HYDROGEN GENERATORS - Parker-Balston

Apr 1, 2022 After a big roll, the same exact thing that happened to Generator #1 yesterday happened to Generator #3 today. This confirms that it was caused by the ship rolling and that there are no issues with the O2 vent. 

Mar 31, 2022 After a big roll, Generator #1 (currently SRA) started beeping continuously and displaying an "O2 vent blocked" message. A hard reset was required to clear the error message. To do this, plug the hard reset dongle (can be found in drawer C-D2) into the port on the back of the hydrogen generator. Turn the unit off, wait a moment, turn the unit back on holding down the middle (RESET) button below the display. This clears the "O2 vent blocked" message. 

Apr 30, 2019 Generator beeps warning if the water level exceeds the max level indicated on the plastic reservoir. If the ship roll is significant, the alarm will sound due to water sloshing above the max level. Pipette water out to get it to stop.