378T software & services update test plans
The following process assumes two things:
1) The Developers have removed all the 'old' IODP software.
2) All boxes are updated to Windows 10.
web application | Chieh | Heather | Eric | Zenon | Alexis | Doris | Johanna | ETs | comments |
---|---|---|---|---|---|---|---|---|---|
Auther | |||||||||
will it run? | pass | pass | |||||||
can you log in? | pass | pass | |||||||
can you edit privs? | I do not have rights to add/remove roles but the functionality appears to be okay - EM. I do not see any Roles to add/subtract (because I do not have privs?) - HBÂ Has been used to add permissions to techs - PF | ||||||||
Change Password | |||||||||
will it run? | pass | pass | pass | pass | |||||
can you change your password? | pass | pass | pass | pass | |||||
DESC Asset Report | |||||||||
does it work? | pass | ||||||||
DESC Value List Manager | |||||||||
does it work? | pass | ||||||||
Drill Report | |||||||||
does it work? | pass | ||||||||
Image QAQC Report | |||||||||
does it work? | pass | ||||||||
LIME (385 data) | |||||||||
will it run? | pass | pass | pass | pass | pass | ||||
can you log in? | pass | pass | pass | pass | pass | ||||
can you cancel a sample? | pass | pass | pass | pass | |||||
can you uncancel the same sample? | pass | pass | pass | pass | |||||
can you edit a sample? | pass | pass | pass | pass | |||||
can you reparent a sample? | pass | pass | pass | pass | |||||
can you cancel a test? | pass | pass | pass | pass | |||||
can you reassign a test? | pass | pass | pass | pass | |||||
LIVE (385 data) | Created a Template and panel successfully; both ran; both showed data (AA) | ||||||||
will it run? | pass | pass | pass | pass | |||||
can you use the WR_DATA template to view data? | pass | pass | pass | pass | Did not work on the first attempt (ZM). | ||||
can you use the READY To SPLIT template to view data? | pass | pass | pass | pass | |||||
can you zoom/unzoom (with buttons)? | pass | pass | pass | pass | |||||
can you zoom/unzoom (with changing core #/depth)? | pass | pass | pass | pass | |||||
LORE (385 data) | |||||||||
can you do a hole summary? | pass | pass | |||||||
can you save as a csv and open in Excel? | pass | pass | Bugs submitted in form. EM | ||||||
can you do a core summary? | pass | pass | |||||||
can you save as a csv and open in Excel? | pass | pass | |||||||
can you do a section summary? | pass | pass | |||||||
can you save as a csv and open in Excel? | pass | pass | |||||||
can you get all the drilling reports? | Pass | ||||||||
can you save as a csv and open in Excel? | Pass | ||||||||
can you get a sample report? | Pass | ||||||||
can you save as a csv and open in Excel? | Pass | ||||||||
can you see the core lab whiteboard? | pass | pass | |||||||
can you access the images from all the dropdowns? | pass | pass | Bug submitted in form. EM; PF fixed. | ||||||
can you access the PMAG reports? | pass | ||||||||
can you save as a csv and open in Excel? | pass | ||||||||
can you access chemistry reports? | pass | pass | |||||||
can you save as a csv and open in Excel? | pass | pass | Bug submitted in form. EM Expanded ALKALINITY has a column for both pH & ph. Can you remove the ph column? EM Question: Why are analyte headers ammonium, calcium, magnesium, phosphate, sulfide showing up in the Expanded TITRA_MAN report? We only use this analysis for chloride titration. EM; PF no idea, removed. | ||||||
can you access x-ray reports? | pass | ||||||||
can you save pXRF as a csv and open in Excel? | pass | ||||||||
can you access strat corr reports? | pass | pass | |||||||
can you save as a csv and open in Excel? | pass | pass | |||||||
Standards | |||||||||
does it work? | pass | pass | |||||||
SCORS Uploader & Manager | |||||||||
will it run? | pass | ||||||||
can you login? | pass | ||||||||
can you load an affine table? | pass | ||||||||
can you load a splice? | pass | ||||||||
can you cancel an affine table? | pass | (Whoever left this message, pls) feel free to clean up a bunch of those old files. | |||||||
can you cancel a splice? | pass | ||||||||
Thin Sections Report Builder | |||||||||
will it run? | pass | ||||||||
can you login? | pass | ||||||||
does it behave? | pass | ||||||||
360 deg Composite Processor??? | |||||||||
does it run? | pass | There are no images to test. It does run and allows me to log in. |
Windows applications (using new samples in 378T) | Chieh | Heather | Eric | Zenon | Alexis | Doris | Johanna | ETs | comments |
