Petrographic Image Capture and Archiving Tool (PICAT):User Guide
Introduction
This system was designed to digitally photograph Petrographic Thin Section samples, capturing the entire slide in an efficient manner in both cross-polarized and single (plane)-polarized states. Microscopes are commonly used for imaging thin sections, but they do not have a field of view wide enough to image the entire specimen. A whole specimen view is needed to provide a map (or guide) showing the relationship of the entire sample to specifically captured higher-detailed regions. The most common method of obtaining a full specimen image has been with the use of a modified flat bed or film scanner. Whereas these methods have been successful in some instances, they are cumbersome, slow, and tedious when the entire workflow process is considered.
The PICAT (Petrographic Image Capture and Archival Tool) provides an easy and versatile imaging process that captures a publication-quality image (8 x 10 inches at 300 DPI) within seconds rather than minutes. When coupled with a workflow-optimized software package, the time, effort, and training needed to capture thin section image data is significantly reduced. The live image displayed on a computer monitor instantly shows errors in exposure, focus, color balance, and cropping. The operator is confident the image is correct before rendering the sample to digital state.
...
1. Select the type of specimen holder (for instructions on changing to the X/Y stage holder see Hardware > Specimen Holders > > Installing the X/Y Style Holder).
...
3. Two polarization filter holders, one above the tray and one under it, are coupled by a magnetic linkage. The entire assembly can assembly can be rotated out of the way as needed. The filters can be rotated independently or together (the former by turning the polarizing the polarizing [bottom] filter while preventing the filter holder from rotating, the latter by turning the white linkage knob).
...
11. The preview window can be ignored/closed without affecting anything. If Adobe Bridge is already loaded, its icon will flash in the the taskbar (burnt-orange square with the letters "Br"). If Bridge is not already loaded, it will open automatically after a few moments.
12. No action needs to be taken in Bridge at this point. Close/minimize Bridge at any time. The camera automatically creates and saves the saves the JPEG and RAW files.
Processing and Uploading the Image
...
4. Once a sample has been selected, the sample text_id field populates and a new label is generated in the top right that indicates the the new filename, based on the sample selected. If multiple pictures were taken of the same sample, the filename is appended with _2, _3, _4, etc., based on the number of previous images taken and processed. Fill out the rest of the values based on the conditions conditions when the image was taken. In the comments field, enter the description from the Thin Section Request form and the name of the requestor (e.g. "Fossiliferous limestone. Jane Doe.")
5. Click Upload to open a window indicating the upload status. When the upload process is finished another dialog box will appear indicating that the process is complete
...
Clean the polarizing filters with the same care as cleaning a lens. Begin with the least aggressive technique, such as a camelhair brush, followed perhaps by lens cleaner and fluid. Never apply friction to the glass before brushing with a camelhair brush to remove grit or particles. Be aware of substances that may be abrasive to the filter surface.
Disassembly of the Filter Module
Note the following when working with or disassembling the filter module:
...
Color Head for Enlarger | Transformer with Voltage Stabilization | Transformer |
---|---|---|
CATALOG #: 204544 | CATALOG #: 204451 | CATALOG #: 204453 |
Equipped with dichroic interference filters, calibrated in desitometric units up to 180. The dials are indirectly illuminated. Continuously adjustable, illuminated density aperature up to 60 densitometric densities (= two aperature settings). Best possible light distribution due to special diffusor plate. Requires transformer 204451 or 204453 to operate. Items Include: | Output power: 100 W Dimensions: approx. 120 x 80 x 140 mm (4.7 x 3.1 x 5.5 in.) Available for 230 V, 240 V, and 120 V Fuse: 1,6 AT (230/240 V) / 2,5 AT (120 V) 4451: Transformer with starting-current limitation and electronic voltage stabilization. For compensation of voltage deviations (for constant color temperature). Input: 230 V, + 10%, -15%, 50 Hz Output: 11,5 V ± 1%, 50 Hz | Output power: 100 W Dimensions: approx. 120 x 80 x 140 mm (4.7 x 3.1 x 5.5 in.) Available for 230 V, 240 V, and 120 V Fuse: 1,6 AT (230/240 V) / 2,5 AT (120 V) 4453: without voltage stabilization Input: 230 V, 50 Hz Output: 12 V, 50 Hz |
