deterioration can also be detected on chromatograms.


Image Modified
Figure 11. Maintenance Program Screen


Reactor life is also influenced by the nature of samples, especially water content. So you may have to check for anomalies on the chromatogram on a regular basis, as follows:

  1. Select View > View Maintenance. The Maintenance Program screen shows the reactor status (Figure 11). To reset the maintenance program, select Edit > Reset Maintenance Program.
  2. Select View > Maintenance Status to see the current settings. The program can be changed in two ways.
  3. Click Edit > Default. Original settings for Left 1 and Left 2 were
    • Lifetime: 200
    • Number of runs to warning message: 10

4. Click Edit > Manual. Based on onshore tests, change the settings to

  • Lifetime: 170
  • Number of runs to warning message: 10

5. Click OK to save changes.

Editing the Method

Select Edit > Edit Element Analyzer parameters. The following parameters will display:
–Temperature tab:

...

  1. On the CHNS screen, select Edit > Sample table (Figure 12, below). Clear the sample table and create two new rows.
  2. Fill in the sample table as follows:

...

  • Sample name = Bypass test
  • Filename = Bypass_test_ddmmyy
  • Type = Bypass
  1. Prepare Blank (tin cup with V2O5) and Bypass using a standard material (~0.6–0.7 mg of sulphanilamide or BBOT with V2O5 and tin cup). Place the Blank in the sample tray of the MAS 200 R autosampler for the FlashEA 1112.
  2. Note: Acquire a chromatogram that contains all the components to be identified.
  3. Click OK to close the sample table.
  4. Select File > Save method to save the sample table.
  5. On the CHNS screen, click the green arrow toolbar icon to start running the samples.
  6. Select View > View sample being acquired to confirm run status.

...

  1. After the Bypass sample is done, select Edit > Component table to identify the analysis peaks (Figure 13, below). Instructions for editing the component table are given in the Eager 300 Software Operating Manual, p. 101–115.
  2. Select File > Load Chromatogram, and select bypass test-ddmmyy (last sample run).

10. Click Open to view the chromatogram.
11. Select Time of any component on the table.
12. Move the cursor near the red-gray arrow on the screen and right-click.
13. Adjust the red line on the screen to intersect the top of the peak.
14. Repeat Steps 11–13 for each peak on the chromatogram, and click OK to save changes.
Warning: If peak retention times have changed a lot, measure an additional Bypass sample to confirm the new peak positions. Possible causes of peak retention time shifts include (1) gas leaks (include outside of the instrument), (2) gas flow changes, (3) atmospheric pressure changes, and (4) others.


Image Modified
Figure 12. Sample Table.
Image Modified
Figure 13. Component Table.

Editing the Integration Parameters

  1. After finishing peak identification in the previous section, select Edit > Edit Method on the CHNS screen. Method editing instructions are shown in the Eager 300 Software Operating Manual, p. 74–76.
  2. Select the Integration parameters 3 tab and select Test Integration (Figure 14).
  3. Select File > Load Chromatogram.
  4. Load the chromatogram for the last test result (Bypass).
  5. Select Detect > Peak Integration.
  6. Four default integration parameters will be indicated. To customize detailed settings, click Time events, then Peak Detect. (Figure 15).
  7. To edit the integration parameter table, select any peak time.
  8. Move the cursor near the red-gray arrow. The cursor will convert to a narrow white- black arrow. Left-click to move the arrow.
  9. Click Type of Event in the table, and then click the green inverted triangle.
  10. Choose a type of event and click OK to close the window.
  11. Click OK to save the modification.
  12. Click OK on the Peak detect screen and then close the Chromatogram screen.



Figure 14. Edit Method Screen.

Figure 15. Peak Detect Screen.

Editing Other Parameters

Select the Calculation parameters 4 tab on the Edit Method screen to select the calibration method. Eager 300 has three calibration methods, details of which are described in the FlashEA 1112 Operating Manual, p. 232–234. Generally, we use Linear fit for sediment samples. K-Factor is sometimes useful for samples that are characterized by low component concentrations.

...

Installing the Reactor in the Furnaces

Warning: The reactor must be installed with the furnaces at room temperature or <50°C with gloves. Do NOT use any mechanical tools to screw or unscrew the fixing nut.

Figure 16. Autosampler Removal.
See the FlashEA 1112 Operating Manual, p. 126–129 and 134. Pages 306–308 are also useful when replacing the O-ring of the coupling unions.

  1. Loosen fixing nut and remove autosampler (Figure 16, above).
  2. Delicately remove the used reactor (Figure 17, below, 2) from the right furnace.
  3. Detach O-ring from the used reactor. If O-ring is undamaged, you can use it again.
  4. Gently cleanup coupling union (Figure 17, 1) and O-ring. Wipe metal with alcohol if needed, but do NOT apply alcohol to O-ring. If using alcohol, apply a small amount of Vaseline to the O-ring to fix the cleaned coupling union.
  5. Put a kimwipe or aluminium foil on the coupling union (Figure 17, 3) to protect cleaned coupling union from dust.
  6. Delicately introduce the reactor into the furnace, ensuring that the tube conical end is turned downward toward the base (Figure 17, 4).
  7. Gently press the edge of the reactor until introduction is complete.
  8. Slip on the O-ring with its conical section turned upward.
  9. Manually screw the fixing nut.

...

Connecting the Water Trap (Adsorption Filter)

Warning: Wear gloves and face protector when preparing the adsorption filter. Work with quartz wool under the draft chamber because quartz wool is an inhalation hazard.

...