---|---|---|---|---|---|---|---|---|---|
Core Lab | |||||||||
Core Entry SampleMaster | |||||||||
will it run? | pass | pass | |||||||
can you login? | pass | pass | |||||||
can you create sections on a new core? This assumes a new hole and core have been created from OPs. | pass | pass | I have created three sections on 378T-U9999A-1H (1, 2, 3). There is a CC that is "all to PAL". | ||||||
can you create an IW, HS and PAL samples on that core? | pass | I made a HS on 1H2 (0-5) and a 10 cm IW on 1H3 (63-73). | |||||||
can you print labels? | pass | ||||||||
does the laser engraver run? | pass | PC needs updating to Win10 first. 25Nov : EM Opening the laser software gives Error - clicking past the error allows me to run the laser engraver without issue. see 'software issue' for screen shot of initial error. HB - ISSUE FIXEDÂ - WORKING Dec 7 | |||||||
Chemistry Lab | |||||||||
SampleMaster | |||||||||
is it installed where it is supposed to be? | pass | ||||||||
will it run? | pass | ||||||||
can you login? | pass | ||||||||
can you create splits from a template on the IW sample created? | pass | ||||||||
can you print labels? | pass | Cannot scan HS/CARB labels. 27Nov : EM First label's barcode 'smaller'. Subsequent ones are fine. | |||||||
Cahn balance | |||||||||
is it installed where it is supposed to be? | pass | ||||||||
will it run? | pass | ||||||||
can you login? | pass | ||||||||
can you weigh and save to the Coulometer analysis? | pass | ||||||||
can you weigh and save to the CHNS analysis? | pass | ||||||||
Chlorinity | |||||||||
is it installed where it is supposed to be? | pass | ||||||||
will it run? | pass | ||||||||
can you login? | pass | Font in login screen is too bad. 'USERNAME' & 'PASSWORD' cut-off. 25 Nov : EM | |||||||
can you run a sample? | pass | ||||||||
does the sample result appear in LORE? | pass | ||||||||
Coulometer | |||||||||
is it installed where it is supposed to be? | pass | ||||||||
will it run? | pass | ||||||||
can you login? | pass | ||||||||
can you run a sample? | pass | ||||||||
does the sample result appear in LORE? | pass | ||||||||
Mettler balance | |||||||||
is it installed where it is supposed to be? | pass | ||||||||
will it run? | pass | ||||||||
Spreadsheet Uploader | |||||||||
can it install? | fail/pass | Buglist reported (25Nov) : EM Now able to run. EM | |||||||
can you login? | pass | ||||||||
can you upload a salinity value? | pass | ||||||||
does that value appear in LORE? | pass | ||||||||
Worklist Generator | |||||||||
can it install? | fail/pass | Buglist reported (25Nov) : EM Now able to run. EM | |||||||
can you login? | pass | ||||||||
can you see the CHNS sample after weighing on the CAHN? | pass | ||||||||
can you create your sample list file? | pass | ||||||||
will EA import the sample list file? | pass | ||||||||
MUT | |||||||||
is it installed where it is supposed to be? (NGA, IC, ALK, EA) | pass | ||||||||
can you upload a NGA result? | pass | ||||||||
can you upload an IC result? | pass | ||||||||
can you upload an ALK result? | pass | ||||||||
can you upload an EA result? | pass | ||||||||
Thin Sections Lab | |||||||||
Image Capture | pass | ||||||||
Is Thin Section Report Writer installed and does it run? | fail | Runs Error 'JSON' when click validate - Error fixed, please retest - TB | |||||||
SampleMaster | |||||||||
is it installed where it is supposed to be? | pass | ||||||||
will it run? | pass | ||||||||
can you login? | pass | ||||||||
can you print labels? | pass | ||||||||
XRay Lab | |||||||||
Mettler balance | |||||||||
is it installed where it is supposed to be? | pass | ||||||||
will it run? | pass | ||||||||
LOI Spreadsheet uploader | |||||||||
will it run? | pass | ||||||||
can you see the results in LORE? | pass | ||||||||
MUT | |||||||||
can you upload a 385 result file? | pass | Was a user error. (.png not .pdf files) | |||||||
can you see the results in LORE? | pass | ||||||||
SampleMaster | |||||||||
is it installed where it is supposed to be? | pass | ||||||||
will it run? | pass | ||||||||
can you login? | pass | ||||||||
can you print labels? | pass | ||||||||
pXRF | |||||||||
laptop updated to Win10? | pass | Laptop updated and communicating with XRF. | |||||||
Publications Specialist | |||||||||
is all the required software installed? | Pass | currently installed L2E, Virtual Photo, and DESCLogik | |||||||
can you run the above software? | Pass | ||||||||