LIMS Component Table
ANALYSIS | TABLE | NAME | ABOUT TEXT |
TSIMAGE | SAMPLE | Exp | Exp: expedition number |
TSIMAGE | SAMPLE | Site | Site: site number |
TSIMAGE | SAMPLE | Hole | Hole: hole number |
TSIMAGE | SAMPLE | Core | Core: core number |
TSIMAGE | SAMPLE | Type | Type: type indicates the coring tool used to recover the core (typical types are F, H, R, X). |
TSIMAGE | SAMPLE | Sect | Sect: section number |
TSIMAGE | SAMPLE | A/W | A/W: archive (A) or working (W) section half. |
TSIMAGE | SAMPLE | text_id | Text_ID: automatically generated database identifier for a sample, also carried on the printed labels. This identifier is guaranteed to be unique across all samples. |
TSIMAGE | SAMPLE | sample_number | Sample Number: automatically generated database identifier for a sample. This is the primary key of the SAMPLE table. |
TSIMAGE | SAMPLE | label_id | Label identifier: automatically generated, human readable name for a sample that is printed on labels. This name is not guaranteed unique across all samples. |
TSIMAGE | SAMPLE | sample_name | Sample name: short name that may be specified for a sample. You can use an advanced filter to narrow your search by this parameter. |
TSIMAGE | SAMPLE | x_sample_state | Sample state: Single-character identifier always set to "W" for samples; standards can vary. |
TSIMAGE | SAMPLE | x_project | Project: similar in scope to the expedition number, the difference being that the project is the current cruise, whereas expedition could refer to material/results obtained on previous cruises |
TSIMAGE | SAMPLE | x_capt_loc | Captured location: "captured location," this field is usually null and is unnecessary because any sample captured on the JR has a sample_number ending in 1, and GCR ending in 2 |
TSIMAGE | SAMPLE | location | Location: location that sample was taken; this field is usually null and is unnecessary because any sample captured on the JR has a sample_number ending in 1, and GCR ending in 2 |
TSIMAGE | SAMPLE | x_sampling_tool | Sampling tool: sampling tool used to take the sample (e.g., syringe, spatula) |
TSIMAGE | SAMPLE | changed_by | Changed by: username of account used to make a change to a sample record |
TSIMAGE | SAMPLE | changed_on | Changed on: date/time stamp for change made to a sample record |
TSIMAGE | SAMPLE | sample_type | Sample type: type of sample from a predefined list (e.g., HOLE, CORE, LIQ) |
TSIMAGE | SAMPLE | x_offset | Offset (m): top offset of sample from top of parent sample, expressed in meters. |
TSIMAGE | SAMPLE | x_offset_cm | Offset (cm): top offset of sample from top of parent sample, expressed in centimeters. This is a calculated field (offset, converted to cm) |
TSIMAGE | SAMPLE | x_bottom_offset_cm | Bottom offset (cm): bottom offset of sample from top of parent sample, expressed in centimeters. This is a calculated field (offset + length, converted to cm) |
TSIMAGE | SAMPLE | x_diameter | Diameter (cm): diameter of sample, usually applied only to CORE, SECT, SHLF, and WRND samples; however this field is null on both Exp. 390 and 393, so it is no longer populated by Sample Master |
TSIMAGE | SAMPLE | x_orig_len | Original length (m): field for the original length of a sample; not always (or reliably) populated |
TSIMAGE | SAMPLE | x_length | Length (m): field for the length of a sample [as entered upon creation] |
TSIMAGE | SAMPLE | x_length_cm | Length (cm): field for the length of a sample. This is a calculated field (length, converted to cm). |
TSIMAGE | SAMPLE | status | Status: single-character code for the current status of a sample (e.g., active, canceled) |
TSIMAGE | SAMPLE | old_status | Old status: single-character code for the previous status of a sample; used by the LIME program to restore a canceled sample |
TSIMAGE | SAMPLE | original_sample | Original sample: field tying a sample below the CORE level to its parent HOLE sample |
TSIMAGE | SAMPLE | parent_sample | Parent sample: the sample from which this sample was taken (e.g., for PWDR samples, this might be a SHLF or possibly another PWDR) |
