...
Once the machine is under vacuum (indicated by the SEM beeping twice), navigate to the Operation Panel (Figure 7). IMAGE GOES HERE
Select an accelerating voltage (range: 1 to 30 kV; see table 1 for working guide).
1-5 kV
Delicate or uncoated samples (e.g. microfossils)
5-10 kV
Coated biological samples (e.g. Au coated microfossils, recommended)
10-30 kV
Carbon-coated thin section samples
Table 1. Working guide for accelerating voltage
- Select which detector to use:
- the secondary electron (SE) detector, which returns an image of the topography of the sample's surface and is generally used for micropaleontological identification with SEM stubs
- or the back-scattered electron (BSE) detector, which returns a value based on the average atomic number (Z) of a given material and is generally used on thin sections or grain mounts.
- User guide can have a more in-depth explanation of the two different detectors and show samples of SE and BSE images.
- Click the START button to generate an electron beam.
- Monitor the emission current, which should be around 110uA. If it has diverged by more than ±20 uA, please notify a technician as this either means the filament is about to die or the beam is not stable, and image quality will be subpar. how to change the filament
...
Align the beam
Recommended settings
Sample | Sample preparation | SEM settings |
---|---|---|
Foraminifera or other large (50 um - 1 mm) particles of geological material | Picked and mounted on an SEM stub, optionally sputter coated |
|
Diatoms or nannofossils | Smear slide, sputter coated with Au/Pd |
|
Polished thin section/polished thick section | Polished thin/thick section placed in thin section holder, fastened down using the clips and carbon tape, carbon coated |
|
Larger amounts of geological material, prepared as a polished grain mount | Grain mount placed in grain mount holder, carbon coated |
|
Additional tips
- Dealing with movement
- Sea state
- Jostling the table
- For high magnification (>5000x; unlikely at sea conditions)
- Change the variable strip aperture
- Increase the Spot Size (higher number)
- Change the accelerating voltage up or down
- Adjust the stigmation
- Focus the microscope/decrease the scan speed
...