SHIL | |||||||||
can you run IMS on the SHIL? | pass | ||||||||
does the VCD print out? | pass | ||||||||
can you upload the image to LORE? | pass | Re-tested SHIL. No problems. | |||||||
can you view the image in LIVE and LORE? | pass | In LORE filenames are lowercase characters. | |||||||
Physical Properties | |||||||||
will IMS run on WRMSL, XIL, NGR, SHMSL? | pass | ||||||||
is MUT installed on the above stations? | pass | WRMSL, NGR and SHMSLÂ pass. XIL doesn't need MUT for uploading the images, but MUT is intalled on the station. | |||||||
can you run a WRMSL section from 378T label created above? | pass | ||||||||
can you run a XIL section from 378T label created above? | pass | ||||||||
can you run a NGR section from 378T label created above? | pass | ||||||||
is NGR Master installed and does it run? | pass | ||||||||
can you run a SHMSL section from 378T label created above? | pass | ||||||||
does the analyses results upload? | pass | WRMSL, NGR and SHMSL pass. | |||||||
can you view the results in LIVE? | pass | WRMSL, NGR and SHMSL pass. XIL is not in LIVE. | |||||||
can you view the results in LORE? | pass | WRMSL, NGR and SHMSL pass. XIL is not in LORE. | |||||||
can you download the results to a CSV and open in Excel? | pass | WRMSL, NGR and SHMSL pass. XIL is not in LORE. | |||||||
is all the required THERMCOM software installed and will it run? | pass | ||||||||
can you view the TC results in LORE? | pass | ||||||||
is the MADMAX software installed? | pass | PC needs updating to Win10 first. 26Nov : EM Updated. Now needs testing. 1Dec : EM LabVIEW to be updated to 2017 from 2014? Retesting? EM | |||||||
can you acquire a wet mass? | pass | ||||||||
can you acquire a dry volume? | pass | ||||||||
can you acquire a dry mass? | pass | ||||||||
can you upload the MADMAX results? | pass | ||||||||
can you view the results from MADMAX in LORE? | pass | ||||||||
All required software on velocity gantry installed and running? | pass | ||||||||
Sample Table | |||||||||
Sample Master | |||||||||
is it installed where it is supposed to be? | pass | ||||||||
will it run? | pass | ||||||||
can you login? | pass | ||||||||
can you create a MAD, PMAG, XRD, CARB sample from a 378T section? | pass | ||||||||
can you print labels? | pass | ||||||||
Core Description | |||||||||
is SampleMaster installed where it is supposed to be? | pass | ||||||||
can SampleMaster run? | pass | ||||||||
can you log into SampleMaster? | pass | ||||||||
can you print a smear slide label? | pass | Bug. Couldn't see samples in the print label pop-up. Bug reported | |||||||
is Image Capture installed where it is supposed to be? | pass | ||||||||
does it run? | pass | Bug Completed & Fixed (AA). | |||||||
is Image Grabber installed where it is supposed to be? | pass | ||||||||
does it run? | pass | ||||||||
is DESCLogik installed where it is supposed to be? | pass | ||||||||
does DESCLogik behave as expected? (Scientist) | pass | ||||||||
does DESCLogik behave as expected? (Admin) | pass | ||||||||
can Virtual Photo Table install and does it run? | pass | ||||||||
Downhole Measurement | |||||||||
is TPFIT uploader installed and does it run? | Pass | Happy now, Eric...? Incredibly!!! 8) EM | |||||||
Paleomagnetics | |||||||||
SRM : MUT running | pass | All old IODP applications still on SRM PC. 26Nov : EM Cleared. 27Nov :EM | |||||||
SRM : SampleMaster running? | pass | ||||||||
D2000 laptop updating to Win10? | pass | ||||||||
will D2000 software run on Win10? | pass | ||||||||
Spinner PC has updated IODP software? | pass | ||||||||
JR-6A data uploaded with MUT? | 'pass' | Cannot fully test upload as this MUT config will not upload results to samples that already have a LIMS result. This message was in fact shown in MUT "no new results" so in that respect, I would say MUT is functioning properly. | |||||||
Kappabridge PC updated to new IODP software? | pass | ||||||||
Kappabridge MUT upload? | pass | ||||||||
Scanning Electron Microscope | |||||||||
is 7 up (SEM uploader) installed? | pass | SEM still on Windows 7. Uploader & MUT runs | |||||||
Does it run? | pass | ||||||||
Can you view results in LORE? | pass | ||||||||
Microscopes | |||||||||
is the required software installed for the microscopes? | pass | ||||||||
Correlator | |||||||||
does the CorrDownloader run? | pass | ||||||||
does Correlator run? | 3rd party tool, no need to test accdg. to PF | ||||||||
Request Code Manager | |||||||||
does it run? | pass |