Welcome
This guide is designed to help the scientists and technicians use the Whole Round Multisensor Logger (WRMSL) and upload the data to LIMS. For detailed information on the WRMSL please refer to the Whole Round Multisensor Logger User Guide.
Processing Samples
Running Cores on the WRMSL
- Open IMS Main from the WRMSL computer desktop (Figure 1).
- The program will initialize all of the instruments and then display the main user interface (Figure 2).
- To set the measurement parameters for each of the instruments select DAQ: Measurement Editor. Pro Tip! Always set measurement intervals as whole number divisors of the distance between the sensors (30 cm), or the movement routine will be highly inefficient!
- In the Measurement Editor window, select the instrument you wish to configure and then click in the Instrument Parameters window (Figure 3).
- In the DAQ Parameters window the user may alter the settings for each instrument and may set the instrument to online or offline mode. Select OK. The changes will be written to the configuration file. (Note: If you hit return, it selects the CANCEL CHANGES button. You must click on the ACCEPT CHANGES button to apply changes.)
- GRA Settings include: Measurement Interval, Edge, Acquire Time, Setting the Instrument Online or Offline, and Setting the Control of the measurements On or Off.
- MS Settings include: Measurement Interval, Edge, Number of Points to Average, Setting the Instrument Online or Offline, Setting the Control of the measurements On or Off, and Setting the Drift Correct On or Off.
- PWL Settings include: Measurement Interval, Edge, Number of Waveforms to Stack, Voltage Target, Setting the Instrument Online or Offline, and Setting the Control of the measurements On or Off.
- Select Finished in the Measurement Editor window. Check that the System Status and DAQ Parameters displayed on the main interface window (left side of page) have updated to the values specified.
- DAQ Parameters (steps 4 to 6) should be used only when the instruments parameters need to be changed.
- Set the water core pusher ID by selecting DAQ: Pusher Setup. In the window, scan the water core label. When the user selects Accept the ID of the water core will be written to the configuration file.
- Use this step only when a different water pusher is used, the program already keeps the information of the regular water pusher.
- To begin measuring a core, place a section on the track. Top (blue endcap) should be placed nearest the sensors and bottom should be nearest the pusher.
- Select Start from the Main Interface window.
- Enter the section label identification in the Whole Round MSL Section Information window (Figure 4). This is typically done using the bar code scanner, but may be done manually or by searching the LIMS database. If scanning the label, make sure the cursor is in the SCAN box before scanning the label.
- Select Measure. The pusher will move the core through the instruments and return to its home position once it reaches its furthest limits.
- During a measurement the user may adjust the number of stacked measurements.
- When the first section is finished the Whole Round MSL Section Information window will appear again.
- If there are more sections to be measure place the next section on the track, scan the label, and press measure. Repeat for all sections of the core.
- For best results it is recommended that sections be measured IN ORDER. If it is not possible to measure the sections in order it is recommended that a pusher core is used after each section.
- Once all of the sections from a core have been measured, place the water core standard on the track and press Pusher End Sequence. The standard will push the final section through the sensors and a standard measurement will be made on each sensor.
- The files will be written to: C:\data\IN.
Figure 1. WRMSL Desktop Icon
Figure 2. Main WRMSL User Interface Window with GRA, MS, and PWL displays.
Figure 3. Measurement Editor Window
Figure 4. Whole Round MSL Section Information window
Calibration Procedures
Calibrations should regularly be performed for each of the sensors. Calibration utilities are available under Instruments.
GRA Calibration
- Select Instruments: GRA Calibrate and the GRA Calibration window will open (Figure 5).
- Place the GRA standard (standard with the telescoped aluminum in Figure 6) on the track with the top (end with blue tape) nearest the sensors.The aluminum must be level, A level line is drawn on the bottom of the core. Also, make sure that the ridge of the bottom endcap fits into the groove of the pusher (Figure 7). You may need to loosen the screws of the black plate on the pusher, just enough for it to slide.
- Press OK. The pusher will move the core into the GRA sensor range.
- The data points will be visible on the GRA Calibration graph. When the calibration measurements have completed, the window will display a density check. You may choose to apply or cancel the values at this time (Figure 8).
- After you select apply the data will be written to the configuration file.
- GRA Calibration is complete.
Figure 5. GRA Calibration window.
Figure 6. GRA Standard with telescoped aluminum at the top of the standard
Figure 7. Correct positioning of the Al-water standard on the track
Figure 8. GRA Calibration after data acquisition
PWL Calibration
- In the Instrument menu, select PWL: Calibrate-Utility and the PWL Calibration Utility window will open (Figure 9).
Follow the steps on the right side of the PWL Calibration Utility window. Note: Liner Delay correction is excluded when using the Aluminum standard.
Place the aluminum standard between the transducers. Close the transducers. Aluminum standard width is 76.25mm. Select Laser Offset Correction. Confirm that the Distance matches the standard's width.
Enter the Aluminum Standard velocity. 6295.00 m/s. Select Determine System Delay. Verify that the Velocity matches the Aluminum Standard's velocity.
Open transducers. Remove Aluminum Standard.
Position the Water Standard between transducers. Close transducers. Turn on Liner Delay. Enter the water's temperature (water room temperature) under Temperature. Select Auto H2O Velocity Update. Liner Thickness is 2.580 mm. Click Determine Liner Velocity.
- Accept Changes.
- Open transducers and remove the water standard.
Figure 9. PWave Calibration Utility.
Adjusting Measurement Parameters for the Waveform
You can use the controls Requested Stack and Threshold in PWL Calibration Utility window to adjust the measurement parameters to obtain a better waveform (Figure 8). These parameters may need to be adjusted if the section material has changed.
Requested Stack: Sets the number of waveforms to average to increase the signal-to-noise ratio. Note: stack should not be any lower than 50.
Threshold: A millivolt value is used to find the first arrival peak. For more information on the first arrival determination, see Appendix 1 in WRMSL User Guide. Note: Threshold bar should be above 10m.
Uploading data To Lims
To up load the data into LIMS, MegaUploadaTron (MUT) must be running. If not already started do the following:
- On the desktop click the MUT icon on the task bar (Figure 9) and Login.
- The LIMS Uploader window will open (Figure 10).
- Check Automatic Upload in the lower right hand corner and click minimize. MUT will run in background. The uploaded data will move from C:\data\IN to C:\data\archive. If there is any problem with the uploading of the file, it will go to C:\data\error.
Figure 9. MUT Icon.
Figure 10. LIMS Uploader Window