Adelphi Twin Jet Ampule Sealer Operating Instructions
Operating Instructions
WARNING: CARE MUST BE TAKEN WHEN SEALING AMPULES BECAUSE GLASS REMAINS HOT AFTER SEALING
Ensure valves are shut
Connect the gas and compressed air and set pressures
Open the gas and air valve and light flame
Adjust gas and air to obtain a flame with inner cones of approximately 5 mm and the ends of the flame just overlapping
Note: If the flames curl up there is too much gas
Adjust the height of the ampule rest so the flames play onto the narrow part of the neck of the ampule where the seal is to be formed
Place the ampule in the flames and rotate the base of the ampule slowly
When the glass stem starts to glow red, use tweezers to hold the top of the ampule whilst still rotating the base. Draw off the top as the glass softens. Practice will be needed to form a good seal.