MI5 Flow Chart
MI5 Tool with downloader
Introduction
The Icefield MI5 Multishot Orientation Tool orients Advanced Piston Corer (APC) cores by taking orientation measurements for a short period of time just prior to when the core is taken. During this time the drill string is kept steady while the tool measures and stores: • azimuth • inclination • toolface gravity • toolface magnetism • total magnetic field strength • magnetic dip angle • probe temperature The orientation tool contains triaxial accelerometers and magnetometers and is run on the APC bottomhole assembly (BHA) within a nonmagnetic collar. The tool is synchronized with the Meazura™ MEZ1000 Palm device (yellow object in photo below) prior to deployment. After deployment, the recorded data are downloaded to the Palm device.
Setting up and Starting a Survey
See Figure 1 for a flow chart of the following steps.
Install Batteries in Multishot tool Open Inclin on Palm
Select SURVEY
Name Survey – Click on the symbol with the hand over a keyboard to enter text (Site--Hole--First Core of survey) (ex. 1440A1H) (8 character max)
Select Done
Select OK
Select BY DELAY
Set Delay to 0 seconds
Select OK
Set Shot Interval to 10 Sec
Select OK
In Header Information window set declination to 0
Select OK
Connect Palm to Probe via cable
Select OK
Note battery voltage (not below 9 volts if starting a 12 hour survey) and tool memory (should be 0% full) Select OK in the information window that appears. Green light will flash every second and red light every ten seconds when tool is collecting data.
Disconnect cable from probe. Select OK on Palm. Replace brass cap on tool.
Survey is now running. Place tool in pressure casing and pass it to the core tech. You do not need to press GO. Do not press DONE until survey is complete. Palm can be shut off/go to sleep while survey is collected.
Figure 1. Flow chart for setting up and starting a survey
Ending a Survey and Collecting Data
See Figure 2 for a flow chart of the following steps.
Remove Tool from Pressure Case
Turn on Palm (Palm should display the Monitor Survey Window)
Select DONE
Palm will go to main Inclin screen (Survey, Collect Data, and Manage Data)
Select Collect Data
Select RECOVER DATA
Select YES when confirmation screen appears
Connect Probe to Palm via Cable
Select YES
Wait for Download to complete. Large surveys will take longer.
Palm will go to the main Inclin Screen (Survey, Collect Data, Manage Data) when download is complete. Palm may go to sleep when download is complete.
Disconnect Probe from Palm
Figure 2. Flow chart for ending a survey and collecting data.
In order to check that the data is now on the Palm
Select MANAGE DATA when recover is complete
Look to make sure the survey appears in BOLD in the Manage Data menu
Survey will only show up if you started and ended the survey with the same Palm. If it doesn't appear and you used the same Palm, try to Recover the data again OR go to HotSync Instructions and attempt to HotSync the data.
Hotsync Data to PC and Process with Inclin
See Figure 3 for a flow chart of the following steps.
Go to Palm Main Window (Click on the little house) Connect Palm to PC via USB cable
Select HotSync on Palm
Press HotSync Symbol (square with a circle and two arrows) in the middle of the screen on Palm
Wait. Large Files may take a few minutes. Palm will sound when HotSync begins and ends. A status bar will appear on the PC and the Palm.
Open the Inclin Data folder from the desktop.
Open the Recover Data Folder from within the Inclin Data folder
Rename the .txt file so that it matches the Survey name assigned when the survey was initiated the hole name (ie U1440A1Htxt)
Log in to Novell
Copy the .raw and the .txt files to Data (JR1)– Data1-- Paleomagnetics CORIENT ICEFIELD folder on server
Ensure all data is backed up. Wait to clear the probe and palm until you need the tool for the next run. This will allow the magnetics technician to check that the data downloaded correctly and fix any outstanding issues.
Figure 3. Flow chart for syncing data to a PC and processing.
.
Clearing Data from the Palm and the Probe
Note: Only clear the Palm and Probe if necessary. Otherwise leave clearing the tool and palm to the Paleomagnetism Technician. Do NOT clear the probe before the data has been hotsynced and backed up to the server.
Reasons to clear the tool and palm:
All three tools have been run and one of them must be deployed before the paleomagnetism technician arrives on shift
One of the probes is not functioning and you must reuse a tool
Clearing the Probe:
Connect Palm to Probe
Open Inclin on Palm
Tap the word Inclin at the top of the main Inclin window (window displaying survey, collect data, recover data)
Select Options menu
Select Clear Probe
Confirm that you want to clear the probe memory. Select NO to continue clearing the probe (window asks if you want to abort the process)
Confirm that you want to continue. Select Yes.
Palm will contact Probe and notify user when probe is cleared.
Clearing the Palm:
The Palm should be cleared before starting a new survey otherwise the data files will contain multiple surveys when HotSynced. Make sure the data is already downloaded and backed up. If possible choose to use a Palm that is already clear.
Select Manage Data from the main Inclin screen on the Palm
Tap on Select Survey at the top of the screen
You may choose to delete selected items (if you have selected specific items) or you may delete all of the surveys on the palm. Typically Delete all will be the option used.
The warning menu will appear explaining that the survey names should not be reused until the probe is also cleared. Select OK.
Confirm that you want to delete all holes by selecting Yes.
A second confirm window will appear telling the user the data has not been backed up. This is due to our method of using Recover Data. As long as the survey name is no longer in bold text and you have confirmed the data is on a PC and on the server, select YES.
Icefield Tool Work Flow
Set up and start survey. The Palm and Tool should not have existing surveys when a new survey is started. If they do, try to use a different tool/Palm if possible. Existing data may indicate that the previous survey has not been downloaded or verified yet.
Record User Name, Tool Number (2043, 2052, or 2007), Palm ID (Vanilla Ice, Ice T, or Ice Cube), Time Survey started (in GMT), Survey Name (Site--Hole-- FirstCore), and any additional comments in blue core orientation notebook.
Deploy Orientation tool downhole. Deployment should be approximately 8-12 hours, unless otherwise specified.
When the survey time is almost over, setup a second tool. Communicate with the Core Tech to let them know that a new tool needs to be deployed.
Deploy second tool
Recover data from the first tool and back it up to the Computer: DATA (JR1): data 1: 13.5 Paleomagnetics CORIENT ICEFIELD core orientation folder.
Once data is backed up, set the tool and Palm aside for the paleomagnetics technician. Once they verify they survey data has downloaded properly, they will clear the tool and Palm
If you have used all three tools and no paleomagnetics technician has cleared the tools, ensure the data has been backed up and clear a tool and palm (using flow chart instructions). Make sure to open the data files to verify the transfer was successful.
Repeat process of recovering and deploying tools. If you have used all three tools and no paleomagnetics technician has cleared the tools, ensure the data has been backed up and clear a tool and palm (using flow chart instructions)
Credits
This document originated from the file MI5Flowchartversion3.pdf (see Archived Versions below for a pdf copy) that was written by B. Novak (2015-03-24). Credits for subsequent changes to this document are given in the page history.
Archived Versions
MI5FlowChart_exp378.pdf: PDF version of this wiki page as of 2020-02-24
MI5Flowchartversion3.pdf: Original file written by B. Novak