Temperature Controller Settings

Temperature Controller settings as of Exp 391.  See the DTB Series manual or the SOLO 4848 manual for more information on how to setup the controller.

Temperature Controller Settings-SHIL


Temperature Controller Settings-SHIL


Operation Mode

Code

Meaning

Selection

r-s

(run or stop)

RUN

SP

decimal point position

1

AL 1H

Upper limit alarm

50

LOC

setting lock mode

off

out 1

display and adjust output value of 1st output group

0

Out 2

display and adjust output value of 2nd output group

0

 

 

 

Regulation Mode

At

auto tuning

off

Pcd0

PID Control setup - set which group to save parameters to

39

Pcd1

 

0

Pcd2

 

0

Pcd3

 

0

Pcd4

 

Auto

Su0

PID setting N=0

39

P0

Proportion band setting N=0

30

C0

Ti setting: n=0

65

d0

Td setting: n=0

63

Cof0

Integral deviation setting: n=0

0

HtPd

Heating/cooling control cycle setting

30

HCPd

Control cycle setting of 2nd output group

5

CoEF

P value of 1st and 2nd output group during dual loop output control

0.1

Dead

deadband 

-1

tPof

regulate temperature deviation value

0

 

 

 

Initial Setting Mode

CnPt

Set Input type

K

tPun

Set temperature unit

C

tP-H

set upper limit of temperature range

999.9

tP-L

set lower limit of temperature range

-199.9

Ctrl

Set control mode

PCd

S-HC

Select heating/cooling or dual output control

H1C2

ALA1

Alarm 1 mode setting

6

ALA2 

alarm 2 mode setting

0

SALA

set system alarm

Off

CoSH

Communication write function enable/disable

On

C-SL

ASCII, RTU Communication formats selection

ASCII

C-no

Communication  address setting

1

bPS

communication baud rate setting

9600

Len

Data length setting

7

Prty

parity bit setting

Even

Stop

stop bit setting

1