TSIMAGE | SAMPLE | standard | Standard: T/F field to differentiate between samples (standard=F) and QAQC standards (standard=T) |
TSIMAGE | SAMPLE | login_by | Login by: username of account used to create the sample (can be the LIMS itself [e.g., SHLFs created when a SECT is created]) |
TSIMAGE | SAMPLE | login_date | Login date: creation date of the sample |
TSIMAGE | SAMPLE | legacy | Legacy flag: T/F indicator for when a sample is from a previous expedition and is locked/uneditable on this expedition |
TSIMAGE | TEST | test changed_on | TEST changed on: date/time stamp for a change to a test record. |
TSIMAGE | TEST | test status | TEST status: single-character code for the current status of a test (e.g., active, in process, canceled) |
TSIMAGE | TEST | test old_status | TEST old status: single-character code for the previous status of a test; used by the LIME program to restore a canceled test |
TSIMAGE | TEST | test test_number | TEST test number: automatically generated database identifier for a test record. This is the primary key of the TEST table. |
TSIMAGE | TEST | test date_received | TEST date received: date/time stamp for the creation of the test record. |
TSIMAGE | TEST | test instrument | TEST instrument [instrument group]: field that describes the instrument group (most often this applies to loggers with multiple sensors); often obscure (e.g., user_input) |
TSIMAGE | TEST | test analysis | TEST analysis: analysis code associated with this test (foreign key to the ANALYSIS table) |
TSIMAGE | TEST | test x_project | TEST project: similar in scope to the expedition number, the difference being that the project is the current cruise, whereas expedition could refer to material/results obtained on previous cruises |
TSIMAGE | TEST | test sample_number | TEST sample number: the sample_number of the sample to which this test record is attached; a foreign key to the SAMPLE table |
TSIMAGE | CALCULATED | Top depth CSF-A (m) | Top depth CSF-A (m): position of observation expressed relative to the top of the hole. |
TSIMAGE | CALCULATED | Bottom depth CSF-A (m) | Bottom depth CSF-A (m): position of observation expressed relative to the top of the hole. |
TSIMAGE | CALCULATED | Top depth CSF-B (m) | Top depth [other] (m): position of observation expressed relative to the top of the hole. The location is presented in a scale selected by the science party or the report user. |
TSIMAGE | CALCULATED | Bottom depth CSF-B (m) | Bottom depth [other] (m): position of observation expressed relative to the top of the hole. The location is presented in a scale selected by the science party or the report user. |
TSIMAGE | RESULT | aperture | RESULT aperture: f-stop of the camera for this image (e.g., f/8) |
TSIMAGE | RESULT | camera | RESULT camera: make and model number of the camera |
TSIMAGE | RESULT | date_taken | RESULT date taken: date/time stamp the image was captured |
TSIMAGE | RESULT | focal_length | RESULT focal length (mm): focal length of the camera |
TSIMAGE | RESULT | iso | RESULT iso setting: ISO setting of the camera (e.g., ISO-4000) |
TSIMAGE | RESULT | jpeg_asman_id | RESULT JPG image ASMAN_ID: serial number of the ASMAN link for the JPG/JPEG image |
TSIMAGE | RESULT | jpeg_filename | RESULT JPG image filename: file name of the JPG/JPEG image |
TSIMAGE | RESULT | macro | RESULT macro lens setting: multiplier from 1x to 5x if the macro lens is used (always 1x if standard lens) |
TSIMAGE | RESULT | polarization | RESULT polarization: indicator for no polarization, single polarization, or cross polarization |
TSIMAGE | RESULT | polarizationangle | RESULT polarization angle (deg.): angle of polarization (if measured; usually null) |
TSIMAGE | RESULT | raw_asman_id | RESULT raw file ASMAN_ID: serial number of the ASMAN link for the raw (camera format) uploader file |
TSIMAGE | RESULT | raw_filename | RESULT raw filename: file name for the raw (camera format) uploader file |
TSIMAGE | RESULT | shutter_speed | RESULT shutter speed (s): shutter speed, usually expressed in a fraction of a second (e.g., 1/640 sec.) |
TSIMAGE | SAMPLE | sample description | SAMPLE comment: contents of the SAMPLE.description field, usually shown on reports as "Sample comments" |
TSIMAGE | TEST | test test_comment | TEST comment: contents of the TEST.comment field, usually shown on reports as "Test comments" |
TSIMAGE | RESULT | result comments | RESULT comment: contents of a result parameter with name = "comment," usually shown on reports as "Result comments" |