Pmag Check Ups
Pmag Check Ups.
Always keep Cryowatch running and visible on multiple screens.
Record temperatures for the SRM every week or so (confluence page is open on SRM computer). There are four values:
shield and SQUIDs on the 202 control-unit (Tsh and Tsq). Turn of the unit after to conserve He.
compressor oil and H2O-in (both from cryowatch).
If the SRM computer needs rebooting, there is the startup protocol posted by the SRM computer.
If it is necessary to switch to the backup Haskris follow the protocol on the main Haskris. Summary as follows.
Turn on backup unit.
Turn both black valves on the plumbing circuit 180° (from up to down).
Turn off main Haskris unit.
Make sure the backup unit has not slightly overflowed during switching (rags as at the backup unit if necessary).
To switch back do the above in reverse.
Charge all 3 palm pilots towards the end of the tie-up. Cables and palm pilots are in the downhole MOT drawer.
Check the sump on the ship chill-water intake. When the sediment reaches the marked line, it should to be emptied. There is a flashlight next to it to make inspection easier.
Turn on the bypass and turn off both valves to the sump.
Use the strap wrench to unscrew the sump.
Dump the sediment and briefly clean the sump.
Thread the sump back on until it stops threading.
Turn back on both valves to the sump.
Turn of the bypass valve.
Verify that the sump is not leaking.
You deserve a beer. Thank you.