APCT-3 Run Procedure
374 | Feb 2018
- Tool assembly
- Check pressure case and cutting shoe for damage. Ensure they are in good working order.
- Check O-Rings on the case. Ensure they are not damaged. Apply a light coat of Dow Corning 111 grease to the O-Rings and install on the case. Ensure threads are clean and apply a light coat of Lubriplate No. 630-2 Multipurpose Grease P/N 07298.
- Install electronics package into the pressure case using the installation tool. Apply a liberal amount of Thermal Joint Compound Type 120 from Wakefield Engineering to the temp sensor and install in the case.
- Install data cable in the electronics package.
- Initialization
- Launch WinTemp from the desktop
- Select-Logger
- Select-Battery.Battery Voltage 3005 mv. If below 2750 mv return logger to shore.
- Select-Exit
- Select-Logger
- Select-Setup. A warning message may appear: "The logger contains data. Please clear memory before proceeding."
- Select-OK
- Select-Logger
- Select-Clear Data. A warning message may appear: "Do you really want to clear data?"
- Select-Yes
- Select-Logger
- Select-Setup (If no additions)
- Select-Start Now!
- Select-OK Select after tool starts to run.
Verification window will come up. Tool can run for 18 hours.
- Deployment and tool retrieval
- Remove cable from electronics package.
- Using the installation tool, insert the electronics package into the cutting shoe.
- Attach cutting shoe to pressure case. Tighten snugly with special wrench.
- Run Tool. Recommended 5 min at mud line and 10 min after shooting into the hole.
- When tool returns from the hole, ensure it is loosened on the drill floor and power washed off on the core receiving deck.
- Put the tool in the vice. Use the wrench to remove the cutting shoe from the pressure case.
- Using the installation tool remove the stainless steel spacer with the O-Ring.
- Data download
- Install data cable in the electronics package.
- Run WinTemp:
- Select-WinTemp from desktop.
- Select-Logger
- Select-Read data. You will see the following message: "The logger is still active. Do you want to stop acquisition?"
- Select-Yes
- Select-File > Export and then save as to copy both WTF and the DAT file
- Browse to core folder:
(e.g. C:\data\in\368\1501C\368-1501C-05H)
- Select-Save
- Select-Logger
- Select-Clear Data. Verification message will appear: "Do you really want to clear the logger memory?"
- Select-Yes
- Select-Setup
- Select-Deactivate
- Data processing
- Launch TPFITnew.exe from the desktop.
- Select-Load Data
- Select-Exp 354 Folder
- Select-.dat file
- Select-Open
- Select-Edit Meta Data Add run#-Hole# mbsf depth-Operator Name
- Select-Accept/Pick/To-Pick Center Window of Curve/Start-Pick End-Pick/Done/Compute Contours-Gives Final Temp
- Select-Make Report.