Skip to end of metadata
Go to start of metadata

You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 15 Next »

Introduction

This user guide will help scientists and technicians correctly use the ASC Scientific IM10/ IM10-30 Impulse Magnetizers (Figure 1). The IM10/IM10-30 Impulse Magnetizer (IM10 and IM10-30) are used to impart an isothermal remanent magnetism (IRM) by exposing the sample to a transient   DC magnetic field at a constant temperature (Do we need a note about DC versus AC fields?). The field is produced by the discharge of energy from a capacitor bank through a coil surrounding the sample cavity. The capacitor bank is charged to the desired voltage (corresponding to the desired field) and then discharged through the coil generating a transient DC field to impart an IRM.

Figure 1: ASC Scientific IM10-30 (left) and IM10 (right) Impulse Magnetizers

For a given applied field, the IRM is the remanent magnetization acquired by magnetic grains with a coercive force less than or equal to the applied field. Once the applied field is greater than all of the grains' coercivities, the remanence reaches a maximum. The field at which the IRM does not get any larger with increasing applied field is the saturation field (Hs) and the resulting remanence is called the saturation isothermal remanent magnetization (SIRM). The opposite DC field necessary to reduce the SIRM is to zero the coercivity of remanence (Hcr). Since SIRM, Hcr, and Hs are dependent on both magnetic grain-size and mineralogy, the IM10/10-30 are ideal for primary magnetic carrier identification.

The IM10 and IM10-30 are capable of applying fields up to 1.2 T and 5T, respectively. The desired applied field is adjustable and controlled by the charging voltage can be approximately adjusted with the voltage adjustment knob. The charging voltages corresponding to specific field levels for the unit on board are given in Appendix A.

Caution: The magnetizer produces a strong magnetic field capable of remagnetizing material within 1 m of the coil.


Procedures

  1. Turn the "Power" switch on. No warm-up period is necessary.
  2. Place the sample in the holder in the desired position. Insert the holder into the sample cavity until it hits the back of the cavity.
  3. Set the voltage adjustment knob so that the ascending voltage displayed on the meter approaches the desired charging voltage slowly. There is a lag time between adjusting the voltage knob and charging of the capacitors to a given voltage. The set voltage is approached asymptotically over a 30- to 60-s time period. The voltage buildup can be monitored via the digital meter. The most accurate and reproducible results will be obtained if the adjusting knob is set to a point slightly above the desired voltage, so that the desired voltage is approached at a rate of 0.5 V/s.
  4. Allow the charging voltage to ascend to the desired value and push the trigger button. For most accurate and reproducible results, do not press the trigger button if the voltage is still building rapidly.
  5. If you overshoot the desired voltage, push the trim button until the displayed voltage drops below the desired value.
  6. Measure the induced magnetization in the cryogenic magnetometer or with the spinner magnetometer.
  7. At the end of each work session, turn the voltage adjustment knob fully counterclockwise and trigger the circuit before turning off the power.
  8. Note: after being magnetized, samples with a high concentration of magnetic minerals may become too strong to be measured in the SRM. In this case a small rock chip can be used for IRM acquisition experiments or SIRM determinations.


Orienting the Samples

ASC Scientific IM10 Impulse Magnetizer

The ASC IM10 impulse magnetizer produces a positive field coming out of the coil towards the user. To apply a field along the desired axis, orient the cube as depicted below. Cubes are drawn as they would be seen by the user as they are placed into the impulse magnetizer. The hatched arrow shown in the following diagrams is the 'up' arrow placed on all cubes during sampling pointing towards the top of the core.  The surface with the arrow is the split surface.


Figure 2: ASC IM10 Sample Cube Orientation


ASC Scientific IM10-30 Impulse Magnetizer

The ASC IM10-30 impulse magnetizer produces a positive field going into the coil, away from the user. To apply a field along the desired axis, orient the cube as depicted below. Cubes are drawn as they would be seen by the user as they are placed into the impulse magnetizer. The hatched arrow shown in the following diagrams is the 'up' arrow placed on all cubes during sampling pointing towards the top of the core.  The surface with the arrow is the split surface.

Figure 3: ASC IM10-30 Sample Cube Orientation


Coil Installation and Removal (IM10-30 only)

The IM10-30 impulse magnetizer is equipped with 3 interchangeable coils numbered 2 through 4 (Figures 4 and 5). Each coil is unique and has its own charging voltage to desired field table. Charging voltage tables for each coil are presented in Appendix X.

Figure 4: IM10-30 Coils

Figure 5: IM10-30 Coils (front)

Coil Change Procedures

  1. Hold the coil with the scribed line on the edge of the front face of the coil pointing up (12 o'clock).
  2. Push the coil into the cavity on the front panel of the unit until the coil protrudes out of the cavity about 2 cm. Align the scribed line on the coil with the scribed line on the front face of the cavity (Figure 6) and push the coil in the remaining distance until it protrudes by only about 2 mm

    Figure 6: Coil change

  3. Screw on the retaining ring until it tightens against the face of the coil. Lightly hand tighten only! You are now ready to use the instrument.
  4. To remove a coil, unscrew the retaining ring. To remove the  #2 coil, insert index and middle fingers of each hand into the sample cavity with your thumbs against the front face of the cavity. Pull the coil straight out. To remove the #3 coil, use the coil extractor tool (Figure 7). Turn the handle of the coil extractor counterclockwise a couple of turns while holding the rubber end. Insert the rubber plug of the extractor into the sample cavity and while pressing it against the side of the cavity, turn the handle clockwise until the plug expands enough to fit snug inside the coil. pull the coil out about 5 cm. Unscrew the handle a few turns until the extractor can easily be removed. Pull the coil straight out. To remove the #4 coil , insert the plastic handle of the extractor slowly into the sample cavity. With a quick jerk, pull the coil out of the cavity about 5 cm. A small finger may also be used to accomplish this. Next, slowly pull the extractor or finger out of the coil and pull the coil straight out.

Figure 7: Coil Extractor Tool


Appendix

A.1 ASC Scientific IM10 Voltage Table


A.2 ASC Scientific IM10-30 Voltage Table




























Credits

Text for this document was taken from the Handbook for shipboard paleomagnetists. ODP Tech. Note, 34., 2007

Archive Versions



  • No labels