NI/LabVIEW O2 monitoring

Chassis : NI-9147 ethernet expansion chassis for C Series modules

Nov 22, 2021 LabVIEW 2021 32-bit installed on ChemTech PC.

Dec 21, 2020 The chassis & power supply are ready to be installed by the Chemistry Technicians where they would like it. We are awaiting for 32-bit LabVIEW to be installed on the ChemTech PC so the programming can commence. Please contact @jr_moortgat (on shore) for questions.

Get MCS to install 32-bit LabVIEW. Dec 17, 2020 Email sent for this request

 

Hardware

MAC 00:80:2F:31:57:92

Physically connected to LAN–070

hostname : NI9147-FCL

IP : 165.91.72.104; Subnet : 255.255.255.128; Gateway : 165.91.72.2; DNS : 165.91.150.5

Software installed

 

Module : NI-9871 4-port RS485 C Series serial interface module

Twinfly P1-150-24 external power supply (+24 V) connected to NI 4POS female microfit pigtail, 2 conductor (P/N 198159-01).

Port 1 : RJ-50 connection to O2 sensor via RS-485.

AMI 2001RS : RS485

pin A - DATA (pos) to RXD (pos). Pins 3 & 7 on connector

pin B - DATA (neg) to RXD (neg). Pins 2 & 6 on connector.

 

LabVIEW programming

This NI link is a good starting point.